Maintenance
Proper care and maintenance of your Bio-Pump
electronic device and should be treated accordingly. Rough and abusive use and/or accidental
impacts may damage the unit. Zefon off ers the following guidelines in caring for your equipment.
1. Keep the unit clean and free of dust and dirt. It may be wiped down with a clean dry cloth.
2. Keep the battery charged. Charge periodically when not used for a prolonged amount of time or
leave plugged in for a continuous trickle charge.
3. Always store Bio-Pump
Plus in its case for proper protection.
®
4. Do not ship the Bio-Pump
Plus using the carrying case as the shipping container. The case is not
®
designed to be a shipping container and damage may occur.
5. Do not attempt to oil the motor. The motor is designed to be maintenance free.
6. Do not store the unit in extreme heat for any extended period of time.
ZA0043 Heavy Duty
Tripod Sampling Stand
Optional Accessories
Heavy Duty Tripod Sampling Stand
Remote 5' Extension Tubing
Parts & Accessories
Bioisolation Filter
AC Power Adaptor, 100v-240v
Battery
Cassettes (10/pk)
®
Air-O-Cell
Cassettes (50/pk)
®
Air-O-Cell
Cassettes (10/pk)
®
Via-Cell
5350 SW 1st Lane, Ocala, FL 34474 USA
Tel: +1 (352) 854-8080; Fax: +1 (352) 854-7480; www.zefon.com
®
Plus is essential for a long useful life. It is a delicate
ZBP-310 Remote Extension
shown attached to Bio-Pump
®
ZA0043
ZBP-310
ZA0047
ZBP-304
ZBP-315
AOC010
AOC050
VIA010
LA03067 Rev 3
Bio-Pump
with TSI Calibrator Kit
Product # ZBP-205-CKIT
Description
The Bio-Pump
Plus is an advanced portable, battery powered air sampling pump designed for the
®
exclusive use with Air-O-Cell
and Via-Cell
®
Specifi cations
Flow Rate:
15LPM
Accuracy:
± 5 %
Housing:
PC/ABS plastic
Battery:
Rechargeable Nickel Metal Hydride (NiMH)
Run Time:
100 fi ve minute samples on new, fully charged battery
Dimensions:
4.5" (w) x 2.75" (h) x 8" (d) inches
Weight:
1.75 lbs
5350 SW 1st Lane, Ocala, FL 34474 USA
Tel: +1 (352) 854-8080; Fax: +1 (352) 854-7480; www.zefon.com
Operating Instructions
Plus
®
cassettes.
®
LA03067 Rev 3

Advertisement

loading

Summary of Contents for Zefon International Bio-Pump Plus ZBP-205-CKIT

  • Page 1 Maintenance Operating Instructions Proper care and maintenance of your Bio-Pump ® Plus is essential for a long useful life. It is a delicate electronic device and should be treated accordingly. Rough and abusive use and/or accidental impacts may damage the unit. Zefon off ers the following guidelines in caring for your equipment. Bio-Pump Plus ®...
  • Page 2 6. Do not open the pump case except to change out an old battery. 7. Do not block the vent holes under the unit. Limited Warranty Zefon International warrants the Bio-Pump ® Plus for a period of 1 year from date of purchase. Zefon will without charge repair or replace, at its option, the defective pump or component parts upon delivery to its repair facility.
  • Page 3 Taking Samples Using Sequential Sampling Mode Pump Overview 1. Setup the pump as described in steps 1-3 for single sampling mode. 2. Using the arrows, select “SEQUENTIAL SAMPLING MODE”, then SEQUENTIAL press SET. SAMPLING MODE 3. Using the arrows, select the number of on/off cycles, then press SET. 4.
  • Page 4 Calibration Sampling Modes It is recommended that the Bio-Pump ® Plus be calibrated and verifi ed at least once every day of use. Bio-Pump ® Plus comes with 3 diff erent sampling modes to choose from: 1. Quick Sample Mode Quick Sample modes allows easy sampling of the most commonly used sampling times of 1, 2, 5, Connect the Calibrator to the Bio-Pump ®...