DUAL CHAMBER
EXTERNAL PACEMAKER
Shree Pacetronix Ltd
Plot no. 15, sector II,
Pithampur, Dist.: Dhar
Madhya Pradesh, India
Email:
pacetronix@hotmail.com
Website: www.pacetronix.com

Advertisement

Table of Contents
loading

Summary of Contents for Pacetronix ZEUS-1

  • Page 1 DUAL CHAMBER EXTERNAL PACEMAKER Shree Pacetronix Ltd Plot no. 15, sector II, Pithampur, Dist.: Dhar Madhya Pradesh, India Email: pacetronix@hotmail.com Website: www.pacetronix.com...
  • Page 2 Revision Date Change Description January 15 , 2019 August 21 , 2020 User Information October 20 , 2021 Specification & Sticker Document: Man_Zeus-1, Rev 03 / 20102021...
  • Page 3 Explanation of Symbols : Symbol Symbol Explanation Operating instructions Type CF applied part. Do not dispose of this product in the unsorted municipal waste stream. Dispose of this product according to local regulations Package contents Product documentation Temperature limitation Serial number Manufacturer Date of manufacture DEFIBRILLATION-PROOF TYPE CF APPLIED PART.
  • Page 4: Table Of Contents

    Contents 1. GENERAL DESCRIPTION ....................6 2. PACKAGE CONTENTS ...................... 8 3. SAFETY FEATURES ......................8 4. SPECIFICATION........................8 Technical Data .................... 9 Mechanical Data ..................11 Urgent Values .................... 11 5. FUNCTION AND DESCRIPTION..................12 Functional Indicators ................12 5.1.1 Pacing ........................
  • Page 5 7. MODE DEFINITIONS ....................... 20 8. DEVICE MAINTENANCE ....................23 8.1. Cleaning..................... 23 8.2. Safety & Technical checks ................. 23 WARNINGS ......................... 23 PRECAUTIONS ......................26 FURTHER INFORMATION ..................29 LIMITED ONE YEAR WARRANTY ................30 -------------------------------------------------------------------------------------------------------...
  • Page 6: General Description

    The Zeus-1 Dual chamber external pacemaker offers short-term pacing to the patient with myocardial infarction or heart block. Note: In rest of the document Zeus-1 Dual chamber external pacemaker will be referred as Zeus-1. In designing these pacemakers, the emphasis has been on producing a rugged battery powered pacemaker that is small and light enough to be comfortably attached to a patient, yet possesses all the attributes of reliable bedside units.
  • Page 7 ZEUS -1...
  • Page 8: Package Contents

    • Minimized susceptibility to electromagnetic and magnetic Interference. SPECIFICATION _________________________________________________________________ The Dual chamber external pacemaker Zeus-1 is the result of longtime experience in development and production of cardiac pacemakers. Zeus-1 provides safe and effective heart stimulation indicated by rhythm and conduction defects, termination of bradycardias at the time of implant or exchange of a pacemaker.
  • Page 9: Technical Data

    Stimulation is indicated by synchronous flashing of ‘PACE’ & Sensing is indicated by synchronous flashing of ‘SENS’. Mains independence battery operation provides a maximum on electric safety for both patient and user. 4.1 Technical Data PARAMETERS VALUES Technical Parameters Description of Dual chamber External Pacemaker Pacemaker Rate...
  • Page 10 Flashing GREEN LED’s on the front panel for both Atrial & Ventricle. Replace battery as Low battery Indicator (RED LED) glows. Low Battery Indicator Battery Replacement time approx 15 Sec. Buzzer Alarm For lead integrity...
  • Page 11: Mechanical Data

    After 20 pacing/Sensing cycles device locked automatically. Lock / Unlock Press “Pause” switch to ‘Unlock’ Safety Cover Acrylic transparent cover Pause Button When pressed output will be stopped for at the most 10 sec. ON/OFF Button Separate ON and OFF button to start and stop the device. Lead Temporary lead 4.2 Mechanical Data...
  • Page 12: Function And Description

    FUNCTION AND DESCRIPTION _________________________________________________________________ Functional Indicators 5.1.1 Pacing Pacing Indication : Flashing LED’s on the front panel for both A & V. Note : These LED’s indicate delivery of a pacing pulse but do not necessarily indicate that the pacing pulse has initiated cardiac stimulation.
  • Page 13: Avi (Msec) (A-V Interval)

    Description of Rate: The pacing rate refers to the average number of pacemaker’s output pulses occurring over a specific period of time, usually one minute. Rate (ppm) = 60000/ Interval (ms) The “60,000” Rule 60,000 divided by the ‘RATE’ equals the ‘INTERVAL’ 60,000 divided by the ‘INTERVAL’...
  • Page 14: A' Sens (Mv) (Atrium Sensitivity)

    5.2.4 ‘A’ Sens (mV) (Atrium Sensitivity) This knob is used to set Atrium sensitivity (mV) values. It allows continuous adjustment of the Sensitivity from 0.4 to 10 mV. Description of Atrium Sensitivity: Sensitivity is the ability of the pacemaker to see intrinsic cardiac events (P Wave). 5.2.5 Mode This knob is used to set desired Mode.
  • Page 15: V' Output (Volts) (Ventricle Pulse Amplitude)

    Description of Blanking period :Blanking is the period following a paced Atrial event for which Ventricle sense amplifier will be kept ‘OFF’. 5.2.7 ‘V’ Output (Volts) (Ventricle Pulse Amplitude) This knob is used to set Ventricle pulse Amplitude in volts values.
  • Page 16: Operating Instructions

    blood pressure, and ensure defibrillation equipment is immediately available. OPERATING INSTRUCTIONS _______________________________________________________________ 6.1 RAP Button Whenever Rapid Atrial Rate is required, press the RAP button and adjust the rate knob for desired Atrial rate. Caution: Only for Atrial 6.2 ON Switch Switches the unit ON by pressing ON button.
  • Page 17: Off Switch

    6.3 OFF Switch Switch the unit OFF by pressing OFF switch twice in 4 sec. Press and held switch for 1 sec and then release and press again. If it doesn’t gets OFF then repeat the process. 6.4 Pause / Unlock Button Pause button is used to stop output of the unit for both chambers.
  • Page 18: Demand And Asynchronous Modes

    ECG should be monitored in order that the S-T segment elevation, indicating contact with the myocardium and impaction, can be noted. b). Connecting the lead to the ZEUS-1 Switch OFF the device (Zeus-1). Connect the proximal electrode to the indifferent (+) red terminal and...
  • Page 19: Battery Replacement

    External pacemaker (Zeus-1) as described in the method below. Method : Twist the terminal caps counter clockwise, push the lead electrodes into the end of the collets terminals, tighten the leads by twisting caps clockwise. Caution : The terminal caps should not be over tightened otherwise the terminal may break.
  • Page 20: Mode Definitions

    MODE DEFINITIONS Single Chamber Modes The Model Zeus-1 will pace in four single chamber modes: AOO, VOO, AAI, and VVI. AOO and VOO Modes The asynchronous mode (A00 and VOO) emit a pacing pulse into one chamber, either the atrium (AOO) or the ventricle (VOO),at the end of the base rate interval, regardless of any intrinsic activity.
  • Page 21 RATE in the absence of detected ventricular activity. The atria channel is continuously blanked. Dual Chamber Modes The Model Zeus-1 will pace in four dual-chamber modes: DOO, DDD, DVI & VDD.
  • Page 22 DOO Mode When pacing in the A-V sequential asynchronous mode (DOO), the pacemaker will pace the atrium and the ventricle asynchronously at the programmed base rate and with the programmed A-V interval. Both channels are continuously blanked. When the V-A interval times out, the pacemaker will: •...
  • Page 23: Device Maintenance

    8.2. Safety & Technical checks Safety and technical checks should be carried out on the Model Zeus-1 at least once every 6 months and after any malfunction or accident. Pacetronix does not recommend field repair of the device. For service or repair contact your local Pacetronix representative at the appropriate address.
  • Page 24 They are inherently sensitive to some external fields. In the presence of excessive levels of interference the Model Zeus-1 may inhibit completely or revert to asynchronous operation, pacing at the rate set by the RATE knob.
  • Page 25 Defibrillation discharges of 360 watt-seconds have not affected the Model Zeus-1 in laboratory tests. However, for maximum safety it is recommended that the paddles not be placed near the Model Zeus-1 or the lead system. Whenever possible, for the safety of the patient, disconnect the pacemaker from the lead system before defibrillating or cardioverting.
  • Page 26: Precautions

    Random Failures The physician should be aware that the Model Zeus-1 Temporary Pacemaker can fail due to a number of reasons, such as random component failure, battery depletion, and mishandling.
  • Page 27 • No output • No sensing • False indicator light signals • Increased or decreased rate, output pulse width, or output amplitude • Reversion to asynchronous pacing and Loss of control of rate, output, sensitivity or power. If loss of control of rate, output, sensitivity or power occurs, and it would be appropriate to temporarily stop pacing the patient, attempt to correct the condition by turning the device off and then on.
  • Page 28 Unauthorized Changes of Pacemaker Settings Do not place the Model Zeus-1 in any area where patients may interact with it. The temporary pacemaker should be placed in an area that minimizes tampering with the device by unauthorized personnel (patients, visitors, etc.).
  • Page 29: Further Information

    FURTHER INFORMATION _________________________________________________________________ For any further information please contact : Shree Pacetronix Ltd. Plot No.15, Sector II, Pithampur, Dist. Dhar (M.P.) 454775 Fax: 07292 400418 Tel: 07292 411105 Email: pacetronix@hotmail.com Website: www.pacetronix.com Notes: _______________________________________________________________ _______________________________________________________________ _______________________________________________________________ _______________________________________________________________...
  • Page 30: Limited One Year Warranty

    The remedies set forth herein shall be the exclusive remedies available arising out of the sale or use of the ‘ZEUS-1’ Dual chamber external pacemaker. No person has any authority to bind Shree Pacetronix Ltd. to any representation or warranty except as set forth herein.