Chapters 1 and 2 of your VT1413C or Agilent/HP E1313 manual. Identifying the Plug-on You’ll find the VXI Technology part number on the connector side of the SCP to the left of the serial number bar code. For the VT1509A, the part number is : VT1509A.
Connecting To The Terminal Module This section shows how to make connections to the Terminal Module. The SCP connections for the Terminal Modules are shown on the stick-on labels that came with the SCP. Use the appropriate label for the type of Terminal Module you have.
Programming With SCPI Commands The SCPI commands shown here are covered in Chapters 3 and 5 of your VT1413C or Agilent/HP E1313 manual. This section will relate those commands to the parameter values which are specific to this Plug-on. Checking the ID To verify the SCP type(s) installed on the VT1413C or Agilent/HP E1313 use the SYSTem:CTYPe? (@<channel>) command.
Page 8
The channel parameter must specify a single channel. To query the filter state of channel 2 send INP:FILT? (@102) query channel 2 enter statement here Querying the While the VT1509A’s amplifiers have fixed gain, the channel gain can be queried. The response to this query will always be 64. To query any channel Channel Gain to determine its gain setting use the INPut:GAIN? (@<channel>) command.
NOTES 1) When OTD is enabled, the inputs have up to 0.2 µA injected into them. If this current will adversely affect your measurement, but you still want to check for open transducers, you can enable OTD, make a single scan, check the CVT for bad measurements, then disable OTD and make your regular measurement scans.
Page 10
VT1509A Register Map Read (returned value) Write( <regvalue>) SCP Register <regaddr> Value SCP ID (9020 Whole SCP Reg 0 00ppp000000 ppp=Plug-on ccc=SCP channel In addition you will access bits in the Card Control register to control Open Transducer Detection. Checking ID of SCP To query an SCP for its ID value, write the following value to Parameter Register 1: SCP number...
Specifications These specifications for the VT1509A reflect the combined performance of the VT1413C or Agilent/HP E1313 and the VT1509A Signal Conditioning Plug-on. These specifications are not to be added to those presented in the VT1413C or Agilent/HP E1313 User’s Manual. General Specifications Measurement ranges ±3.9 mV to ±0.25 V FS...
Page 12
(90 days) 23°C±1°C (with *CAL? done after 1 hr warm up and CAL:ZERO? within Measurement accuracy 5 min.). If autoranging is ON, add ±0.02% FS to accuracy specifications. DC Volts For Agilent/HP E1313, multiply Noise Spec. by 1.4. Gain 64 Range Linearity Offset Error...
Page 13
(simplified specifications, see temperature accuracy graphs in VT1413C or Agilent/ Measurement accuracy HP E1313 manual for details) Temperature (cont.) Thermocouples (cont.) Type S A/D Filter 0 to 100°C 100 to 200°C 200 to 800 C 800 to 1750°C 2.85°C 1.35°C 0.70°C 0.65°C 2.85°C...