Cleaning; Safety Switch; Calibration - VERDER CARBOLITE GERO RHF 14/8 Installation, Operation And Maintenance Instructions

1400 c rapid heating furnace
Table of Contents

Advertisement

5.2.1

Cleaning

Soot deposits may form inside the furnace, depending on the process. At appropriate
intervals remove these by heating as indicated in the General Operation Notes.
The product's outer surface may be cleaned with a damp cloth. Do
not allow water to enter the interior of the case or chamber. Do not
clean with organic solvents.
5.3

Safety Switch

When correctly functioning, the safety switch will isolate all live conductors (live and
neutral connections) within the heating element circuit(s) when the product door is
opened. The safety switch should be checked regularly to ensure that this occurs.
The safety switch should not fail under normal working conditions, however rough
handling, exposure to corrosive materials/ environments, or exceptionally frequent use,
could compromise the safety system.
Weekly check:
The following check can be carried out by a general operator:
On the temperature controller, set a safe temperature above ambient. The heater
l
lights should illuminate.
Open the door and check the heater lights. They should no longer be illuminated.
l
If the heater lights remain illuminated when the door is open, discontinue use and
contact Carbolite Gero Service.
Annual check:
The following checks should be carried out by a qualified electrician, as specified in the
"Maintenance Schedule" section of this manual:
Remove the element access panel and take a voltage measurement from the heat-
l
ing element terminals. Do not attempt to take a reading from the heating element
itself as surface oxidation will give an unreliable contact.
Ensure that power to the heating elements is switched off when the door is opened.
l
Contact Carbolite Gero Service and discontinue use of the product if it is found that the
heating elements are not fully isolated during these checks.
5.4

Calibration

After prolonged use, the controller and/or thermocouple may require recalibration. This
is important for processes that require accurate temperature readings or for those that
16