READ THIS BEFORE USE.
The following basic safety precautions should always be taken.
NOTE
PLEASE KEEP THE USER'S MANUAL NEARBY THE GLUCOSE TEST METER.
Thank you for choosing the CERA-CHEK 1070 Blood Glucose Monitoring System. The CERA-CHEK 1070 is superior in accuracy, retest results, and portability. Stored test results are easily retrievable so that you may compare them to laboratory results with your physician.
This allows for better management of your diabetes and periodic blood glucose testing. Please be sure to read this manual before using the CERA-CHEK 1070 Blood Glucose Monitoring System. If you have any questions, please contact your local healthcare professional.
The CERA-CHEK 1070 Blood Glucose Monitoring System is intended to help people manage their diabetes with measuring their glucose level with fresh capillary whole blood from the fingertip, the palm, the forearm, the Upper arm, the thigh, the calf for self testing. It also provides professional healthcare personnel with the helpful information with measuring glucose concentration in fresh capillary whole blood as well as fresh venous blood. The CERA-CHEK 1070 Blood Glucose Monitoring System is intended for in vitro diagnostic use.
The test is based on the measurement of an electrical current generated by the reaction of glucose with the reagent of the test strip.
The test meter measures the current and displays the corresponding blood glucose level. The strength of the current produced by the reaction depends on the amount of glucose in the blood sample.
Your test meter has been designed, tested, and proven to work together as a system to produce accurate blood glucose test results. Use only with the manufacturer's compatible test strip and glucose control solution with this test meter.
Your system includes
NOTE
CERA-CHEK 1070 Blood Glucose Test Strips(including code key) and Glucose Control Solution are sold separately.
Insert Port
The insert port is where you insert the test strips for testing as well as the code key for calibration.
Power Button
Turn the meter ON or OFF Enter the "Set" mode Save the date, time, measurement unit and alarm settings.
Display
Your test results are display here.
Setting Buttons ( )
Change the date, time, measurement unit and alarm settings Recall previous test results from the meter memory.
Data Port
Send stored data to a computer via an optional data transfer cable to view.
Your test strip measures the concentration of glucose in whole blood. Blood is applied to the absorbent hole of the test strip and is automatically drawn into the reaction zone.
Contact Port: Insert this end of the test strip into the test meter. Push it in firmly until it will go no further.
Grip: Hold this part to insert or remove the test strip.
Confirmation Window: This is where you confirm if enough blood has been applied to the absorbent hole of the test strip.
Absorbent Hole: Apply a drop of blood here. The blood will automatically flow into the reaction zone.
When the using CERA-CHEK 1070 Blood Glucose Contact Monitoring System for the first time, or before using a Port new box of test strips, you will need to calibrate the test meter. There is a code key in every box of test strip. Calibration can be easily done by inserting the code key into the test meter. The code key must match the code numbers on the test strip vial and LCD display.
Please check the right direction to insert the code key &RGH1XPEHU into the test meter.
NOTE
To replace the battery
Replacing the battery will not delete test meter settings unless you
NOTE
Recheck and reset the clock setting as necessary after battery replacement to ensure time is set correctly. To set the meter clock, see Test Meter Set-up below.
Set-up Mode: The meter will enter the set-up mode when replace the battery. With your meter turned on or off, press the Power button for 3 seconds to enter the setup mode. To exit the set mode in any step of the setting, press the Power button for 3 seconds. The meter will save the settings and it will turn off.
Please always set the date and time before using the CERA-CHEK 1070 Blood Glucose Monitoring System for the first time or after replacing the battery. Do not perform a blood glucose test until you have set-up the test meter.
NOTE
Pressing the Setting buttons ( ) continuously will pass the numbers quickly.
NOTE
Default alarm times are 6:30, 9:30, 12:30, 18:30 and 21:30. You can choose from these or set your own time.
Special caution is needed to choose the measurement unit since the test results can be different according to your choice of measurement unit. Please consult with your diabetes healthcare professional about the measurement unit.
NOTE
Your test meter displays 7, 14 and 30days average which you can access from the test meter memory. These averages are calculated from results obtained during 7 14-30days from the current date and time settings. When you set a new date and time, the 7, 14-30day averages may change. After 2 minutes of inactivity while the test meter is in the setting mode, it will shut down automatically.
Insert the code key completely into the code key port with the code number facing up and the code number will appear on the LCD display. Make sure the code numbers on the display match the code key and the test strip vial. If all of the numbers match, you may begin testing.
NOTE
If the code numbers displayed on the screen do not match the code key and the test strip vial, try calibrating again or contact your local distributor.
Checking the System with Glucose Control Solutions CERA-CHEK 1070 Glucose Control Solution contains glucose that reacts to the test strips. By comparing the result tested using the control with the expected range printed on the test strip vial label, it is able to check that the test meter and the test strips are working together as a system and that you are performing the test correctly. It is very important that you do this simple check routinely to make sure you get accurate results.
Perform a Glucose Control Solution Test if
NOTE
The results obtained from your Glucose Control Solution test do not indicate your personal glucose levels.
Use CERA-CHEK 1070 Blood Glucose Test Meter only with CERA-CHEK 1070 Test Strips. Glucose control solution is sold separately. Do not use the control solution after expiration date as printed on the vial. Use the control solution within 4 months after opened. It is recommended to record the discard date on the coming 4 months on the glucose control solution vial as a reminder when to dispose the opened control solution.
[For example: You open the control solution on January 1st. Count forward four months to April 1st. This is your discard date. Record this date on the side of the control solution vial.]
Always close the cap on the Glucose Control Solution immediately after use.
If the test result is out-of-the range numbers printed on the test strip vial, repeat the test. If you continue to have out-of-range results, it means that the system may not be working properly. Do not use the product and please contact your local distributor for assistance.
Please wash and dry your hands prior to starting.
You have the option to get a drop of blood from an alternative site (forearm, upper arm, palm, thigh and calf) in addition to blood taken from your fingertip. When blood glucose is changing rapidly, fingertip samples show these changes more quickly than alternative site samples. Please read this section to understand which option is right for you.
Do not use Alternative Site Testing under the following conditions.
A lancet should only be used once. Please dispose the used lancet in a safe manner in order not to cause accidental injury.
Never share a lancet of the lancing device with another person.
Always use a new lancet.
Make sure the code number shown on the display matches the code number printed on the test strip vial. If the code numbers are different, please refer to the section of "Coding the Test Meter", for the procedure of coding.
NOTE
If you do not apply a blood sample to the test strip within 2 minutes, the test meter will automatically shut down. You must remove and reinsert the test strip to restart the test procedure.
The blood sample should fully fill the confirmation window in order for the countdown to begin. If you find that the confirmation window is not fully filled when the test meter is counting, DO NOT try to add more blood to the test strip. Discard the test strip and retest with a new one.
Please discard the used test strip immediately after use.
DO NOT REUSE THE TEST STRIP.
CERA-CHEK 1070 can store up to 1000 test records. Each record includes the test result, time and date. The meter will also calculate the average values of records from the last 7, 14, 30days.
NOTE
If you are using the meter for the first time, the meter display will show three dashed lines (---) and the 'memory' symbol ( ).This means that no data has been stored in memory.
NOTE
If there are already 1,000 records in the memory, the oldest record will be erased to make room for a new one. Your most recent result will display first.
Extreme caution should be used when clearing the memory. This is not a reversible operation.
Time of day | Plasma Glucose range (mg/dL) for people without diabetes. | Your target range (mg/dL)/(mmol/L) |
Before a meal | Less than 100mg/dL (5.6 mmol/L) | |
2 hours after a meal | Less than 140mg/dL (7.8 mmol/L) |
Source: American Diabetes Association; Diadetes Care, January 2007, Volume 30, Suppl 1 S42~S47
CERA-CHEK 1070 provides a comfortable function to our users; blood glucose management software that can transfer and show analytical interpretation of all the stored data of blood glucose results from your glucose meter into your computer.
※ To use the blood glucose management software, you should purchase separately the installation CD and the USB data transfer cable.
NOTE
When you disable the cable during the transfer of data, the data may not be transferring properly. Please keep the transfer cable intact until the transfer is completed.
In case you need more details about the blood glucose monitoring software of CERACHEK 1070, please contact your local distributor.
To avoid contaminating your test meter and test strips, please wash your hands and test site with soap and water. Make sure to dry your hands and test site thoroughly.
Your test meter does not require special maintenance. There is no special cleaning required because no blood or control solution comes in contact with the test meter. Avoid getting dirt, dust, blood, or water inside the test meter through the test port. A damp cloth and mild detergent can be used to wipe the outside of the test meter. Your test meter is a precision instrument. Please handle it with care.
Message me | Possible cause | What You Should Do |
![]() | Test result level is over 900mg/dL. (50.0mmol/L) | Repeat the test with a new test strip. If this message shows again, please contact your doctor immediately. |
![]() | Test result level is less than 10mg/dL. (0.6mmol/L) | Repeat the test with a new test strip. If this message shows again, please contact your doctor immediately. |
![]() | Prepare to replace the battery. | Recommending to replace the battery. |
![]() | The battery power is too low for testing. | Replace the battery at once. |
![]() | The operating temperature of the system is not the mentioned range. 4°C ~ 40°C (39°F ~ 104°F). | Allow the test meter and test strips to warm up or cool down slowly (30 minutes) until the error message goes away. |
![]() | Test strip is used. | Repeat test with a new test strip. |
![]() | Failure to operate during a test. | Repeat test with a new test strip. |
![]() | Test strip was removed during a test. | Repeat test with a new test strip. |
![]() | Damaged code key or the code key was removed during a coding the meter. | Re-coding the Test Meter If the problem still continues, please contact your local service center. |
![]() | The code is not set. | Insert the code key that acco mpanied the box of the test strips. |
Product Name: CERA-CHEK 1070 Blood Glucose Monitoring System
Model: G 300
Calibration: Plasma (YSI 2300 STAT PLUS)
Sample material: Fresh whole blood from capillary and venous
Sample Size: 0.5
Hematocrit Range: 10% ~ 70 %
Test Time: 5 seconds
Dimensions: 53.6 x 94 x 13.9 (mm)
Weight: 35 g
Power source: One CR2032 3V lithium coin battery
Display: LCD
Memory: 1000 measurement results with date and time
Automatic Turn Off: Shuts down after 2 minutes of inactivity
Measurement Units: Either mg/dL or mmol/L
Operating Condition: 4°C~40°C (39°F~104°F), below 85%, relative humidity
Altitudes up to 4,000m (13,200ft)
Measurement Range: 10~900 mg/dL (0.6~50 mmol/L)
Storage/Transport Condition: -10°C~60°C (14°F~140°F)
(Blood Glucose Test Meter) below 85% relative humidity, 700 to 1,060 hPa
The accuracy of the CERA-CHEK 1070 Blood Glucose Monitoring System was evaluated by comparing blood glucose results obtained from 100 subjects with replicate measurement. Readings obtained with the CERA-CHEK 1070 Blood Glucose Monitoring System were compared to those obtained using an YSI Model 2300 Glucose Analyzer.
Accuracy results for glucose concentration < 75mg/dL (4.2 mmol/L) | |||
Within ± 5 mg/dL (Within ± 0.28 mmol/L) | Within ± 10 mg/dL (Within ± 0.56 mmol/L) | Within ± 15 mg/dL (Within ± 0.83 mmol/L) | |
70%(14/20) | 95%(19/20) | 100%(20/20) | |
Accuracy results for glucose concentration ≥ 75 mg/dL (4.2 mmol/L) | |||
Within± 5% | Within± 10% | Within± 15% | Within± 20% |
78%(141/180) | 89%(160/180) | 100%(180/180) | 100%(180/180) |
The precision of CERA-CHEK 1070 Blood Glucose Monitoring System was estimated with venous blood samples and Glucose Control Solutions in the laboratory.
Within Run Precision | ||
Blood average | 45 mg/dL(2.5 mmol/L) 98 mg/dL(5.4 mmol/L) 129 mg/dL(7.2 mmol/L) 228mg/dL(12.7 mmol/L) 326 mg/dL(18.1 mmol/L) | STD= 2.4 mg/dL CV = 3.1 % CV = 3.2 % CV = 3.0 % CV = 2.2% |
Day to Day Precision | ||
Control average | 43 mg/dL(2.4 mmol/L) 108 mg/dL(6.0 mmol/L) 304 mg/dL(16.9 mmol/L) | STD= 2.3 mg/dL CV = 3.2 % CV = 2.8% |
Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.
Download Cera-Chek 1070 - Blood Glucose Monitoring System Manual