Model 090B82 CALIBRATION ADAPTOR Installation and Operating Manual For assistance with the operation of this product, contact PCB Piezotronics, Inc. Toll-free: 800-828-8840 24-hour SensorLine: 716-684-0001 Fax: 716-684-0987 E-mail: info@pcb.com Web: www.pcb.com...
Page 2
Piezotronics, user servicing or repair is critical test. not recommended and, if attempted, may void the factory warranty. Routine PCB Piezotronics maintains an ISO- maintenance, such as the cleaning of 9001 certified metrology laboratory and electrical connectors, housings, and offers calibration services, which are...
Page 3
50% of the general contact numbers are: replacement cost returned item(s). PCB will provide a price PCB Piezotronics, Inc. quotation replacement 3425 Walden Ave. recommendation for any item whose Depew, NY14043 USA repair costs would exceed 50% of...
OPERATION MANUAL FOR BALLISTICS PRESSURE SENSOR CALIBRATION ADAPTOR FOR AMMUNITION TESTING Model 090B INTRODUCTION INSTALLATION OF THE CARTRIDGE CASE The Series 090B Calibration Adaptors are designed to facilitate calibration of the Model 117B Conformal (Refer to Installation Dwg. No. 090-2 -90) Pressure.
OPERATION MANUAL FOR BALLISTICS PRESSURE SENSOR CALIBRATION ADAPTOR FOR AMMUNITION TESTING Model 090B Insert the empty cartridge case into the chamber COLLECTING THE DATA body (Item 1) as shown. Inspect the cartridge case retainer (Item 3) for an O-ring groove, It is important to proceed carefully while performing which will be found on certain calibrators.
OPERATION MANUAL FOR BALLISTICS PRESSURE SENSOR CALIBRATION ADAPTOR FOR AMMUNITION TESTING Model 090B Without dropping pressure, immediately OPERATION proceed to next pressure level, set pressure and record reading. Apply pressure slowly, starting with the lowest pressure working toward the highest, set the pressure NOTE: For sensors with full scale range of to specific values precisely and read corresponding...
Page 9
OPERATION MANUAL FOR BALLISTICS PRESSURE SENSOR CALIBRATION ADAPTOR FOR AMMUNITION TESTING Model 090B Figure 1 Drawing Number: 21105 Revision: NR...
Page 10
OPERATION MANUAL FOR BALLISTICS PRESSURE SENSOR CALIBRATION ADAPTOR FOR AMMUNITION TESTING Model 090B Figure 2 Drawing Number: 21105 Revision: NR...
Page 11
Model 117B75 CONFORMAL BALLISTICS PRESSURE SENSOR Installation and Operating Manual For assistance with the operation of this product, contact PCB Piezotronics, Inc. Toll-free: 800-828-8840 24-hour SensorLine: 716-684-0001 Fax: 716-684-0987 E-mail: info@pcb.com Web: www.pcb.com...
Page 12
Piezotronics, user servicing or repair is critical test. not recommended and, if attempted, may void the factory warranty. Routine PCB Piezotronics maintains an ISO- maintenance, such as the cleaning of 9001 certified metrology laboratory and electrical connectors, housings, and offers calibration services, which are...
Page 13
50% of the general contact numbers are: replacement cost returned item(s). PCB will provide a price PCB Piezotronics, Inc. quotation replacement 3425 Walden Ave. recommendation for any item whose Depew, NY14043 USA repair costs would exceed 50% of...
Page 15
Component Name Hazardous Substances Lead Mercury Cadmium Chromium VI Polybrominated Polybrominated (Pb) (Hg) (Cd) Compounds Biphenyls Diphenyl (Cr(VI)) (PBB) Ethers (PBDE) Housing PCB Board Electrical Connectors Piezoelectric Crystals Epoxy Teflon Electronics Thick Film Substrate Wires Cables Plastic Solder Copper Alloy/Brass This table is prepared in accordance with the provisions of SJ/T 11364.
Page 16
OPERATION MANUAL FOR CONFORMAL BALLISTICS PRESSURE SENSOR Series 117B Once the proper thickness spacer is found, removal INTRODUCTION and re-installation now becomes a routine matter. The Model 117B Conformal Sensor measures INSTALLATION pressure inside metallic cartridges and paper or plastic shotshell, during firing of the round, without drilled holes in the round.
Page 17
OPERATION MANUAL FOR CONFORMAL BALLISTICS PRESSURE SENSOR Series 117B the curvature. In these latter cases, the Once the proper thickness is found for sensor forward side is identified with the perfect flushness, the sensor may be legend “FWD” etched on the guide collar to removed and reinstalled using this same which the slotted clamp is attached.
OPERATION MANUAL FOR CONFORMAL BALLISTICS PRESSURE SENSOR Series 117B LOW-END NONLINEARITY It will be noticed during calibration, especially with metallic cartridges, that a certain amount of pressure is necessary before linear output is attained from the conformal sensor. This is due to the fact that since the cartridge cases are made nominally smaller in diameter than the chamber to allow easy insertion and extraction, the case must be expanded slightly by the internal...
Page 19
OPERATION MANUAL FOR CONFORMAL BALLISTICS PRESSURE SENSOR Series 117B Drawing Number: 21086 Revision: NR...
Page 20
OPERATION MANUAL FOR CONFORMAL BALLISTICS PRESSURE SENSOR Series 117B Drawing Number: 21086 Revision: NR...
Page 21
AN IMPROVED TECHNIQUE FOR UTILIZATION OF CONFORMAL BALLISTICS SENSOR CALIBRATION DATA The PCB Series 117B Conformal Ballistics Sensor (patented) measures chamber pressure in a novel manner. The sensitive surface of the sensor, when properly installed in the gun chamber, conforms or blends in precisely with the inside surface of the chamber actually forming a portion of the wall adjacent to the cartridge case.
Page 22
AN IMPROVED TECHNIQUE FOR UTILIZATION OF CONFORMAL BALLISTICS SENSOR CALIBRATION DATA output Sensitivit (Eq. 1) input The units for sensitivity are commonly pC/psi or mV/psi. Note that in this case, the sensitivity is identical to the slope of the line, since the line does intersect the origin by previous definition.
Page 23
AN IMPROVED TECHNIQUE FOR UTILIZATION OF CONFORMAL BALLISTICS SENSOR CALIBRATION DATA At the lower pressures, the sensor has very little output (Eq. 4) because there is a certain amount of clearance between the cartridge case and the chamber wall. Obturation of the cartridge occurs in this region.
Page 24
AN IMPROVED TECHNIQUE FOR UTILIZATION OF CONFORMAL BALLISTICS SENSOR CALIBRATION DATA The straight line is the best line that can be constructed Set range of charge amplifier output to 1000 pC/volt (or through the linear portion of the curve as shown in other desired range).
Page 25
AN IMPROVED TECHNIQUE FOR UTILIZATION OF CONFORMAL BALLISTICS SENSOR CALIBRATION DATA Dial Gain Standardization The indicated peak pressure P2 from the conformal, which through scale factoring can be read to read Many charge amplifiers feature a calibrated gain pot for directly in psi (on a digital peak meter for example) is dialing in sensor sensitivity for the purpose of setting simply added to the offset pressure to arrive at true peak...
Page 26
AN IMPROVED TECHNIQUE FOR UTILIZATION OF CONFORMAL BALLISTICS SENSOR CALIBRATION DATA Calibrate Signal Insertion Another method that is useful especially when utilizing a storage oscilloscope readout is to use a charge calibrator to inject a full scale charge signal into the charge amplifier.
Page 27
AN IMPROVED TECHNIQUE FOR UTILIZATION OF CONFORMAL BALLISTICS SENSOR CALIBRATION DATA * Adds offset pressure (not available from PCB). Drawing Number: 21210 Revision: NR...
Page 28
Model Number Revision: R CHARGE OUTPUT PRESSURE SENSOR ECN #: 46956 117B-40K Performance ENGLISH OPTIONAL VERSIONS [2][3] Sensitivity(± 20 %)(for straight oil) 0.110 pC/psi 0.016 pC/kPa Optional versions have identical specifications and accessories as listed for the standard model Measurement Range 10 to 35 kpsi 68,950 to 241,300 kPa except where noted below.
Page 29
Model 090B82 CALIBRATION ADAPTOR Installation and Operating Manual For assistance with the operation of this product, contact PCB Piezotronics, Inc. Toll-free: 800-828-8840 24-hour SensorLine: 716-684-0001 Fax: 716-684-0987 E-mail: info@pcb.com Web: www.pcb.com...
Page 30
Piezotronics, user servicing or repair is critical test. not recommended and, if attempted, may void the factory warranty. Routine PCB Piezotronics maintains an ISO- maintenance, such as the cleaning of 9001 certified metrology laboratory and electrical connectors, housings, and offers calibration services, which are...
Page 31
50% of the general contact numbers are: replacement cost returned item(s). PCB will provide a price PCB Piezotronics, Inc. quotation replacement 3425 Walden Ave. recommendation for any item whose Depew, NY14043 USA repair costs would exceed 50% of...
Page 33
Component Name Hazardous Substances Lead Mercury Cadmium Chromium VI Polybrominated Polybrominated (Pb) (Hg) (Cd) Compounds Biphenyls Diphenyl (Cr(VI)) (PBB) Ethers (PBDE) Housing PCB Board Electrical Connectors Piezoelectric Crystals Epoxy Teflon Electronics Thick Film Substrate Wires Cables Plastic Solder Copper Alloy/Brass This table is prepared in accordance with the provisions of SJ/T 11364.
Page 34
PCB Piezotronics Inc. claims proprietary rights in REVISIONS the information disclosed hereon. Neither it nor any reproduction thereof will be disclosed to others DESCRIPTION without the written consent of PCB Piezotronics Inc. UPDATED O-RING 43263 MODEL 117B75 CONFORMAL BALLISTIC PRESSURE SENSOR...
Page 35
Model 117B75 CONFORMAL BALLISTICS PRESSURE SENSOR Installation and Operating Manual For assistance with the operation of this product, contact PCB Piezotronics, Inc. Toll-free: 800-828-8840 24-hour SensorLine: 716-684-0001 Fax: 716-684-0987 E-mail: info@pcb.com Web: www.pcb.com...
Page 36
Piezotronics, user servicing or repair is critical test. not recommended and, if attempted, may void the factory warranty. Routine PCB Piezotronics maintains an ISO- maintenance, such as the cleaning of 9001 certified metrology laboratory and electrical connectors, housings, and offers calibration services, which are...
Page 37
50% of the general contact numbers are: replacement cost returned item(s). PCB will provide a price PCB Piezotronics, Inc. quotation replacement 3425 Walden Ave. recommendation for any item whose Depew, NY14043 USA repair costs would exceed 50% of...
Page 39
Component Name Hazardous Substances Lead Mercury Cadmium Chromium VI Polybrominated Polybrominated (Pb) (Hg) (Cd) Compounds Biphenyls Diphenyl (Cr(VI)) (PBB) Ethers (PBDE) Housing PCB Board Electrical Connectors Piezoelectric Crystals Epoxy Teflon Electronics Thick Film Substrate Wires Cables Plastic Solder Copper Alloy/Brass This table is prepared in accordance with the provisions of SJ/T 11364.
Page 40
PCB Piezotronics Inc. claims proprietary rights in REVISIONS the information disclosed hereon. Neither it nor any reproduction thereof will be disclosed to others DESCRIPTION without the written consent of PCB Piezotronics Inc. .565 .025 14.35 UPDATE TO CURRENT STANDARDS...