Thanks so much for choosing our product. This high-tech infrared thermometer is used to measure the surface temperature of the forehead and convert it to the actual human temperature by an algorithm (adjustment mode). It can help you learn about the health of you and your family quickly at any time and anywhere.
Product name: Infrared thermometer
Product model: FDIR-V12
LCD display description
The thermometer mainly consists of plastic shell, infrared temperature sensor, PCBA, buzzer chip, LCD display and battery. Product front and back view shows below:
Model | FDIR-V12 |
Power supply | DC 3V,2*1.5V AAA batteries |
Measuring range | Measurement Range for Body: 32.0°C~42.9°C(89.6°F~109.2°F) Measurement Range for Object Surface: 0°C~60.0°C(32.0°F~140.0°F) |
Measuring accuracy | Body Mode: ±0.2°C /0.4°F,if temperature at 35.0°C ~ 42.0°C ±0.3°C/0.5°F,if temperature lower than 35.0°C or higher than 42.0°C Object Mode: ±1°C |
Memory function | 19 groups of data memory |
Display resolution | 0.1°C/0.1°F |
Clinical repeatability | ±0.3°C/0.5°F |
Measuring position | Forehead |
Operation condition temperature | Temperature: 50.0°F~104.0°F (10.0°C~40.0°C) Relative humidity: ≤ 95%RH Atmospheric pressure: 70 kPa ~ 106 kPa |
Transport/Storage condition temperature | Temperature: -13.0°F~131.0°F (-25.0°C~55.0°C) Humidity: ≤ 95%RH,non-condensing Atmospheric pressure: 70 kPa~106 kPa |
Grade of waterproof | IP22 |
Electric shock | Internally powered ME equipment |
Applied part | Type BF applied part, including the whole unit |
Battery life | 2 years/1000 measurements |
Product size | 145mm*41mm*49mm |
Product weight | 93g |
Service life | 5 years |
Warranty | 2 years |
Software version | V1.0 |
This product mainly adopts infrared temperature measurement method to measure the temperature of forehead and can be used for newborns, children and adults. We recommend adults to operate the thermometer instead of babies and children.
Measuring inflammation, trauma or postoperative lesions.
Put two AAA-batteries into the battery house on the back of the device. At the point the product will start self-inspection, and then turn into power off automatically.(if the battery power is low when starting up, please replace the battery to ensure adequate power supply)
If infrared thermometer beeps once, it means everything goes well. If the measurement result is 37.6°C or above, the infrared thermometer will beep four times. Measurement readings will be displayed on screen.
Measurement range for body | Buzzer warning | Color of backlight | ||
32.0≤ T ≤ 37.5(99.6°C°F(89.6) °C °F) | The buzzer warning by sending out a long "beep" | Green | ||
37.6≤ T ≤ 42.9(109.2°C°F(99.7) °C °F) | The buzzer warning by sending out a "beep" "beep" "beep" "beep"..."beep" "beep" "beep" "beep"... "beep" "beep" "beep" "beep" | Red | ||
T | The buzzer warning by sending out a "beep" "beep" "beep" "beep" | Green | ||
T > 42.9(109.2Displaying Hi° F)° C | The buzzer warning by sending out a "beep" "beep" "beep" "beep" | Green | ||
Measurement range for object | Buzzer warning | Color of backlight | ||
0°C(32.0°F) ≤ T ≤ 60.0°C (140.0°F) | The buzzer warning by sending out a long "beep" | Green | ||
T | The buzzer warning by sending out a "beep" "beep" "beep" "beep" | Green | ||
T > 60.0(140.0Displaying Hi° F)° C | The buzzer warning by sending out a "beep" "beep" "beep" "beep" | Green |
After power on, press [Memory/Voice Button] for about 2 seconds to turn on or off the mute mode.
Once power on, object mode and body mode can be switched by pressing [Mode Switch Button] slightly.
If the device is not in use, it will Power off automatically in 30 seconds.
The thermometer will detect the power automatically after boot-up. If the power is low but still sufficient enough for usage, the low power symbol will appear on the screen along with the measurement result. However, when the battery is running too low, the low power symbol will be flashing on the screen, and after 8 seconds, it will automatically shutdown. You must replace new batteries to continue usage.
Please follow the related national laws of disposing the abandoned batteries.
Please do not throw the batteries directly to the garbage can. Please take out the batteries if the device is not used for long periods of time.
Please do not put the batteries in the fire.
The temperature of human body belongs in a range and the range may vary among people. Temperature of individuals may also vary from time to time. We recommend that you know your normal temperature measured on the forehead, so that you therefore have a starting point for any temperature differences.
Self-diagnosis and self-treatment based on the measurement results could be dangerous, please contact the doctor for advisory and provide him/her the measurement results for reference.
The thermometer has already been calibrated at the time of manufacture. We suggest to change into new devices after 2 years from purchase, or seek assistance for calibration from professional organizations before usage. If the thermometer is used as prescribed, no further calibration will be necessary. Please consult Seagull Healthcare's quality department for technical questions about measurement results.
Problem or error message | Checklists | Solution |
No response | The batteries are dead? | Replace batteries with new ones. |
The batteries type are wrong or reversed polarity? | Remove the batteries and adjust them correctly. | |
Poor battery connection? | ||
![]() | Temperature is measured beyond temperature range from 32.0 ℃ ~ 42.9 ℃ (89.6 ℉ ~ 109.2 ℉ ). Hi: means that the temperature measured is too high Lo: means that the temperature measured is too low | Refer to the user manual and measure the temperature again. |
![]() | The ambient temperature is beyond the operating range of 50 ℉ ~104 ℉(10℃ ~40℃) | Keep the thermometer in the environment where the temperature is between 50 ℉ ~104 ℉(10℃ ~40℃ for 30 )minutes. |
![]() | Battery is running low | Please replace the battery in time not to disturb usage. |
![]() | The battery is almost dead and can not be used anymore. | Must replace new battery to continue usage. |
![]() | Hardware damage | Please contact distributor for support |
We only use original accessories. Check the accessory list as below to ensure if the package delivered is complete.
Quantity | Parts |
1pcs | FDIR-V12 device |
1pcs | User Manual |
Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.