Summary of Contents for SunTech Vet30

  • Page 1 Continuous Vital Signs Monitor for Companion Animals User Manual...
  • Page 2: Changes

    All content in this manual is the proprietary information of SunTech Medical and is provided solely for purposes of operation of the SunTech Vet30. This manual and the Vet30 described in it are protected under copyright law under which they may not be copied, in whole or in part, without written consent of SunTech Medical.
  • Page 3: Welcome To The Suntech Vet30

    The Vet30 is a portable device, weighing approximately 2.65 lb (1202 g). Control buttons allow the user to turn on or off the device and stop/start a BP measurement. The touchscreen display shows the device status and measurement information, and allows for selection of settings.
  • Page 4 Use of Vet30 Use only veterinary BP cuffs supplied by SunTech Medical. Observe the patient carefully during the measurement. Accuracy of any BP reading or oxygen saturation measurement may be affected by the position of the patient, their physical condition and use outside of the operating instructions detailed in this guide. The interpretation of BP and oxygen saturation measurements should only be made by a veterinarian.
  • Page 5: Table Of Contents

    8.5 Reviewing Data and Creating Reports..... 31 9. Alarms ..............33 9.1 Technical Alarms ..........33 9.2 Clinical Alarms ..........33 10. Taking Care of the SunTech Vet30 ..... 34 10.1 Battery ............34 10.2 Cleaning............34 10.3 Preventative Maintenance ......35 10.4 System Components ........
  • Page 6: Warnings And Precautions

    WARNING: DO NOT use if device is dropped and/or is damaged. Have a qualified service representative check the unit before using again. Repairs should only be conducted by an authorized SunTech Medical service representative. WARNING: Before performing a procedure when device is not plugged into main power, ensure there is sufficient battery charge.
  • Page 7 DO NOT dispose of battery in domestic waste! Instead, please dispose of in an environmentally responsible way, or return the battery to SunTech Medical. A prepaid return label can be obtained. Please see our website for more information about our environmental policy at http://www.suntechmed.com/about-suntech/environmental-...
  • Page 8: Icons And Symbols

    Polarity of DC power connector IEC 60417-5639 Rechargeable battery ISO 7000-1641 Consult Instructions for Use This product meets the requirements of the applicable EU Directive Directives WEEE Directive Disposal in compliance with WEEE Directive 80-0075-00-RevK SunTech Vet30 User Manual | 8 2022-July-31...
  • Page 9 Exits or closes a screen SunTech Design Goes back to previous screen SunTech Design Deletes all stored memory data SunTech Design Averages selected BP readings SunTech Design Begins STAT Mode Bluetooth Bluetooth 80-0075-00-RevK SunTech Vet30 User Manual | 9 2022-July-31...
  • Page 10: Getting To Know Suntech Vet30

    BP Cuffs • Sensors • Temperature Probes The Vet30 has been tested with specific parts of the system used within the patient environment. The parts of the system that can be used in the patient environment are: • Vet30 monitor •...
  • Page 11: Front Panel

    CAUTION: Read all warnings, cautions, and instructions provided prior to using the device. Insert AC adapter plug into the Vet30. Plug other end of the AC adapter into a power supply to begin charging the battery. Only use AC Adapter supplied by SunTech Medical.
  • Page 12 If applicable, attach sensor to extension cable and lock plastic hinged cover to prevent For additional information on how to take an SpO measurement, see Section 7. accidental disconnection. 80-0075-00-RevK SunTech Vet30 User Manual | 12 2022-July-31...
  • Page 13 AccuVet SpO B) Masimo SpO Attach the LNC Patient Cable to the Vet30. Orient the connector of the cable to mate with the patient cable connector on the back of the Vet30 and insert. Masimo SpO Attach the LNC Patient Cable connector to a LNCS reusable sensor connector. Orient sensor connector (1) to patient cable connector (2) as shown.
  • Page 14: Display

    Animal Selection Screen When you turn on your SunTech Vet30, you will see the Animal Selection Screen. Even if you are not ready to measure BP on a patient, select either Large or Small animal so that you can proceed to the Main Screen. This screen can be...
  • Page 15 Larger view of the current Larger view of the current Larger view of the current temp reading BP reading % and HR reading • Plethysmograph waveform TIP: The waveform is not normalized. 80-0075-00-RevK SunTech Vet30 User Manual | 15 2022-July-31...
  • Page 16: Start/Stop Button

    4.2 Volume The SunTech Vet30 monitor’s default setting is Speaker Off so that patients are not disturbed by sounds. To toggle between Speaker Off, Low Volume, and High Volume, press the Speaker button. When sound is OFF, you still will hear a very quiet clicking when you touch buttons but no beeps.
  • Page 17 Language. Select desired language. Clinical Alarms The SunTech Vet30 allows clinical alarms for all values (SYS, DIA, MAP, HR, Temp, and SpO ). Factory setting is OFF. • To change alarm values, select the Settings screen in the Icon Menu, then choose Clinical Alarms.
  • Page 18 Power Management The Vet30 includes an Auto OFF feature which turns the monitor off after a selected period. The Auto OFF timer is disabled when powered by an AC Adapter and during Interval Measurement Mode. To edit the Auto OFF timer, select the Settings screen in the Icon Menu, then choose Power Management.
  • Page 19 Use the + and – buttons to select the hour and minute. In AM/PM format, touch the AM or PM button to toggle between them. • To exit without changing the time, touch Cancel. To save the time, select Set. 80-0075-00-RevK SunTech Vet30 User Manual | 19 2022-July-31...
  • Page 20 Multiple languages are available including English, French, Italian, German, Spanish and Portuguese. To access the languages, select the Settings screen in the Icon Menu, then choose Language. • Touch the desired language to select it. Press X to save and exit. 80-0075-00-RevK SunTech Vet30 User Manual | 20 2022-July-31...
  • Page 21: Blood Pressure Measurement

    WARNING: The cuff should not be applied over a wound as this can cause further injury. SunTech recommends that the cuff be placed on a front limb while the patient is lying on the right or left side. This helps ensure that the cuff is at heart level, which is best for measurement accuracy. Also, the patient is less likely to retract the front limb when the cuff gently squeezes it during the measurement.
  • Page 22: Taking A Bp Measurement

    TIP: For information on averaging BP measurements, see Section 8.2. 5.4 Interval BP Mode In Interval BP Mode, the SunTech Vet30 automatically takes a BP reading at set time intervals. The time interval can see Section 4.3 be changed through the Interval BP settings ( •...
  • Page 23: Stat Mode

    WARNING: Check the limb frequently to ensure that operation of the monitor does not result in prolonged impairment of the circulation of the patient. During STAT Mode, the SunTech Vet30 will automatically take continuous BP measurements for 10 minutes. •...
  • Page 24: Taking A Temperature Measurement

    TIP: Clean the probe before and after each use making sure to remove all bioburden. 1. Place temp probe on patient and connect to Vet30. Position patient so that they are lying down. Apply disposable, non-rigid probe cover if using. Place the probe either in the esophagus or rectum.
  • Page 25: Masimo Pulse Oximetry

    NOTE: A functional tester cannot be used to assess the accuracy of the pulse oximetry function of the Vet30. NOTE: High-intensity extreme lights (such as pulsating strobe lights) directed on the sensor, may not allow the Vet30 to obtain vital sign readings.
  • Page 26 NOTE: Additional information specific to the Masimo sensors compatible with the Vet30, including information about parameter/measurement performance during motion and low perfusion, may be found in the sensor's directions for use (DFU). NOTE: Cables and sensors are provided with X-Cal™ technology to minimize the risk of inaccurate readings and unanticipated loss of patient monitoring.
  • Page 27: Accuvet Pulse Oximetry

    Connect the sensor connector and the patient cable connector, securing with the protective cover. complete instructions on connecting the Masimo sensor and patient cable to the Vet30, see Section 3. To disconnect the sensor: Lift the protective cover to gain access to the sensor connector.
  • Page 28 Refer to the Directions for Use included with the Masimo sensor for additional application information. 7.2 AccuVet Pulse Oximetry NOTE: A functional tester cannot be used to assess the accuracy of the pulse oximetry function of the Vet30. Where to Apply the AccuVet Y-lingual SpO Sensor Clean the sensor before use.
  • Page 29: Heart Rate Measurement

    Measurement If a sensor is connected to the Vet30 and to the patient, but not displaying a reading, observe the display for one minute to determine if the monitor is searching for a reading. When the monitor is searching for a signal, the dashes in the SpO section of the screen will flash.
  • Page 30: Averaging Measurements

    8.3 Bluetooth Connection The SunTech Vet30 can pair with a PC within a 30-foot radius to transfer data via Bluetooth. To pair with a PC, the Vet Data Capture PC application must first be installed on the PC. The application is available for download from the SunTech website at https://www.suntechmed.com/vet30-data-capture.
  • Page 31: Data Transfer

    To begin a data transfer, touch the Data Transfer icon ( ) in the Icon Menu of the Vet30. A status message will pop up during the transfer on the Vet30 and will appear at the bottom of the Vet Data Capture application.
  • Page 32 Once data transfer is complete, select the data points to be included in the report by holding Ctrl and clicking on each one. For large amounts of consecutive data points, hold Shift and click the first and last data points to be included. 80-0075-00-RevK SunTech Vet30 User Manual | 32 2022-July-31...
  • Page 33: Alarms

    To test the clinical alarms, turn the volume of the monitor on and set the alarm at a value likely to be exceeded with the next reading. Perform the reading. When the value exceeds the limit set, the alarm should display and a beep should sound. 80-0075-00-RevK SunTech Vet30 User Manual | 33 2022-July-31...
  • Page 34: Taking Care Of The Suntech Vet30

    The Vet30 has an internal user-replaceable battery that is charging whenever the monitor is plugged into a power source. The last segment of the battery indicator will turn red when battery charge is low. During continuous...
  • Page 35: Preventative Maintenance

    System Self Checks The Vet30 performs a range of system checks during normal operation. If the monitor detects a problem, it will display a status message recommending a troubleshooting action or to contact SunTech Customer Service. Replaceable Parts CAUTION: Replace user replaceable parts that are broken, worn, missing, damaged, incomplete, or contaminated.
  • Page 36: System Components

    CAUTION: Use only approved accessories. Replace battery only with SunTech part number 17-0026-00. Expected Service Life Monitor: 7 years (dependent upon internal BP pump usage) AccuVet SpO sensor: 3 years Masimo Patient Cable: 17,520 hours of patient monitoring time AccuVet temperature probe: 3 years Lithium-ion Battery: 250 cycles (before capacity falls below 80% of original capacity) 10.4 System Components...
  • Page 37: Accessories & Replacement Parts

    52-0025-00 Pack of 3 sizes (small, medium, large) Rechargeable Lithium-Ion Battery 17-0026-00 7.2 V, 2.2 Ah, 15.8 Wh Carrying Case 45-0010-00 Nylon bag that fits monitor and accessories USB Bluetooth Transmitter 45-0008-00 80-0075-00-RevK SunTech Vet30 User Manual | 37 2022-July-31...
  • Page 38: Status Messages

    10.6 Status Messages If the SunTech Vet30 has a problem performing a task, a short beep will occur (if speakers are ON), the START/STOP button will turn Red and a Status Message will appear on the monitor screen. Take action as directed on the screen, or as suggested in the table below.
  • Page 39 10.7 SpO Status Messages If the SunTech Vet30 encounters a problem related to the Masimo SpO functionality, a warning sign will appear on the main screen. Touch the warning symbol to open the SpO focus screen, where further information will be available about what went wrong.
  • Page 40: Frequently Asked Questions

    Check that the SpO sensor is fully plugged into the back of Communications use or system communications the Vet30. If error continues, service or a sensor replacement failure. may be required. 11. Frequently Asked Questions How many measurements can I store in Memory? You can store up to 960 measurements.
  • Page 41: Limited Warranty

    United States prepaid to the factory. SunTech Medical, Inc. will repair any component(s) or part(s) that it finds to be defective during the period of this limited warranty. Should a defect become apparent, the original purchaser should first notify SunTech Medical, Inc.
  • Page 42: Factory Default Settings

    Clinical Alarms – Low SpO Clinical Alarms – High SpO 100% Speaker Status Auto OFF Timer 10 minutes Interval BP Rate 3 minutes Language English Startup Animal Selection Date Format MM/DD/YYYY Time Format 12-hour Bluetooth Status 80-0075-00-RevK SunTech Vet30 User Manual | 42 2022-July-31...
  • Page 43: Performance Specifications

    Range 0-100% 0-100% Accuracy ±2% at 70%-100%, for patients ±2% at 70%-100%, for patients weighing >30 kg weighing >30 kg Undefined, for patients weighing Undefined, for patients weighing <30 kg <30 kg 80-0075-00-RevK SunTech Vet30 User Manual | 43 2022-July-31...
  • Page 44: Radio Frequency Compliance Requirements

    This is a Class I device that contains a wireless transmitter which can be used in at least one EU member state. There are no restrictions of use. 13.5 Electromagnetic Compatibility (EMC) Changes or modifications to the SunTech Vet30 that are not approved by SunTech Medical may cause EMC interference problems with this or other equipment. 80-0075-00-RevK...
  • Page 45 WARNING: Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the Vet30 system, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
  • Page 46 Guidance and Manufacturer’s Declaration – Electromagnetic Immunity The Vet30 Monitor is intended for use in a Professional Healthcare Facility within the electromagnetic environment specified below. It is not intended for helicopter transport, hospital ambulance or home use. It is not intended for near active HF SURGICAL EQUIPMENT and the RF shielded room of an ME SYSTEM for magnetic resonance imaging, where the intensity of EM DISTURBANCE is high.
  • Page 47 (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m) and E is the Immunity Test Level in V/m. �� = √�� �� 80-0075-00-RevK SunTech Vet30 User Manual | 47 2022-July-31...
  • Page 48 380 – 390 TETRA 400 Pulse Modulation b) 18Hz 430 – 470 GMRS 460, FM c) 5 kHz FRS 460 deviation 1 kHz sine LTE Band 13, Pulse 704 - 787 Modulation b) 217Hz 80-0075-00-RevK SunTech Vet30 User Manual | 48 2022-July-31...
  • Page 49: Appendix A: Service Screens

    Repairs should be undertaken only by personnel trained or authorized by SunTech Medical. The SunTech Vet30 BP monitor includes a service mode where service information can be accessed, factory default settings can be restored and hardware verification checks can be performed. The service screens are accessed by holding down the power button for approximately 6 seconds when turning the monitor ON.
  • Page 50: Service Information

    There may be a charge associated with the verification if the transducer is not out of calibration. SunTech suggests this check be performed every 2 years or if there is suspicion that the monitor may be out of calibration.
  • Page 51: Air Leak Check

    6 mmHg/min. During manufacturing at SunTech Medical, acceptable air leakage is less than 3 mmHg/min. Both of these pass criteria will not affect the performance or accuracy of the NIBP module so the SunTech Vet30 uses the 6 mmHg/min pass criteria. If you do not have access to this equipment or prefer to have someone else perform the air leak check, the monitor can be sent to SunTech following the procedure described in the Limited Warranty section of this manual.
  • Page 52: Restore Factory Defaults

    If fail, then there is an air leak within the pneumatic system. Try repeating the check with a different cuff making sure all connections are sealed. If it still fails, you may send the monitor and cuffs back to SunTech Medical following the procedure in the Limited Warranty section of the manual.