The manufacturing process operates under Class 1 of the Medical
Devices Directive by fully complying with the essential
requirements for the CE marking of this product.
Flammability
The Pro-Mat-2 Mattress meets the requirements of BS7177:2008 for medium hazard.
The interior materials meet the requirements of The Furniture & Furnishings (Fire
Safety) Regulations 1988.
User Instructions
The Pro-Mat-2 Mattress has been designed for use on mesh, slat & solid mattress
supports and can be used on profiling & static bed bases, by patients in the 'at risk' user
category
with a weight guide of up to 185kgs / 29 stones. (NOTE: You should check
the Safe Working Load of the bed on which the mattress is to be used to ensure
compatibility with mattress & user.)
Ensure that the surface of the bed is clear from debris before positioning the mattress –
debris can damage the coating of the fabric.
Frequent turning & rotation of multi-turn foam mattresses is essential to ensure that
loading is evenly spread across both sides & both ends in order to minimise foam core
fatigue in use.
The cover on this mattress is removable for cleaning purposes & to allow for inspection
of the cover and the foam core for damage and contamination. It is recommended that
frequent inspections of the cover are made to check for damage such as holes or cuts.
Always ensure that, if removed, the cover is returned to the original foam core to avoid
cross-contamination. Using this cover on a different foam core may also render the
warranty invalid.
Instructions for use are screen printed onto the cover of the product for ease of
reference.
Indication of use based upon interface pressure readings – not a prescriptive criteria. Always defer to clinician's
judgement.
Cleaning
For general cleaning and superficial dirt, use a disposable wipe with a solution of warm
water and neutral detergent. Do not use abrasive or phenolic-based cleaners. Do not
mangle or iron.
To disinfect in situ, use a 0.1% sodium hypochloride solution (1,000 ppm chlorine).
A removable cover, identified as contaminated should be laundered at 65°C for not
less than 10 minutes or at 95°C for not less than 3 minutes. Do not mangle or iron.
Spin and tumble-dry at not more than 130°C. Thoroughly rinse and dry the cover
before re-use or storage. Store in a cool dry area.
Contaminated foam cores cannot be decontaminated and should be disposed of safely,
as should any cover damaged by cuts, tears or 'sharps' penetration.
Warranty
Covered against faulty manufacture and materials by replacement.
Normal expected wear and tear in use is not covered.
All warranties are subject to the terms and conditions of trading.
Replacement value under warranty reduces proportionally to the period in use up to 3
years after purchase.
If making a claim under warranty please quote the production identification batch
number printed on the interior foam core.
Storage & Care
Store the mattress flat – do not fold
Do not place heavy or sharp objects on the mattress surface when in use / in store.
Do not allow sharp objects to penetrate the cover coating (incl. hypodermic needles)
Keep the mattress clean
Regularly inspect the interior core and exterior cover of the mattress for damage or
contamination.
Do not store in damp conditions