Norav Medical NR-1207-E Installation And Operation Manual

Norav Medical NR-1207-E Installation And Operation Manual

Nr recorder
Hide thumbs Also See for NR-1207-E:

Advertisement

Quick Links

NR Recorder
Model NR-1207-E
Installation and Operation Guide
(for iOS Devices)
Doc P/N: 2018-005|Norav Medical
Revision 2.0
Updated 16/12/19

Advertisement

Table of Contents
loading

Summary of Contents for Norav Medical NR-1207-E

  • Page 1 NR Recorder Model NR-1207-E Installation and Operation Guide (for iOS Devices) Doc P/N: 2018-005|Norav Medical Revision 2.0 Updated 16/12/19...
  • Page 2: Copyright

    The NR Recorder is tested and certified for the following standards: EN60601-1 ■ EN60601-2-47 ■ ■ IEC60601-11 Protection type and class: BF, Internally powered ■ Defibrillation protection: yes, with Banana cable only. ■ Norav Mobile ECG Doc P/N: 2018-005| Norav Medical Updated 16/12/19 | Revision 2..0)
  • Page 3: Disclaimer

    Any action for breach of warranty shall be commenced within one (1) year of said breach or be forever barred. Any repairs made to the product that are not covered by the warranty shall be billed to the customer. For service or technical support contact the local supplier or Norav Medical. Document History Rev.
  • Page 4: Table Of Contents

    First Time the Norav Mobile ECG App Settings ..................... 13 Performing the ECG test ............................14 Saving Test Results and Generating Report ......................16 Tests Archive ................................16 Enhanced Features ..............................17 Norav Mobile ECG Doc P/N: 2018-005| Norav Medical Updated 16/12/19 | Revision 2..0)
  • Page 5: Document Conventions

    Magnetic and electrical fields can have an influence on the function of instruments. Ensure that all non ■ Norav Medical equipment which is operated nearby complies with the EMC requirements (regulations for Electro Magnetic Compatibility). X-Ray, CTs, etc. can cause interference to other equipment, as a result of their authorized higher emission of electromagnetic interference.
  • Page 6 NOTE: The manufacturer is not responsible for any Radio or TV interference caused by unauthorized modifications to this equipment. Such modifications could void the user’s authority to operate the equipment. NOTE NOTE: Install hardware only after software installation. Norav Mobile ECG Doc P/N: 2018-005| Norav Medical Updated 16/12/19 | Revision 2..0)
  • Page 7: Intended Use

    Populations of patients at an age or period in which a routine baseline evaluation of ECG characteristics is ■ desired. Contraindications for Use and Adverse effects The device has no contraindications or adverse events. Norav Mobile ECG Doc P/N: 2018-005| Norav Medical Updated 16/12/19 | Revision 2..0)
  • Page 8: Unpacking

    Unpacking Open the device package and verify the following package content: Description NR1207E-D-00E-02 NR Recorder, model NR-1207-E C10-B-E-EI 10 lead ECG cable for NR Recorder, banana ends, IEC, EI C-USB-EXT-2M USB cable 2M C-ADAPT-BANANA-SNAP-01 Banana to Snap adapters pack of 10 BAT-ALK-P-AA-EN91 Battery, Alkaline Primary, 1.5V, Size AA...
  • Page 9: Installing The Norav Mobile Ecg App

    2. When the ‘Norav Mobile ECG’ app description appears- tap Get. You might need to sign in with your Apple ID and password or use Touch ID or Face ID if you have them set up. Norav Mobile ECG Doc P/N: 2018-005| Norav Medical Updated 16/12/19 | Revision 2..0)
  • Page 10: Installing The Nr Recorder

    1. Attach the Patient Cable: Insert the ECG cable connector into the slot on top of the NR-1207-E unit. Make sure to insert the cable connector until both of two latches of the cable connector are locked on the unit.
  • Page 11 Press the button to return to the Settings screen. Select the Save as Default option the Bluetooth setting then press button to save the changes. Norav Mobile ECG Doc P/N: 2018-005| Norav Medical Updated 16/12/19 | Revision 2..0)
  • Page 12: Pairing The Nr Recorder With Ios Mobile Device

    Settings select Bluetooth make sure it’s turned ON select the NR Recorder name in the found Bluetooth accessory list tap Pair on the Bluetooth Pairing Request message Norav Mobile ECG Doc P/N: 2018-005| Norav Medical Updated 16/12/19 | Revision 2..0)
  • Page 13: First Time The Norav Mobile Ecg App Settings

    1. Tap the Norav Mobile ECG shortcut. When the App screen will appear—open Settings. 2. Enter the Physician and Clinic information then exit to App main screen by tap the X icon Norav Mobile ECG Doc P/N: 2018-005| Norav Medical Updated 16/12/19 | Revision 2..0)
  • Page 14: Performing The Ecg Test

    Hook up the Patient ■ Connect the LIMB leads: Connect the CHEST leads V1 V2 Entering Patient Information ■ Tap Start New Test Enter Patient Details then tap Start Norav Mobile ECG Doc P/N: 2018-005| Norav Medical Updated 16/12/19 | Revision 2..0)
  • Page 15 Wait about 10 secs you will see the Stop button flashes (means it’s ready to save). Tap Stop button to save the test and to enter Report Mode. Norav Mobile ECG Doc P/N: 2018-005| Norav Medical Updated 16/12/19 | Revision 2..0)
  • Page 16: Saving Test Results And Generating Report

    To access the test report—tap the selected patient row. To search patient by ID or by Name enter the looking value in the Search input field Norav Mobile ECG Doc P/N: 2018-005| Norav Medical Updated 16/12/19 | Revision 2..0)
  • Page 17: Enhanced Features

    1-5 sec except lead strip from 1-10 sec 12 Leads report showing all 1-10 sec. 1x6 report showing 1-10 sec of limb leads (I, II, III, aVR, aVL, aVF). Norav Mobile ECG Doc P/N: 2018-005| Norav Medical Updated 16/12/19 | Revision 2..0)
  • Page 18 Ready template ECG editable remarks In order to save time to the technician or the doctor, we’ve added ready remarks template (for ECG). The remarks are editable and managed. Norav Mobile ECG Doc P/N: 2018-005| Norav Medical Updated 16/12/19 | Revision 2..0)
  • Page 19 App. *Simulator mode will be labeled as Hazards in: ECG leads, summary report and database. Norav Mobile ECG Doc P/N: 2018-005| Norav Medical Updated 16/12/19 | Revision 2..0)
  • Page 20 PDF Report (e.g. other App). When this option is disabled and email recipients is definded (in the ‘Email To’ fields) it will open the default email client. Norav Mobile ECG Doc P/N: 2018-005| Norav Medical Updated 16/12/19 | Revision 2..0)

Table of Contents