Page 1
USER MANUAL Vacc-Safe PHARMACY REFRIGERATOR Warning: Before using this product, please read this manual carefully and keep it for future reference. The design Model: VS350 and specifications are subject to change without prior notice for product www.vaccsafe.com.au improvement.
Safety warnings Warning: Flammable refrigerant is adopted for the product. WARNING: Keep all ventilation openings in the enclosure or, in the structure for building-in, clear of obstruction. WARNING: Do not use mechanical devices or other means to accelerate the defrosting process, other than those recommended by the manufacturer.
Product features This product can be used to store human biological samples, medicines, vaccines, biological products, reagents, and materials that need to be stored at 2~8°C. Product function Accurate temperature control & traceable data Microcomputer temperature control display with an accuracy of 0.1°C Temperature can be set between 2°C to 8°C, and in the box, it fluctuates ≤2°C, adjustable and accurate.
Precautions and safety measures Dear users of Vacc-Safe: Hi! Thank you for using a Vacc-Safe pharmacy refrigerator. For your convenience to understand this manual, use the product, and prevent personal injuries and goods damages, please read the following instructions thoroughly.
Page 6
If the power cord needs to be lengthened, the sectional area of the lengthened cord shall not be less than 2.5mm . At the same time, the length is not longer than 3 m. Otherwise, it may cause a fire or electric shock. In case of flammable gas leakage, it is necessary to close the gas leakage valve and open doors/windows, for ventilation.
Page 7
If the pharmacy refrigerator is not working properly, please unplug the power plug. It may cause electric shock or fires if it continues to operate under abnormal conditions. Before any repair or maintenance of the pharmacy refrigerator, always disconnect the power supply of the pharmacy refrigerator to prevent electric shock or personal injuries.
Page 8
Do not climb on the pharmacy refrigerator or put anything on the pharmacy refrigerator, otherwise it may fall over and cause personal injuries or damages to the pharmacy refrigerator. Don't use the pharmacy refrigerator in the open air. When the pharmacy refrigerator is wet by rain, it may cause electric leakage or electric shock.
Page 9
--------------------------------------------------------------------- Note There should be no obstruction around the pharmacy refrigerator and the ventilation should be smooth. It is necessary to check the pharmacy refrigerator setting when restarting the pharmacy refrigerator after power failure or power off. C hanges in settings may damage stored items. O nce the power of the pharmacy refrigerator is cut off, it will take more than five minutes to turn it on again to avoid damages to the compressor or system.
Precautions for use Before putting items into the device, please make sure that the internal temperature in the cabinet has reached the set temperature, and then put the items into the pharmacy refrigerator .i n batches. The volume of the articles in the box should not exceed 1/4 of its volume at a time to prevent the temperature from rising too much.
Product installation Product handling --------------------------------------------------- Protect the pharmacy refrigerator when moving it, same as shown as below photo, please move it by handcart with cushion. Remove all packing materials and bottom cushion, then move into position for placement. After moving it to appropriate location, wait for 2 hours before powering on. Installation environment ------------------------------------------ The pharmacy refrigerator is only for indoor use, with second pollution level and...
Page 12
Installation site ------------------------------------------------------ D o not install the device in a narrow and confined space and install it in a place that can ensure its normal access, to avoid damage to storage goods by maintenance difficulties and delayed repair in case of device failure. The installation floor must be solid, flat, non-combustible and able to bear the weight of the device during operation.
Page 13
Preparation before use ------------------------------------------- 1. Remove the outer packaging of all products (including the protective foam in the packing box). 2. Check the accessories; please check the contents in the box according to the packing list. If there is any discrepancy, please contact the after-sales service immediately.
Page 14
A special person shall be responsible for the operation of this product, and check and record it every day. If the temperature in the cabinet is too high or too low, the goods shall be transferred to other pharmacy refrigerator, and the goods shall not be stored again until the fault is eliminated.
Page 15
Initial start-up ------------------------------------------------------ Please observe the following rules when conducting initial start-up: 1.Under no-load condition, please connect the power plug to right plugboard, and the plugboard should not be shared with other appliances. 2.After turning on the power supply, turn on the battery switch of the device, check if the inner LED light works properly.
Wire shelf Glass door body Universal wheel Schematic Diagram VS350 Due to the improvement of the product and different models, the actual product may be different from the diagram. Please refer to the actual product! The diagram is only for the description of functional parts.
Operating Instruction Operation and display panel Temperature (°C) Set/Mute Temperature + Temperature - Light Alarm Description of digital tube display Under normal circumstances, the display board displays the actual tempera- ture in the refrigerator; During parameter setting, the display board flashes at an interval of 1 second to display the set temperature value;...
Page 18
Instruction Long press Short press (5s) Set/Mute Enter/Exit the multiplex key parameter setting Up key Plus 0.1 Single key Down key Reduce 0.1 Light switch Turn on/o ffthe light Combination Set/Mute + Up key Lock, unlock Note: 1. When a fault alarm is generated, press any key to mute. 2.
Page 19
User parameter table Parameter Description Scope Note number Set the temperature of 2-8℃ Default 5.0°C Pharmacy refrigerator chamber 0.0~10.0℃ High temperature alarm value Default 3.0 0.0 Cancel the alarm 0.0~10.0℃ Low temperature alarm value Default 3.0 0.0 Cancel the alarm The set temperature range of the Pharmacy refrigerator chamber is 2-8°C.
Page 20
4. Alarm description In case of sensor failure, communication failure, high and low temperature alarm or door switch alarm in the refrigerator chamber, the display board will display the fault code, and the buzzer will "beep" at an interval of 1 second. Press any key to mute, but the fault code will not disappear until the fault is removed.
Maintenance and care Warning T o prevent electric shock or personal injuries, please make sure that power is disconnected before having any repairs or maintenance on the device. W hen maintaining the device, make sure that you do not inhale surrounding medications or suspended particles that could damage your health.
Page 22
End-of-life period: W hen the device reaches the end of its service life, it should be scrapped and must be handed over to qualified professional recycling organizations for scrapping and disposal in accordance with local regulations and ordinances; non-professionals should not disassemble or decompose the device without authorization;...
Specifications VS350 316L 220-240V~/50Hz 190W/1.0A Packing list MC-5L316B *The warranty card and certificate are included in the instruction and the specific packing list shall prevail in accordance with the real product.