Maxim RKI GX-6000 Quick Start Manual

Advertisement

Maxim Quick Start Guide
RKI GX-6000
(PID)
The RKI GX-6000 is a PID designed to screen
air for total VOCs in the ppm or ppb range.

Normal Operation:

Press and hold the "Power" button, allow the
instrument to warm-up for approx. 10
minutes.

Fresh Air Calibration

Move to a clean air location, attach the zero filter to the end of
the inlet. Press and hold "Air" button, release when prompted,
values should be approx. zero.

Enable Data-Logging:

With the instrument powered off.
Press and hold the "Air" and "Shift" Buttons", while holding
press the "Power" button, when unit beeps let go of all three
buttons. Select "Maintenance" and enter the password as
"0006"
Scroll down and select "Log Setting", next "Log Interval"
Enter preferred log interval, select enter and then escape
On the next screen, scroll and select "Start Measure", this enters
the unit back into normal mode.
Please note the instrument will data-log automatically when
powered on based on your defined interval.
** Please note that the USB-Infrared cable is required for data
downloading to PC software**
www.maximenvironmental.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the RKI GX-6000 and is the answer not in the manual?

Questions and answers

Summary of Contents for Maxim RKI GX-6000

  • Page 1 Maxim Quick Start Guide RKI GX-6000 (PID) The RKI GX-6000 is a PID designed to screen air for total VOCs in the ppm or ppb range. Normal Operation: Press and hold the “Power” button, allow the instrument to warm-up for approx. 10 minutes.
  • Page 2 Perform Calibrations To perform a calibration you will need the calibration gases (10 ppm Isobutylene for ppb mode, 100 ppm Isobutylene for ppm mode), a demand flow regulator and some tubing. While in normal operation, press and hold the “shift” but- •...

Table of Contents