Amplifier Instruction
Range:
Type:
Supply:
Output:
Revision: V1.0
Manual
LETIX
L-ANV-0-0-R-AC
110-240V AC
0-10V

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the L-ANV-0-0-R-AC and is the answer not in the manual?

Questions and answers

Summary of Contents for Absolute Measurement L-ANV-0-0-R-AC

  • Page 1 Amplifier Instruction Manual Range: LETIX Type: L-ANV-0-0-R-AC Supply: 110-240V AC Output: 0-10V Revision: V1.0...
  • Page 2: Table Of Contents

    Digital Display Not Illuminated When Indicator Is Switched On........... 9 Verify The Transducer Connections ....................9 Check The Display Is Reading Correctly When The Transducers Are Loaded ....... 9 _____________________________________________________________________________________________________________________ Absolute Measurement Ltd. (T/A CMC Controls) t: (0) 1732 763278 | e: sales@absolutemeasurement.co.uk | w: absolutemeasurement.co.uk...
  • Page 3: General Description

    The Indicator can accept full scale input voltages of either up to +/- 40 mV (foil strain gauges) or up to +/- 400 mV (semi-conductor strain gauges) for full display with selectable decimal point position and scaling, and a zero offset for both positive and negative sensing. _____________________________________________________________________________________________________________________ Absolute Measurement Ltd. (T/A CMC Controls) t: (0) 1732 763278 | e: sales@absolutemeasurement.co.uk | w: absolutemeasurement.co.uk...
  • Page 4: Installation

    10 11 12 13 EXC- SEN- SIG- SIG+ SEN+ EXC+ Used _____________________________________________________________________________________________________________________ Absolute Measurement Ltd. (T/A CMC Controls) t: (0) 1732 763278 | e: sales@absolutemeasurement.co.uk | w: absolutemeasurement.co.uk Unit 19 | Silver Birches Business Park | Aston Road | Bromsgrove | B60 3EU...
  • Page 5: Transducer Electrical Connection Diagrams

    SIG- SIG+ EXC+ PLEASE NOTE Connect the cable screens (green/yellow) to clean earth close to the LETIX Tension Indicator. _____________________________________________________________________________________________________________________ Absolute Measurement Ltd. (T/A CMC Controls) t: (0) 1732 763278 | e: sales@absolutemeasurement.co.uk | w: absolutemeasurement.co.uk Unit 19 | Silver Birches Business Park | Aston Road | Bromsgrove | B60 3EU...
  • Page 6 WIRING DIAGRAM For 2 ULTRA Transducer EXC- SIG- SIG+ EXC+ _____________________________________________________________________________________________________________________ Absolute Measurement Ltd. (T/A CMC Controls) t: (0) 1732 763278 | e: sales@absolutemeasurement.co.uk | w: absolutemeasurement.co.uk Unit 19 | Silver Birches Business Park | Aston Road | Bromsgrove | B60 3EU...
  • Page 7: Supply Connection Diagram

    LETIX Type: L-ANV-0-0-R-AC Revision: V1.0 Supply Connection Diagram _____________________________________________________________________________________________________________________ Absolute Measurement Ltd. (T/A CMC Controls) t: (0) 1732 763278 | e: sales@absolutemeasurement.co.uk | w: absolutemeasurement.co.uk Unit 19 | Silver Birches Business Park | Aston Road | Bromsgrove | B60 3EU...
  • Page 8: Supply Voltage

    IMPORTANT Before calibration, ensure that the dipswitches at the rear of the unit are set to the UP position. After calibration, set them to the DOWN position. This will stop the machine operator from altering the settings and prevent re-calibration of the indicator. _____________________________________________________________________________________________________________________ Absolute Measurement Ltd. (T/A CMC Controls) t: (0) 1732 763278 | e: sales@absolutemeasurement.co.uk | w: absolutemeasurement.co.uk...
  • Page 9: Direct Calibration - Full Scale Setting

    IMPORTANT Once this has been done, set the dipswitches at the rear of the unit to the DOWN position. This will stop the machine operator from altering the settings and prevent re-calibration of the indicator. _____________________________________________________________________________________________________________________ Absolute Measurement Ltd. (T/A CMC Controls) t: (0) 1732 763278 | e: sales@absolutemeasurement.co.uk...
  • Page 10: Troubleshooting

    Indicator repaired (or credited if found faulty under warranty). PLEASE NOTE This unit CANNOT be repaired in the field, and must be returned to CMC Controls. _____________________________________________________________________________________________________________________ Absolute Measurement Ltd. (T/A CMC Controls) t: (0) 1732 763278 | e: sales@absolutemeasurement.co.uk | w: absolutemeasurement.co.uk...