Page 2
♦ ♦ PRECISION INSTRUMENTS FOR TEST AND MEASUREMENT IET LABS, INC. www.ietlabs.com Long Island, NY • Email: info@ietlabs.com TEL: (516) 334-5959 • (800) 899-8438 • FAX: (516) 334-5988...
Page 3
WARRANTY We warrant that this product is free from defects in material and workmanship and, when properly used, will perform in accordance with applicable IET specifi cations. If within one year after original shipment, it is found not to meet this standard, it will be repaired or, at the option of IET, replaced at no charge when returned to IET.
Page 4
WARNING OBSERVE ALL SAFETY RULES WHEN WORKING WITH HIGH VOLTAGES OR LINE VOLTAGES. Dangerous voltages may be present inside this instrument. Do not open the case Refer servicing to qualifi ed personnel HIGH VOLTAGES MAY BE PRESENT AT THE TERMINALS OF THIS INSTRUMENT WHENEVER HAZARDOUS VOLTAGES (>...
1689-9640 Overview The optional Digibridge IEEE-488.2 interface board provides IEEE-488.2 (GPIB) and RS-232 Serial interfaces to the Digibridge 1689/1693 series instruments via 24-pin GPIB and 9-pin male serial connectors. The interface can be used to connect directly to a controller or to a bus comprising of multiple instruments and a controller.
1689-9640 GPIB Address DIP Switch The GPIB Address of the Digibridge is set by the SW2 DIP Switch. The numeric value is set by setting the switch on or off for the corresponding bit with the OFF position for binary 1 and ON position for binary 0.
1689-9640 Jumpers The Digibridge IEEE-488.2 interface board has three jumpers, JP1, JP2, and JP3. For normal operation, all three jumpers should be open (no shorting blocks inserted). JP1 enables the RS-232 interface driver auto power off. JP2 forces firmware update mode via RS-232 serial interface. If JP2 is set, the board will be forced to await firmware update and the Digibridge will be held in reset state.
1689-9640 Legacy Digibridge IEEE-488 Commands The legacy IEEE-488.1 Digibridge commands are supported by the IEEE-488 interface board for compatibility with software and drivers written for the original Digibridge IEEE-488.1 interface boards. Category Description Type Code Display Bin # 2 byte...
Page 11
1689-9640 Range Control Hold range 2 byte Hold range 1 2 byte Hold range 2 2 byte Hold range 3 2 byte Hold range 4 2 byte Autorange 2 byte Frequency Value (in kHz) ═ f Floating point Zero Calibration...
Page 12
1689-9640 Normal routine (highest speed) Measurement Routine 2 byte Signal reversal 2 byte (use for low frequency hum rejection) Shorting between measurements 2 byte (used for reducing transient-recovery time within measuring circuit, when measuring with bias; NOT FOR DISCHARGING THE DUT)
1689-9640 IEEE 488.2 / SCPI Digibridge Command Summary The IEEE-488.2 / SCPI commands are used to communicate with the Digibridge via the IEEE-488 or RS-232 serial interfaces replacing and/or supplementing the legacy commands. Command Parameter(s) Description Legacy CONFigure: PARameter Selects the measured parameter...
Page 14
1689-9640 RANGe Sets the measurement range HOLD Hold range Hold range 1 Hold range 2 Hold range 3 Hold range 4 AUTo Autorange DELay Sets the trigger delay time <numeric> Real value in ms J<numeric>; ITFactor Sets the integration-time multiplying factor <numeric>...
Page 15
1689-9640 BIN: Sets the limit entries for binning <numeric> Bin # B<numeric> HIGH <numeric> Real value high limit in % H<numeric>; <numeric> Real value low limit in % L<numeric>; BINSummary Enables or disables bin summary output DISable Disable output ENABle...
Page 16
1689-9640 COMMunicate: GPIB: ADDRess? Reads the instrument GPIB address CONFigure Enables or disables the GPIB interface ENABle Enable GPIB interface DISable Disable GPIB interface (if using RS-232) SERial: CONFigure Enables or disables the RS-232 serial interface ENABle Enable RS-232 interface...
Page 17
1689-9640 OPERation: [EVENt]? Reads the Event Register CONDition? Reads the Condition Register ENABle <numeric> Sets the Enable Mask Register ENABle? Reads the Enable Mask Register QUEStionable? Reads the Event Register QUEStionable: [EVENt]? Reads the Event Register CONDition? Reads the Condition Register ENABle <numeric>...
1689-9640 :CONFigure:APERture Syntax: :CONFigure:APERture {FAST|MEDium|SLOW } Description: Sets the measure rate. Parameter: Value Description FAST Sets measurement rate to fast MEDium Sets measurement rate to medium SLOW Sets measurement rate to slow Equivalent legacy commands: S0, S1, S2 Equivalent key:...
1689-9640 Description: Set the measurement range. Parameter: Value Description HOLD Hold range Hold range 1 Hold range 2 Hold range 3 Hold range 4 AUTo Autorange Equivalent legacy command: R0, R1, R2, R3, R4, R5 Equivalent key: :CONFigure:DELay Syntax: :CONFigure:DELay <numeric>...
1689-9640 Value Description <numeric> # of measurements Equivalent legacy command: A<numeric>; Equivalent key: :CONFigure:FREQuency Syntax: :CONFigure:FREQuency <numeric> Description: Sets the frequency used for measurement signal. Parameter: Value Description <numeric> Frequency real value in kHz Equivalent legacy command: F<numeric>; Equivalent key: none...
1689-9640 :CONFigure:CURRent Syntax: :CONFigure:CURRent <numeric> Description: Set the measurement constant current level. Parameter: Value Description <numeric> Set measurement current to real value (in mA) Equivalent legacy command: i<numeric>; Equivalent key: Note: The :CONFigure:CURRent command is available only on the Digibridge 1693M model.
1689-9640 Equivalent key: none :CONFigure:NOMinal Syntax: :CONFigure:NOMinal <numeric> Description: Set the nominal real value <numeric> in ohms, henries, or farads. Parameter: Value Description <numeric> Nominal real value Equivalent legacy command: N<numeric>; Equivalent key: :CONFigure:BLANking Syntax: :CONFigure:BLANKing <a numeric> <b numeric>...
1689-9640 :CONFigure:BIN:HIGH Syntax: :CONFigure:BIN:HIGH <numeric> Description: Set the limit entry bin high limit (in %). Parameter: Value Description <numeric> Real value high limit in % Equivalent legacy command: Hh; Equivalent key: :CONFigure:BIN:LOW Syntax: :CONFigure:BIN:LOW <numeric> Description: Set the limit entry bin low limit (in %).
1689-9640 Parameter: Value Description NONe No data output Bin # QBIN QDR, Bin # RBIN RLC, Bin # RQDR RLC, QDR RQBin RLC, QDR, Bin # RQBFail RLC, QDR, Bin # on FAIL only RQBPass Same on FAIL; Bin # on PASS...
1689-9640 :SYSTem:COMMunicate:GPIB:ADDRess? Syntax: :SYSTem:COMMunicate:GPIB:ADDRess? Description: Reads the GPIB address of the Digibridge. (Query Only) The GPIB address can only be changed by changing the SW2 DIP switch setting on the Digibridge IEEE.2 Communication board. Parameter: Value Description 1 to 30...
1689-9640 :STATus:OPERation:CONDition? Syntax: :STATus:OPERation:CONDition? Description: Reads the value of the Operation Status Condition Register. (Query Only) Equivalent legacy command: none Equivalent key: none :STATus:OPERation:ENABle Syntax: :STATus:OPERation:ENABle <numeric> :STATus:OPERation:ENABle? Description: Sets the Operation Enable Mask Register to specified numeric value. :STATus:OPERation:ENABle? reads the current value of the Operation Enable Mask Register without clearing the register value.
1689-9640 :STATus:QUEStionable:ENABle Syntax: :STATus:QUEStionable:ENABle <numeric> :STATus:QUEStionable:ENABle? Description: Sets the Questionable Enable Mask Register to specified numeric value. :STATus:QUEStionable:ENABle? reads the current value of the Questionable Enable Mask Register without clearing the register value. Parameter: Value Description 0 to 255 Numeric value range...
Syntax: *IDN? Description: Reads out the Digibridge instrument information (manufacturer, model number, serial number, FPGA version number, firmware version number) of the Digibridge. Example: “IET Labs, 1693M, 20110228, 1.0, 1.0” Equivalent legacy command: none Equivalent key: none *OPC Syntax: *OPC Digibridge IEEE-488.2 Board...
1689-9640 Description: Sets the OPC bit (bit 0) of the Standard Event Status Register when all pending operations have completed or ended. Equivalent legacy command: none Equivalent key: none *OPC? Syntax: *OPC? Description: Reads 1 if all pending operations have completed or ended.
1689-9640 Equivalent legacy command: none Equivalent key: none *TST? Syntax: *TST? Description: Initiates self-test and returns 0 if all tests passed or error code if any test failed. Equivalent legacy command: none Equivalent key: none *WAI Syntax: *WAI Description: Waits until all commands previously sent have completed.