Rotem DRM-3000 Operating Manual

Rotem DRM-3000 Operating Manual

Multi-probe area meter

Advertisement

DRM-3000
Multi-Probe Area Meter
Operating Manual
Ver. 1.11

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the DRM-3000 and is the answer not in the manual?

Questions and answers

Summary of Contents for Rotem DRM-3000

  • Page 1 DRM-3000 Multi-Probe Area Meter Operating Manual Ver. 1.11...
  • Page 2 DRM-3000...
  • Page 3 Under no circumstances can the detectors be connected or disconnected when the Unit is under power. Considering local regulation in force: person and equipment protection against electric shocks, Rotem Industries Ltd. recalls that ANY WORK POWERED EQUIPMENT SHALL MANDATORILY PERFORMED BY QUALIFIED AND AUTHORIZED PERSONNEL.
  • Page 4 Added: 1.02 18/07/2019 1. 8.4 Setting up the Meter 2. IP Cam instructions 1.03 15/08/2019 6.3.1 Changed Power Input Specs Added calibration position of DRM-3000 for internal detector 1.04 03/09/2019 Updated screens Updated Section 7 Updating the DRM- 1.05 10/10/2019...
  • Page 5: Table Of Contents

    Detectors .......................... 12 6.1.1 Connecting the Detectors ..................12 Response Times for the GM Tubes .................. 13 4.1.1 Automatic Range Switching for the DRM-3000 ............ 13 Power ..........................14 6.3.1 Power Consumption .................... 15 The Aux Connector ......................15 The LAN connector ......................17 6.5.1...
  • Page 6 RMVC accessible functions .................... 38 Defining the IP connection ....................38 Real time Monitoring ......................39 9.2.1 Connecting to the DRM-3000 ................39 Downloading the data ....................... 39 Setting up the Meter ......................40 Real time Calibrating ....................... 42 10.1 Setup ..........................
  • Page 7: Introduction

    9001) is needed for reception of telemetry data. 1.2 Field of use The DRM-3000 has been designed to be mounted in a strategic position providing up to four radiation levels simultaneously thus providing the users, a comprehensive overview of the radiological conditions in the work place.
  • Page 8: Drm-3000 Main Advantages

    1.3 DRM-3000 main advantages  Single Meter Supporting 4 detectors – More cost effective  Large Color Display with color coded background– Radiological conditions are clearly shown  TCP/POE Data transmission and power supply – No need for external power sources ...
  • Page 9: Supported Detectors

    3 Supported Detectors The following detectors are supported by the DRM-3000. The maximum cable length between the Meter and Detector is 100 yards. Description Measuring Range 0.5 µSv/h to 10 mSv/h Wide Range Gamma (50 µR/h to 1R/h) 1 µSv/h to 1 Sv/h High Range Gamma (0.1 mR/h to 100 R/h)
  • Page 10: Description

    (optional) 7 inch HD Display High Impact Plastic Casing Touch Panel Internal External Power Buzzer Plug Rear Cover Backup Battery Pack WRM-2 mounting pads (Optional) CPS Stack Board Detector CPU Board 4.2 Connector View Page | 10 DRM-3000 Operating Manual...
  • Page 11: Mounting The Drm-3000

    5.1 Wall Mounting The DRM-3000 is supplied with a wall bracket and 4 Wall mounting screws. Attach the 4 screws to the back cover of the Meter and attach the Wall bracket to the wall taking care that the key slots are facing upwards .
  • Page 12: Operation

    User can set the required Units of Measurement, High (20mA) Range and Low (4mA) Range in the DRM-3000 using the RMVC software. The DRM-3000 is also capable of supporting the full range of Rotem Detectors as well as others like the Ludlum 42-30 Rem Ball Neutron detector, the Ludlum Prescila Neutron Detector and the SFP-100 Alpha Beta detector from V.F.
  • Page 13: Response Times For The Gm Tubes

    The switching between the low and high range GM-Tube is performed at 1,500 mR/h when the radiation field increases (low range detector is disconnected), and at 1,000 mR/h when the radiation field decreases (low range detector is connected). Page | 13 DRM-3000 Operating Manual...
  • Page 14: Power

    3. Internal Battery which provides 24 hours operation at background radiation levels. The DRM-3000 will draw power in the following priority: 1. AC Power, 2. POE and 3. Internal Battery. This means that if all three are connected the Meter will use the AC Power only and will not use the POE or internal battery.
  • Page 15: Power Consumption

    2. 1 x dry contact, N.O. or N.C. dry contact or 24VDC For meters with CPS firmware version 1.19 and above, once powered up the DRM-3000 will activate the solenoid to provide the connection shown below. This upgrade provides the user with an indication on loss of power to the DRM-3000.
  • Page 16 Pin #6 as long as the over threshold condition continues. The DRM-3000 is also capable of supplying an external power source up to 10 µA 100 mV DC or 4A 250 VAC without being connected directly to the AC Power and without the DC power board.
  • Page 17: The Lan Connector

    To create a clean Ethernet or POE installation, we offer a back cover of the DRM-3000 with a hole for the Ethernet cable. This way the DRM-300 can be mounted flush to the wall.
  • Page 18: Installing The Wrm2 Transmitter

    6.6.2 Configuring the WRM2 Module The WRM2 Transmitter should be configured to the following parameters in order to enable it to be compatible to the DRM-3000 6.6.3 Inserting the WRM2 Module Under no circumstances can the WRM-2 transmitter be connected or disconnected to its mounting pads while the Unit is under power.
  • Page 19: Pushbuttons, Icons And Menus

    6.7.1 Background Colors The DRM-3000 provides a number of background colors on the display to help users identify the radiological conditions and reasons for the audible alarms. Each of the four display segments shows its own individual color, while the external LED...
  • Page 20: Pushbuttons

    Penguin shown top left hand corner Open embedded loading Desktop followed by Startup Screen Upper Led flashes green and buzzer sounds and after a few seconds the Display shows the Radiation Data. The Startup Screen Page | 20 DRM-3000 Operating Manual...
  • Page 21: Icons

    6.7.3.2 Powering Down The Meter Press Continuously on the On/Off Button until the Meter turns off 6.7.4 Icons The DRM-3000 displays both the power and communication statuses at all times. 6.7.5 Menus 6.7.5.1 The Menu Structure 1. START SCREEN – Control which ICONS and which screen segments to Display 2.
  • Page 22 3. METER DETAILS – Show the parameters of the Meter 4. DISPLAY OPTIONS – Provides the user with tools to set the display options. 5. Q-SETUP Password required to access the Q-SETUP menu Page | 22 DRM-3000 Operating Manual...
  • Page 23 Confirm Button changes the level of brightness In the case of more than one external detector connected to the DRM-3000, the user can select which of the four can be in focus and displayed in a full screen mode. The Focus on the specific detector is used to enable the user to query and/or modify a parameter.
  • Page 24 Confirm Button and the Sub-Menu will appear 3. METER DETAILS Click on the Up or Down Button to navigate to the required command [METER DETAILS] Click on the Confirm Button and the Sub-Menu will appear Page | 24 DRM-3000 Operating Manual...
  • Page 25 Click on the Confirm Button and the following screen will appear 4.1. Display Dose First click on the Confirm Button adds the display of the Accumulated Dose to the display of the dose rate. Page | 25 DRM-3000 Operating Manual...
  • Page 26 Confirm Button adds the display of the Dose Rate to the display of the accumulated dose displayed. Second click on the Confirm Button displays only the Dose Rate. 4.3. Display a Graph Go back to the DISPLAY OPTIONS Menu and select GRAPH Page | 26 DRM-3000 Operating Manual...
  • Page 27 Up or Down Button to navigate to the required command [Q-SETUP] Click on the Confirm Button and you will need to enter a password to access the Q-Setup Menu. The password is Up, Select, Up, Down and Confirm or The following screen will appear: Page | 27 DRM-3000 Operating Manual...
  • Page 28 Up or Down Button to navigate to the required command WRM (on/off) Each click on the Confirm Button will provide/cut power to the WRM2 module. The WRM2 icon will be displayed whenever power is applied to the module. Page | 28 DRM-3000 Operating Manual...
  • Page 29 The Dose Alarm is set through the WebiSmarts Software and is useful in providing an alarm if the accumulated dose in a room is above the expected. 6.7.5.6 Calibrating the external detectors using the DRM-3000 Touch panel Click on the...
  • Page 30: Communication

    Byte 10 to byte 16 Dose Reading (xxxxx.x) or 88888.8 in case the Dose is disabled Byte 17 = 0x30 Byte 18 to byte 22 rate reading X.YZW * 10 Byte 23 = 0x30 Byte 24 to byte 25 check sum Byte 26 CR Page | 30 DRM-3000 Operating Manual...
  • Page 31 Status Status See Table below for Status 9 Status 3 lsb CS msb CS lsb Status Message - Byte 9 Overflow Threshold Loss of external Low Battery Error with Status Code Alarm power Detector Page | 31 DRM-3000 Operating Manual...
  • Page 32: Updating The Firmware Of The Drm-3000

    Click OK to continue… 1. Click Connect in the main menu: 2. Select the drop down box of the DRM-3000 option, which will allow you to select which specific detector you want to connect to. Select the required detector Page | 32...
  • Page 33 Upon selection of the either of the TCP connection, the Connect Screen updates as follows: The IP address is shown on the Display of the DRM-3000 under the Meter Details heading (page 24). Insert the correct IP address, check that the port is 5000 and click Connect. The...
  • Page 34 Press “Send File” to start the update. Make sure the progress bar is showing progress and the screen of the DPU-3000 displays “UPDATING…”. The Following message will appear when the update is complete: Programming Completed Successfully! Pls switch of and power up to reset Page | 34 DRM-3000 Operating Manual...
  • Page 35: Updating The Cpu Firmware

    Verify that the startup menu reflects the modifications and shows the latest version of firmware. 8.2 Updating the CPU Firmware THE DRM-3000 CPU update package contains a number of files and should be placed in a separate folder on your desktop.
  • Page 36 Next run update_DRM_App.bat by double clicking on the batch file, the following screen will appear: Insert the IP address of the meter again and press ENTER. The following screen appears after a successful update of the firmware: Page | 36 DRM-3000 Operating Manual...
  • Page 37: Replacing The Firmware Sd Card

    1. Open the back lid of the instrument 2. Gently push the cover of the SD card in the direction shown below 3. Take out current SD card and put the new one instead, make sure it fits the slot. Page | 37 DRM-3000 Operating Manual...
  • Page 38: Rmvc Accessible Functions

    IP address according to the MAC address of each DRM-3000. In a direct connection the network adapter provides an Automatic Private IP Addressing (APIPA) to the DRM-3000 each time they are connected and after a period of up to 30 seconds. The range of IP addresses is 169.254.X.X...
  • Page 39: Real Time Monitoring

    9.2 Real time Monitoring 9.2.1 Connecting to the DRM-3000 To connect to the DRM-3000, you should select the following options: DRM-3000 Meter or Int. Detector or Detector #2 or Detector#2 … TCP Connection IP address according to IP address shown on the METER DETAILS screen.
  • Page 40: Setting Up The Meter

    9.4 Setting up the Meter Click Device Setup, the following screen appears: Page | 40 DRM-3000 Operating Manual...
  • Page 41 If a remote camera is required, the URL of the camera is inserted here and the picture will occupy one of the four display slots. The URL address is normally supplied by the manufacturer of the camera i.e., http://admin:admin@xxx.xxx.xxx.xxx/h264 Page | 41 DRM-3000 Operating Manual...
  • Page 42: Real Time Calibrating

    The DRM is positioned in the calibration chamber as shown below and care should be taken to ensure that both GM tubes are situated within the same homogeneous dose rate field. The DRM-3000 is placed in a jig upside down to allow easy accessibility to the power and communication cables.
  • Page 43 Calibration of the external detectors is achieved using a designated USB to Detector cable. The calibration factor is saved in the detector and therefore the calibration of the external detectors can be achieved independently of the DRM-3000 Meter. The calibration procedure for the internal and external wide range detectors is identical to the calibration procedure of the TelePole.
  • Page 44: Setup

    6. Connect the cable to the Computer and Ethernet connector on the Meter. 7. Click Connect in the main menu: 8. Prior to clicking on RMC, ensure that the DRM-3000 is communicating to the software (radiological information is displayed in Meter Rate / Meter Dose and Meter Serial No. is displayed).
  • Page 45: Recommended Calibration Points

    – over range 1500 – 400 mR/h. Higher calibration points may be used if the customer verifies acceptability of high range response in switch over range. 6. The above entered information may be saved as a template to be used for future Page | 45 DRM-3000 Operating Manual...
  • Page 46 17. Once the desired meter dose rate is obtained; click on the associated Calibration Point (e.g., F1) update and then exit. 18. The Cal. Point Value will populate when the update box is checked: Page | 46 DRM-3000 Operating Manual...
  • Page 47 21. Once the Calibration Certificate is started, the user may add a custom Header or Footer to the document prior to printing. Select either Add Company Header or Add Footer, and select the file to be utilized. Page | 47 DRM-3000 Operating Manual...
  • Page 48 Page | 48 DRM-3000 Operating Manual...
  • Page 49 Page | 49 DRM-3000 Operating Manual...