  • Page 2 Thank you very much for purchasing your new ULTRA X moisture tester. Please read the following instructions carefully. [Geben Sie ein Zitat aus dem Dokument oder die Zusammenfassung eines interessanten Punktes ein. Sie können das Textfeld an einer beliebigen Stelle im Dokument positionieren. Verwenden Sie die Registerkarte 'Textfeldtools', wenn Sie das Format des Textfelds 'Textzitat' ändern möchten.] ULTRA X instruments normally work for many years.
  • Page 3 2. First measurement with the factory settings There are only three steps to do: Tare the balance to zero with the Tare button. Place your sample material on the drying pan and spread it out to an equal surface for a good drying Wait some seconds until the weight is shown steady in the display.
  • Page 4 Konsta-Stop (automatic mode) or Timer Display shows 0,000 g press SET 2 seconds 12,3 min. press „change“ blinks 12,3 set time with button „change“, confirm with SET 10,3 blinks set time with button „change“, confirm with SET 10,0 blinks press SET Timer set to 00,0 sets the instruments into the automatic mode.
  • Page 5: Settings Level

    4. Parameter setting Different parameters can be adjusted by using the Tare together with the Change button. This is only possible with the radiator in the backwards position and the display showing the weight. Push the Tare button very short will tare the balance to zero, push it some seconds will start the parameter setting mode.
  • Page 6 Calibrating the balance  1 (very calm), 2, 3, Stability of the balance 4 (turbulent)  Uhr, Set time and date Datum, Monat, Jahr  Config Settings can be printed by the „Change“button  0,000 g Standard Settings are underlined and bold Meaning of the results: „FEUCHTE“...
  • Page 7 Please note that the standard settings will be correct for the most materials with sample sizes of approx. 10 g, which is our recommendation as a good sample weight. ) “PFl” is the “PreFlash” feature, which is useful for very wet samples to shorten the measuring time.
  • Page 8 6. Calibration of the balance Display shows Press the „change“ button, after a few seconds the value of the calibration weight appears in the display. Place the calibration weight on the middle of the scale pan. Display disappears. After a few seconds appears --- 200,000 g 7.
  • Page 9 9. Failures shown in the display ERR 2 No sample on the pan ERR 3 Overflow in ATRO range ERR 4 No pan on the balance, pan holder touches housing ERR 5 Overflow sample weight ERR 10 - ERR 16 Place is not calm, pan holder is bent, dirt around pan holder others...
  • Page 10 DESCRIPTION OF INTERFACE FOR ULTRA X SERIES 20 . 1 ( U X 2011 / U X 2031 / U X 2081 ) 1. Interface type: RS 232 / V.24 unidirectional 2. Transmission speed: 1200 Baud 3. Interface operation: Asynchronous 4.