Brainlab 17902 System And Technical User Manual

Wall touch screen

Advertisement

Quick Links

WALL
TOUCH SCREEN
Ref-No. 17902
System and Technical User Guide
Revision 1.0
Date of issue: 2023-06-28 (ISO 8601)
Copyright 2023, Brainlab AG Germany. All rights reserved.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 17902 and is the answer not in the manual?

Questions and answers

Summary of Contents for Brainlab 17902

  • Page 1 WALL TOUCH SCREEN Ref-No. 17902 System and Technical User Guide Revision 1.0 Date of issue: 2023-06-28 (ISO 8601) Copyright 2023, Brainlab AG Germany. All rights reserved.
  • Page 3: Table Of Contents

    4.2.2 Test Steps ............................28 4.3 Recurrent Test ...........................29 4.3.1 Inspection Tasks..........................29 4.3.2 Measuring Protective Earth Resistance....................31 4.3.3 Measuring Equipment Leakage Current ...................32 4.4 Recurrent Test for Medical Electrical Systems ................33 System and Technical User Guide Rev. 1.0 Wall Touch Screen Ref-No. 17902...
  • Page 4 6.5 Compliances ............................47 6.5.1 Electromagnetic Emissions .......................47 6.5.2 Electromagnetic Immunity ........................48 6.5.3 RF Communications Equipment ......................50 6.5.4 Tested Cables and Accessories .......................51 6.5.5 Supported Resolutions ........................52 INDEX ..............................53 System and Technical User Guide Rev. 1.0 Wall Touch Screen Ref-No. 17902...
  • Page 5: General Information

    Feedback Despite careful review, this user guide may contain errors. Please contact us at user.guides@brainlab.com if you have improvement suggestions. Manufacturer Brainlab AG Olof-Palme-Str. 9 81829 Munich Germany System and Technical User Guide Rev. 1.0 Wall Touch Screen Ref-No. 17902...
  • Page 6: Legal Information

    Environmental Impact As a manufacturer, Brainlab complies with ISO 14001. To do so, Brainlab supports customers in minimizing the environmental impact of all Brainlab medical equipment during its expected service life and offers guidelines for proper end-of-life management.
  • Page 7 Brainlab requests suppliers to inform whether and to what extent substances from the list are used in their products. For further information regarding REACH processes at Brainlab, contact us here: www.brainlab.com/reach/...
  • Page 8: Symbols

    Reference (article) number NOTE: Indicates the Brainlab product number. Swiss authorized representative Date of manufacture NOTE: The date is expressed as in ISO 8601 as YYYY-MM-DD. Manufacturer Keep dry System and Technical User Guide Rev. 1.0 Wall Touch Screen Ref-No. 17902...
  • Page 9 U.S. federal law restricts this device to sale by or on order of a physician MR conditional NOTE: Presents no known hazards in a specific MR environment. Fragile, handle with care Do not stack This side up Recyclable System and Technical User Guide Rev. 1.0 Wall Touch Screen Ref-No. 17902...
  • Page 10 NOTE: Indicates the range of atmospheric pressure to which the medical device can be safely exposed. Tested by SGS Ground Terminals Brainlab logo DisplayPort interface USB interface IEC (C14) C14 socket Brazilian label for electrical safety China RoHs System and Technical User Guide Rev. 1.0 Wall Touch Screen Ref-No. 17902...
  • Page 11: Using The System

    Medical Technician Brainlab-trained employee Brainlab Staff Brainlab-trained employee Intended Use Environment The device is intended to be used in Professional Healthcare Facility environments as defined by IEC 60601-1- 2:2014. System and Technical User Guide Rev. 1.0 Wall Touch Screen Ref-No. 17902...
  • Page 12 0.50 mT field contour (5 Gauß volume). Clinical Benefit As standalone hardware, the device does not have any clinical benefits nor clinical performance. In combination with Brainlab software applications displayed on the device, it supports the performance of these software applications. Expected Service Life The device provides an expected service life of five years.
  • Page 13: Compatibility With Medical Devices

    Compatible Non-Brainlab Medical Devices The device can be connected to other (Medical) computing units with DisplayPort 1.2 interface and, if interaction with the touch panel is desired, a USB interface. System and Technical User Guide Rev. 1.0 Wall Touch Screen Ref-No. 17902...
  • Page 14: Training And Documentation

    • Read this guide carefully before handling the equipment • Have access to this guide at all times Available User Guides and Leaflets Available user guides vary depending upon the Brainlab product. If you have questions regarding the user guides you received, please contact Brainlab support. Guide...
  • Page 15: System Components

    SYSTEM COMPONENTS System Components General Information The system (Wall Touch Screen, 17902) is a 55" touch display that can display one or more software applications. Possible uses include (but are not limited to) data management, image viewing, planning or conferencing.
  • Page 16 Therefore, please refer to the device description of the respective navigation software application. Mounting Options There are three mounting options for the Wall Touch Screen: • On-Wall • In-Wall (flush with the wall) • In-Wall with decorative panel System and Technical User Guide Rev. 1.0 Wall Touch Screen Ref-No. 17902...
  • Page 17: System Setup

    Please note that local laws take priority over the above-mentioned requirements. In doubt, consult your local representative or the technical service department. System and Technical User Guide Rev. 1.0 Wall Touch Screen Ref-No. 17902...
  • Page 18: System Use

    The Wall Touch Screen is not in standby mode. Solid The Wall Touch Screen is in standby mode. Blinking The Wall Touch Screen is not in standby mode and searching for a signal. System and Technical User Guide Rev. 1.0 Wall Touch Screen Ref-No. 17902...
  • Page 19: Proper System Handling

    The device is not suitable for use in the presence of flammable anesthetic mixtures containing air, oxygen or nitrous oxide. Do not operate device in an environment with enriched oxygen. System and Technical User Guide Rev. 1.0 Wall Touch Screen Ref-No. 17902...
  • Page 20: Use In An Mr Environment

    Brainlab is not responsible for any changes to system behavior or the consequences thereof, if: • Any system components or the system's installation are altered by the user. • The system is used or operated differently than the descriptions provided in the Brainlab user guides.
  • Page 21: Type Plate

    Figure 3 Type Plate Location for In-Wall Mounting The type plate for the in-wall mounting option is located at the top right hand side of the frame ①. System and Technical User Guide Rev. 1.0 Wall Touch Screen Ref-No. 17902...
  • Page 22 Type Plate ① Figure 4 System and Technical User Guide Rev. 1.0 Wall Touch Screen Ref-No. 17902...
  • Page 23: Cleaning

    No Liquids Ensure that liquids do not enter the system, as this could damage the component and/or the electronics. System components are not protected against the ingress of liquids. System and Technical User Guide Rev. 1.0 Wall Touch Screen Ref-No. 17902...
  • Page 24: Cleaning The Wall Touch Screen

    If applicable, carefully clean interfaces, ensuring that no liquids enter the system. Clean and disinfect all surfaces using a compatible disinfectant (see list of Brainlab ap- proved disinfectants), following the disinfectant manufacturer’s recommendations. System and Technical User Guide Rev. 1.0 Wall Touch Screen Ref-No. 17902...
  • Page 25: Electrical Safety Inspection

    Only trained and skilled personnel (in the following named as "medical technician") are allowed to perform electrical safety tests. The test must be performed by a qualified engineer who: • Is qualified to carry out safety inspections on electrical medical equipment. System and Technical User Guide Rev. 1.0 Wall Touch Screen Ref-No. 17902...
  • Page 26 • Is up-to-date with current local regulations regarding industrial and non-industrial accident prevention. • Informs Brainlab immediately in writing if the equipment is deemed unsafe. If you require help from a Brainlab support specialist, contact Brainlab support. System and Technical User Guide Rev. 1.0 Wall Touch Screen Ref-No. 17902...
  • Page 27: Test Precautions

    Refer to the corresponding manufacturer’s instructions for the measuring device you are using. NOTE: If you are unsure whether your measuring device is suitable, contact Brainlab support. Test Guidelines When performing recurrent tests: •...
  • Page 28: Test Steps

    Test each piece of equipment of the ME SYSTEM individually, according to steps 1-3 for ME EQUIPMENT. Report and evaluate these results. Touch current Functional test Document results Evaluate results Check and prepare for normal use. System and Technical User Guide Rev. 1.0 Wall Touch Screen Ref-No. 17902...
  • Page 29: Recurrent Test

    Enter the maximum measured value in the table below: Test Step Normal Condition Passed? Protective earth resistance Equipment leakage current Serial number of measuring device: ________________________ Calibration valid through (date): ________________________ Test performed (date): _________________by: __________________________________ System and Technical User Guide Rev. 1.0 Wall Touch Screen Ref-No. 17902...
  • Page 30 Recurrent Test Related Links Malfunctions on page 40 System and Technical User Guide Rev. 1.0 Wall Touch Screen Ref-No. 17902...
  • Page 31: Measuring Protective Earth Resistance

    Connect the measurement tip of your measuring device with the conductive parts shown in the table below and measure the resistance. If a test does not pass, device must be repaired by Brainlab support. After repair, repeat the entire electrical safety test from the beginning.
  • Page 32: Measuring Equipment Leakage Current

    • All devices powered with mains power • Only measuring points without PE connec- tion Equipment Leakage Current Measurement Points Point Tested Measured Value Test Passed? Housing Wall Display System and Technical User Guide Rev. 1.0 Wall Touch Screen Ref-No. 17902...
  • Page 33: Recurrent Test For Medical Electrical Systems

    Enter the maximum measured value at each required measurement point in the table below: Test Step Normal Condi- Single Fault Passed? tion Condition Touch current Serial number of measuring device: ________________________ Calibration valid through (date): ________________________ Test performed (date): _________________by: __________________________________ System and Technical User Guide Rev. 1.0 Wall Touch Screen Ref-No. 17902...
  • Page 34: Measuring Touch Current

    • Fully assembled with all required PE and da- ta connections for intended use • All mains power switches in operating posi- ≤ 0.5 mA tion (On) • All devices powered with mains power System and Technical User Guide Rev. 1.0 Wall Touch Screen Ref-No. 17902...
  • Page 35: Maintenance

    • Perform a visual inspection as described on page 37. • Perform an electrical safety test as described on page 25. If any damage is detected, place the device out of operation, mark as such and contact Brainlab support. System and Technical User Guide Rev. 1.0 Wall Touch Screen Ref-No. 17902...
  • Page 36 • Mechanical safety tests as described on page 37. If any damages are detected, place the device out of operation, mark as such and contact Brainlab support. Related Links Tested Cables and Accessories on page 51 System and Technical User Guide Rev. 1.0 Wall Touch Screen Ref-No. 17902...
  • Page 37: Maintenance Procedures

    Connect a viable image source and touch host (e.g., computer) and power on the devices. • Check that screen shows an image without obvious issues • Check that touchscreen interaction is accurate If any damage is detected, place the device out of operation, mark as such and contact Brainlab support. Mechanical Safety Tests...
  • Page 38 • Sealing material, depending on condition of existing sealing. (Can be ordered as spare part. Contact Brainlab for further information). Proceed as follows to close the Wall Touch Screen after is has been put into service position (flipped down 90°): System and Technical User Guide Rev. 1.0 Wall Touch Screen Ref-No. 17902...
  • Page 39 The sealing has the same rectangular shape and dimensions as the display and should be installed accordingly. Warning Be careful not to squeeze fingers or cables when closing the display. System and Technical User Guide Rev. 1.0 Wall Touch Screen Ref-No. 17902...
  • Page 40: Malfunctions

    Malfunctions Malfunctions System Damage or Failure If any of the following occur, do not continue to use the system and contact Brainlab support: • The mains power cable or plug is damaged. • Liquid has been spilled into the device.
  • Page 41: Return Instructions

    • Provides you with a cost estimate for repair or replacement • Informs you when your system is expected to be operational again NOTE: Do not return the device without the RMA number. This is provided to you by Brainlab when you contact support for repair or replacement.
  • Page 42 Return Instructions System and Technical User Guide Rev. 1.0 Wall Touch Screen Ref-No. 17902...
  • Page 43: Compliances And Specifications

    Certificate/Approval IEC 60601-1:2005+A1:2012+A2:2020 (Edition 3.2) IEC 60601-1-2:2014+AMD1:2020 (Edition 4.1) NRTL: Certificates • AAMI ANSI IEC 60601-1-2:2007/(R)2012 • CSA C22.2#60601-1 Issued: 2008/02/01 Ed:2 • CE, RoHS, REACH, POP, WEEE System and Technical User Guide Rev. 1.0 Wall Touch Screen Ref-No. 17902...
  • Page 44: Physical Specifications

    Disconnect it from the mains power supply and contact Brainlab support. The device in its various variants complies with the requirements of the IP classification standard IEC 60529: On-Wall IP20 In-Wall IP44 System and Technical User Guide Rev. 1.0 Wall Touch Screen Ref-No. 17902...
  • Page 45: Technical Specifications

    8 bit per channel Number of colors 16.7 million colors Lifetime backlight 50,000 hours Pixel class ISO 9241-307 (pixel error class II) Touch technology Projected capacitive, multi-touch Glass Anti-reflective coating System and Technical User Guide Rev. 1.0 Wall Touch Screen Ref-No. 17902...
  • Page 46: Environmental Requirements

    If the device has been stored in conditions outside the range of the operating conditions, it requires an adaption time for use of a minimum of 24 hours. System and Technical User Guide Rev. 1.0 Wall Touch Screen Ref-No. 17902...
  • Page 47: Compliances

    If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally. System and Technical User Guide Rev. 1.0 Wall Touch Screen Ref-No. 17902...
  • Page 48: Electromagnetic Immunity

    1 kHz Dwell Time = 30 second 100V 60Hz 120V 60Hz 11 Power frequency magnetic field EN 61000-4-8 • 30 A/m (37.7 µT) at 50 Hz and 60 Hz 220V 60Hz 240V 50Hz System and Technical User Guide Rev. 1.0 Wall Touch Screen Ref-No. 17902...
  • Page 49 If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the Wall Touch Screen system. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. System and Technical User Guide Rev. 1.0 Wall Touch Screen Ref-No. 17902...
  • Page 50: Rf Communications Equipment

    433 MHz 3 V/m > 60 s ISO/IEC 18000-63 Type C 860 - 960 MHz 54 V/m > 10 s ISO/IEC 18000-4 Mode 1 2.45 GHz 54 V/m System and Technical User Guide Rev. 1.0 Wall Touch Screen Ref-No. 17902...
  • Page 51: Tested Cables And Accessories

    Malaysia, Hong Kong, China, Japan, Israel, Denmark, Switzerland, Brazil, Argentina, Tai- wan, Italy PureLink DisplayPort cable PI5000-050, 5 me- DisplayPort USB 2.0 AM USB BM Space Shuttle Hi-Tech, 1.5 meter System and Technical User Guide Rev. 1.0 Wall Touch Screen Ref-No. 17902...
  • Page 52: Supported Resolutions

    • 800 x 600 • 720 x 400 • 640 x 480 Signals outside of the specified resolutions may not be supported. For more information on input resolution contact Brainlab support. System and Technical User Guide Rev. 1.0 Wall Touch Screen Ref-No. 17902...
  • Page 53: Index

    27 registration, evaluation and authorization of chemicals REACH..................7 resolutions.................. 52 returning equipment addresses................41 RF communications equipment..........50 safe disposal................6 safety test................... 25 support numbers................5 System and Technical User Guide Rev. 1.0 Wall Touch Screen Ref-No. 17902...
  • Page 56 Art-No. 60922-11EN *60922-11EN*...