Hf generators for operation of ultrasound transducers (22 pages)
Summary of Contents for BANDELIN SONOREX TECHNIK TG Series
Page 1
User instructions High-performance ultrasonic generator series TG for the operation of ultrasonic cleaning tanks (TM), ultrasonic immersible transducers, fl at transducer plates, or reactors TG 3003 TG 1503 valid for: Compact housing TG 1503 and TG 3003 23526b GB/2019-07...
Page 2
Copyright & limit of liability This document may not be reproduced, either in full or in extracts, without prior approval of BANDELIN electronic GmbH & Co. KG, hereinafter referred to as BANDELIN. The German-language original is the binding version of this document. Any differences in the translation are not binding and have no legal effect.
Page 3
Operating sequence Identifi es instructions that must be followed in the described instructions sequence. 23526b GB/2019-07 3 / 25 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
General use ........................16 Maintenance and cleaning ..................17 Cleaning and care ......................17 Warehousing / storage ....................17 Maintenance and repair .....................18 Maintenance .......................18 4 / 25 23526b GB/2019-07 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
Page 5
- not applicable - Taking the unit out of service ..................24 Key words - not applicable - ....................24 23526b GB/2019-07 5 / 25 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
(6), the mains cable (7) and the fan (8) to cool the HF units are located on the rear side. SONOREX TECHNIK TG 1503 SONOREX TECHNIK TG 3003 6 / 25 23526b GB/2019-07 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
The ultrasonic generator is operated from the front and back. Operation is generally conducted on a table or on the wall. 23526b GB/2019-07 7 / 25 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
13 A Mains cable Mains fuse 16 A Degree of protection IP 20 Weight (net) 6 kg 10 kg 8 / 25 23526b GB/2019-07 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
Page 9
It is suitable for use in facilities and areas which are directly connected to a public low- voltage supply network, e.g. medical laboratory facilities. 23526b GB/2019-07 9 / 25 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
• In the case of defects, only authorised skilled personnel or the manufacturer may repair the generator. 10 / 25 23526b GB/2019-07 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
HF cables with plug connectors should be protected against splash water or moisture (protection class: IP 51). 12 / 25 23526b GB/2019-07 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
Verify that the power switch is in the "off" (0) position. Connect the cables (HF cable and mains cable). 23526b GB/2019-07 13 / 25 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
The LED on the switch lights up green if the switch is pressed and shows that the HF unit is ready for operation. 14 / 25 23526b GB/2019-07 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
- If one of the red LEDs lights up or starts fl ickering after start-up, an operating fault has occurred, see chapter 6.2. 23526b GB/2019-07 15 / 25 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
To turn the generator off, press the mains switch on the rear. For additional care instructions, see chapter 5. 16 / 25 23526b GB/2019-07 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
Warehousing / storage During extended periods without use, the generator should be stored in a cool, dry place. 23526b GB/2019-07 17 / 25 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
Preparation: - Switch on ACTIVE on all HF units. (LED should light up green.) 18 / 25 23526b GB/2019-07 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
Page 19
Problems with Transducer frequency Flickers Flickers Flickers the ultrasonic of generator and Consult manufacturer transducers transducer different? 23526b GB/2019-07 19 / 25 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
During the error state, an error signal is emitted at the remote control connection on the rear of the generator (see chapter 2.2.3). 20 / 25 23526b GB/2019-07 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
SONOBLOC tube reactor. During unattended operation, appropriate safety measures should be taken in order to avoid damage caused by dry running. 23526b GB/2019-07 21 / 25 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
The form can be downloaded from the Internet as a PDF fi le: www.bandelin.com - Downloads ... 22 / 25 23526b GB/2019-07 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
The time can be shortened by turning the knob in the opposite direction. 23526b GB/2019-07 23 / 25 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
BANDELIN recommends the TICKOPUR or STAMMOPUR concentrates by DR. H. STAMM GmbH, which have been especially developed for ultrasound use and which optimally utilise the ultrasound.
Page 25
Note: The instructions for use in this and other languages, as well as further information, can be found on the enclosed CD.
Need help?
Do you have a question about the SONOREX TECHNIK TG Series and is the answer not in the manual?
Questions and answers