DE3200 Series Video Microscopes DE3200 Lab Video Microscope DE3200-S2 Bare Fiber + Video Microscope User’s Guide...
Page 2
Connector adapters are compatible with the DE2503 and DE2600 OptiSpec® Microscopes. Within this manual is all the information needed to operate and maintain the DE3200 Series Video Microscopes. Domaille Engineering continues to design innovative new products while maintaining and further developing existing ones.
Centering the illumination ........................10 Cleaning objective lenses ........................10 Adjusting the Magnification ........................10 Cleaning the CCD Camera ........................11 Adjusting the knob tension on the DE3200 Series .................. 11 Troubleshooting Advice ..........................13 LIMITED WARRANTY............................ 15 Product Summary ............................16...
Features: DE3200 DE3200-S2 High resolution video and CCD Camera High resolution video and CCD Camera Five position nosepiece Five position nosepiece Long working distance optics for UPC and APC inspection Long working distance optics for ...
Attaching the Micrometer (DE3200-S2) The micrometer is packaged separately to prevent damage during shipping. To reduce wear on the battery, it has been inserted backwards for shipping. The micrometer nut 1. On the back on the micrometer remove the battery plate. Flip battery and replace battery plate.
Engineering for more information on digital camera options or video frame grabbers. Operation: Focusing and Focus Tension Adjustments The focus adjustment knob on the DE3200 series is conveniently located on each side of the Modular Focus Base. Both coarse and a fine adjustment are provided.
Determining Actual Magnification: The best method to calculate the exact “total” magnification used on the DE3200 series is to measure the cladding on the video display. Take the measurement in millimeters and divide that amount by the 125 micron cladding.
Setting APC Adapters on the DE3200: Use the two end positions on the DE3200 Series stage for setting up APC 2.5 or 1.25mm adapters. A third mounting hole is provided to angle the adapter at 8 degrees. Insert the connector key up on the left hand position and key down on the right hand position.
for 1.25mm and 2.5mm fibers. The Quick Release V-Groove allows the operator to switch between multiple fiber sizes, this adapter allows for fiber sizes from .125mm to 2.66mm. These adapters can be switched out for different applications by following the diagram below: Preventive Maintenance Caution: ...
Centering the illumination Centering the illumination of the DE3200 Series is critical in order to maximize performance of the microscope. All scopes are set up from the factory, but replacing lamps or periodic cleaning of may require the operator to re-center the fiber optic bundle over the vertical illuminator.
Caution: Blow air across and not directly at the CCD sensor of the camera. Adjusting the knob tension on the DE3200 Series The tension on the X-axis adjustment knob may need adjustment from time to time depending on the frequency of use.
Page 12
2. Using a spanner or special adjustment tool available from Domaille Engineering, LLC. Adjust the spanner nut clockwise to increase the tension on the knob. Likewise turn the spanner nut counterclockwise to decrease the tension.
Troubleshooting Advice Image is Not Clear Try adjusting the focus mechanism to bring the image into focus. Increase or decrease the light intensity on the fiber optic light source. Adjust to 5x objective and verify fiber is in the field of view. No Illumination Check the set-up, confirm that all connections are made and that everything is...
Page 14
Make sure the shutter on the CCD is set to the OFF position. Image Drifts In and Out of Focus Tighten the focus tension knob. Tighten the side adjustment screw on the connector bracket. Image is Not Centered Between Objectives Check that each objective is screwed on tight.
LIMITED WARRANTY Domaille Engineering, LLC (“Domaille”) products are warranted by Domaille to be free from defects in workmanship and materials for a period of one-year from the original purchase date. This warranty covers defects in materials or workmanship only and does not include damage due to abuse, misuse, problems with electrical power, problems with compressed air supply, servicing not authorized by Domaille, failure to properly care for and maintain the products, or normal wear and tear.
Product Summary Domaille Engineering’s mission is to provide our customer with the highest quality solution through the innovative use of technology that allows our customers to gain a competitive edge. Our core competencies in engineering and precision manufacturing provide OEM’s and fiber optic customers effective and cost efficient solutions.