Oxygen Saturation is a percentage of Oxyhemoglobin (HbO2) capacity, compounded with oxygen, by all combinative hemoglobin (Hb) capacity in blood. In other words, it is consistency of Oxyhemoglobin in blood. It is a very important parameter for the Respiratory Circulation System. Many respiratory diseases can result in oxygen saturation being lowered in human blood. Additionally, the following factors can reduce oxygen saturation: Automatic regulation of organ dysfunction caused by Anesthesia, Intensive Postoperative Trauma, injuries caused by some medical examinations. That situation might result in light-headedness, asthenia, and vomiting. Therefore, it is very important to know the oxygen saturation of a user so that doctors can find problems in a timely manner. The fingertip pulse Oximeter features small size, low power consumption, convenient operation and portability. It is only necessary for a user to put one of his fingers into the fingertip photoelectric sensor for diagnosis, and a display screen will show oxygen saturation. It has been proven in clinical experiments that it also features high precision and repeatability.
Principle of the Oximeter is as follows: A mathematical formula is established making use of Lambert Beer Law according to Spectrum Absorption Characteristics of Reductive hemoglobin (RHb) and Oxyhemoglobin (HbO2) in glow and near-infrared zones. Operation principle of the instrument: Photoelectric Oxyhemoglobin Inspection Technology is adopted in accordance with Capacity Pulse Scanning and Recording Technology, so that two beams of different wavelength of lights (660nm glow and 940nm near infrared light) can be focused onto a human nail tip through a clamping finger-type sensor. A measured signal obtained by a photosensitive element, will be shown on the Oximeter's display through process in electronic circuits and microprocessor shown on the Oximeter's display through electronic circuits and a microprocessor.
Diagram of Operation Principle
Follow local ordinances and recycling instructions regarding disposal or recycling of the device and device components, including batteries.
Fingertip PULSE OXIMETER is a portable non-invasive, spot-check, oxygen saturation of arterial hemoglobin (SPO2) and pulse rate of adult and pediatric user at home, and hospital (including clinical use in internist/surgery, Anesthesia, intensive care etc). It is not for continuously monitoring.
Notes:
The Pulse bar graph displays corresponding with the user's pulse beat. The height of the bar graph shows the user's pulse strength.
Notes:
EMC of this product complies with IEC60601-1-2 standard.
The materials which the user can come into contact have no toxicity and no action on tissues comply with ISO10993-1, ISO10993-5 and ISO10993-10.
Electromagnetic emissions-for all EQUIPMENT and SYSTEMS
Guidance and manufacturer's declaration – electromagnetic emission | ||||||
The Pulse Oximeter is intended for use in the electromagnetic environment specified below. The customer or the user of the Pulse Oximeter should assure that it is used in such an environment. | ||||||
Emission test | Compliance | Electromagnetic environment – guidance | ||||
RF emissions CISPR 11 | Group 1 | The Pulse Oximeter uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment. | ||||
RF emission CISPR 11 | Class B | The Pulse Oximeter is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply. network that supplies buildings used for domestic purposes. |
Problems | Possible reason | Solution |
SPO2 or PR can not be shown normally |
|
|
SPO2 or PR is shown unstably |
|
|
The Oximeter can not be powered on |
|
|
The screen are suddenly off |
|
|
There are no user-serviceable parts inside the oximeter. The cover should only be removed by qualified service personnel.
If you are uncertain about the accuracy of any measurement, check the user's vital signs by alternate means; then make sure the oximeter is functioning correctly.
Do not spray, pour, or spill any liquid on the oximeter, its accessories, connectors, switches, or openings in the enclosure as this may damage the oximeter.
Symbol | Definition |
![]() | CE Mark: The Product system conforms to essential requirements of the Medical Device Directive 93/42/EEC. |
![]() | Type BF equipment (Refer to IEC 60601-1:1995) |
| Attention, consult accompanying documents. |
﹪SPO2 | Oxygen saturation |
![]() | Pulse rate |
![]() | Symbol for the marking of electrical and electronics devices according to Directive 200296/EC. The device, accessories and the packaging have to be disposed of waste correctly at the end of the usage. Please follow Local Ordinances or Regulations for disposal. Note: The Oximeter is applied to this regulation. |
Note: The illustration used in this manual may differ slightly from the appearance of the actual product.
www.FaceLake.com
37 Sherwood Terrace, Unit 124,
Lake Bluff, IL 60044 USA
Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.