Parker Laboratories Thermasonic Gel Warmer Manual
Also See for Thermasonic Gel Warmer:

Advertisement

Parker Laboratories Thermasonic Gel Warmer Manual

IMPORTANT SAFEGUARDS

This symbol when indicated on the equipment means:
Read these operating instructions fully before use and pay particular attention to sections containing this symbol.

Always observe the following safety precautions:

  1. READ THESE INSTRUCTIONS.

  2. To provide continued protection against risk of electric shock, connect to properly grounded outlets only.
  3. Do not immerse in water.
  4. Unplug from outlet when not in use and before cleaning. Allow unit to cool before cleaning. Before moving, disconnect at the power supply socket.
  5. Do not operate any appliance with a damaged cord or plug or after a malfunction, or has been damaged in any manner. Return unit to an authorized service facility for examination, repair or adjustment.
  6. Do not use outdoors.
  7. Do not use appliance for anything other than intended use.
  8. The use of attachments not recommended by the manufacturer may result in fire, electric shock or injury to persons.
  9. SAVE THESE INSTRUCTIONS.

INCLUDED IN PACKAGE


Thermasonic Gel Warmer

Operating Instructions

*Power Cord – IEC320/C14 Compatible (230V units supplied with European plug CEE 7/7)


Screws and Wall Anchors

ASSEMBLY

INSTALLATION

The Thermasonic Gel Warmer should be placed away from drafts, in an upright position on a flat, stable surface. Do not lay the unit flat on any side. The unit may also be wall mounted. See wall mounting instructions. Plug female end of power cord into IEC320/C14 receptacle and male end into an appropriate electrical outlet matching the voltage listed on the electrical information label on the back of the gel warmer.

*NOTE: UK power cord with a BS -1363 plug (Part #: UKCORD) is available in the United Kingdom, as an accessory from your distributor.

WALL MOUNTING

The Thermasonic Gel Warmer can be wall mounted with use of the supplied screws and wall anchors. Follow these steps to wall mount.

  1. Once wall mount location has been identified, make two marks on the wall 1.75 inches (4.45 cm) apart. When making marks, use a level to assure the Gel Warmer will appear straight when complete.
  2. If mounting into wood, install screws, and allow 0.1" (2.5mm) clearance between the wall and the screw head. If mounting into drywall, the use of wall anchors is necessary.

    Never drill or install screws in a wall with wires or plumbing running through it.
  3. Carefully position the Gel Warmer so the large round wells (see Figure 2) on the back of the unit will pass over the screw heads.
  4. Press the Gel Warmer against the wall and slide downward, trapping the screw head in the grooves just above the wells.

OPERATION

HEAT INDICATOR LAMP

The Heat Indicator Lamp will blink while the unit is approaching the set temperature. Once that temperature has been reached, the lamp will remain illuminated. Should the unit restart the heating process, the Heat Indicator Lamp will again begin to blink.

OPERATING THE GEL WARMER

The Gel Warmer will warm a full (40°C) bottle of gel to 104°F (40°C).


Temperature can vary. Always test a small quantity of gel before applying.

Allow approximately two hours for a full bottle of gel to reach proper temperature. Your Thermasonic Gel Warmer is designed for continuous operation, and it is not necessary to disconnect the warmer when not in use.

EROR CODES

The Thermasonic Gel Warmer contains a temperature sensor to consistently monitor functions and a Microtemp® thermal cutoff to protect against overheating.

If the temperature sensor is not operating at all, the Heat Indicator Lamp will blink once every five seconds. If a short circuit occurs, the Heat Indicator Lamp will blink twice every five seconds. In either circumstance, the heater will not operate.

TECHNICAL SPECIFICATIONS

This equipment is intended for indoor use and will meet its performance figures within the ambient temperature range of 68˚F (20˚C) to 95˚F (35˚C), with maximum relative humidity of 80% (non-condensing). Installation Category II (transient voltages). Pollution Degree 2 in accordance with IEC 664. Suitable for operation at altitude of up to 6,500 feet.

Specifications:

Supply Voltage Range: 120V~ +/- 10%, 50/60 Hz,

230V~ +/- 10%, 50/60 Hz (CE version)

Power Rating: Model 82-01: 12W, 0.1A,

Model 82-01-20: 12W, 0.06A

MAINTENANCE AND SERVICE

No routine maintenance is required and there are no user serviceable parts.

Cleaning

Disconnect the Gel Warmer from the electrical outlet prior to cleaning. The Gel Warmer may be cleaned by using a soft, damp cloth and a mild detergent or hospital grade disinfectant, such as ProteX™ ULTRA Wipes. Do not allow the electronic controls to get wet. Do not use an abrasive cleaner or isopropyl alcohol to clean the Gel Warmer. Do not submerge or immerse the Gel Warmer in water.

Documents / Resources

Download manual