These bead baths require the use of Lab Armor beads to function. The units are not watertight and cannot be used as water or oil baths. Warning: This product contains chemicals, including triglycidyl isocyanurate, known to the State of California to cause cancer as well as birth defects or other reproductive harm. For more information, go to www.P65Warnings.ca.gov.
Page 3
CS-BB20 Bead Bath 100 – 120 Voltage Installation and Operation Manual Part Number (Manual): 4861834 Revision: October 23, 2019 P a g e cascadesciences.com Tel. 503 847-9047...
CERTIFICATIONS This certificate satisfies NRTL safety requirements TÜV SÜD CUE Certificate Number: U8 14 01 64872 030 These units are CUE listed by TÜV SÜD as bead baths for appropriate professional, industrial, or educational use. TÜV SÜD America Inc. is an OSHA recognized NRTL and a Standards Council of Canada accredited certification body.
Page 6
CERTIFICATIONS P a g e cascadesciences.com Tel. 503 847-9047...
UNIT SPECIFICATIONS These bead baths are 100 – 120 volt units. Please refer to the unit data plate for individual electrical specifications. Technical data specified applies to units with standard equipment at an ambient temperature of 25°C and at nominal voltage. The temperatures specified are determined in accordance with factory standard following DIN 12880 respecting the recommended wall clearances of 10% of the height, width, and depth of the inner chamber.
UNIT SPECIFICATIONS EIGHT Unit Weight with Beads Shipping Unit Weight Empty 88.0 lb / 39.9 kg 104 lb / 47 kg* 22.0 lb / 10.0 kg *Shipping weight includes 15 liters of beads. IMENSIONS By Inches Exterior W × D × H Interior W ×...
NGINEERING MPROVEMENTS Cascade Sciences continually improves all of its products. As a result, engineering changes and improvements are made from time to time. Therefore, some changes, modifications, and improvements may not be covered in this manual. If your unit’s operating characteristics or appearance differs from those described in this manual, please contact your Cascade Sciences dealer or customer service representative for assistance.
EADS EQUIRED CS-BB20 bead baths come packaged with 15 liters of aluminum-coated thermal beads. The manufacturer recommends using this full volume of beads to fill the tank to 1” (25 mm) below the top of the bath. This level helps hold product containers in place.
RECEIVING YOUR UNIT NSPECT THE HIPMENT When a unit leaves the factory, safe delivery becomes the responsibility of the carrier. • Damage sustained during transit is not covered by the manufacturing defect warranty. • Save the shipping carton until you are certain that the unit and its accessories function •...
RECEIVING RIENTATION MAGES CS-BB20 Bead Bath Gable Cover Handle Tank Interior Gable Cover Control Panel 12 | P a g e cascadesciences.com Tel. 503 847-9047...
Page 13
RECEIVING Back of Unit Gable Cover Vent Power Cord Inlet with Fuse Holder Data Plate 13 | P a g e cascadesciences.com Tel. 503 847-9047...
RECEIVING ECORDING LATE NFORMATION Record the unit serial number and model number below for future reference. Tech Support needs this information to provide accurate help during support calls and emails. The data plate is located on the back of the unit, next to the power cord inlet. •...
INSTALLATION NSTALLATION ROCEDURE HECKLIST For installing the unit in a new workspace location. Pre-Installation Check that the required ambient conditions for the unit are met, page 16. Check that the spacing clearance requirements are met, page 16. Unit dimensions may be found on page 7. •...
INSTALLATION EQUIRED MBIENT ONDITIONS These units are built for use indoors at room temperatures between 15°C and 30°C (59°F and 86°F) at no greater than 80% Relative Humidity (at 25°C / 77°F). Operating outside these conditions may adversely affect the unit temperature performance. When selecting a location to install the unit, consider all environmental conditions that can adversely impact its temperature performance.
INSTALLATION OWER OURCE EQUIREMENTS When selecting a location for the unit, verify each of the following requirements is satisfied: Power Source The power source must match the power requirements listed on the unit data plate (located on the back of the unit, next to the power cord inlet). The bead bath is intended for 100 –...
INSTALLATION ENERAL OWER AFETY Your unit and its recommended accessories are designed and tested to meet strict safety requirements. It is designed to connect to a power source using the specific power cord type shipped with the unit. For continued safe operation of your unit, always follow basic safety precautions including: Always plug the unit power cord into a protective earth grounded electrical outlet that •...
INSTALLATION EVELING The bead bath is equipped with non-adjustable rubber feet to raise it off the counter and prevent sliding. Ensure that the unit is on a flat and level surface prior to placing the unit in operation. NSTALL THE Install the unit in a workspace location that meets the criteria discussed in the previous entries of the Installation section.
INSTALLATION EIONIZED AND ISTILLED ATER Do not use deionized water to clean the unit or cover, even if DI water is readily available in your laboratory. Use of deionized water may corrode metal surfaces and voids the manufacturing warranty. • The manufacturer recommends the use of distilled water in the resistance range of 50K •...
INSTALLATION SSEMBLE THE OVER Note: The unit ships with the handle installed on the underside of the cover. Uninstall the handle from the underside of the cover. Align the handle with the screw holes on the cover body. Insert the 2 included screws through the underside of the cover body and into the holes on the handle.
Page 22
INSTALLATION 22 | P a g e cascadesciences.com Tel. 503 847-9047...
GRAPHIC SYMBOLS The unit is provided with graphic symbols on its exterior. These identify hazards and adjustable components as well as important notes in the user manual. Symbol Definition Consult the user manual Consulter le manuel d'utilisation Temperature display Indique l'affichage de la température Over Temperature Limit system Thermostat température limite contrôle haute AC Power...
Page 24
SYMBOLS 24 | P a g e cascadesciences.com Tel. 503 847-9047...
CONTROL OVERVIEW Control Panel Over Temperature Protection Control (OTP) This graduated dial sets the mechanical heating cutoff for the Over Temperature Protection system (OTP). The OTP helps prevents unchecked heating of the tank in the event of a hardware failure or external heat spike.
Page 26
CONTROL 26 | P a g e cascadesciences.com Tel. 503 847-9047...
OPERATION HEORY OF PERATION Heating When powered, the bead bath heats to and then maintains an operator-selected temperature setpoint. The unit temperature controller senses the temperature of the beads via a solid-state probe located in the front, right corner of the tank. When the controller detects that the temperature of the beads has dropped below the target setpoint, it pulses power to the heating elements inside the bead bath walls and floor.
OPERATION UT THE ATH INTO PERATION Perform the following steps and procedures to put the unit into operation after installing it in a new workspace environment. Plug in the Power Cord Attach the power cord that came with the unit to the power inlet receptacle on the back of the unit.
OPERATION RMOR EADS Warning: The tank is not watertight. Do not fill the tank with liquid. Use beads only. Avertissement: Le réservoir n'est pas étanche. Ne remplissez pas le réservoir de liquide. Utilisez des perles seulement. Reminder: The beads be should be cleaned and disinfected before filling the tank. See page 36. Always wear gloves when handling beads to avoid contamination.
OPERATION ET THE EMPERATURE ETPOINT Perform the steps below to adjust the setpoint to your process or application temperature. Set the Over Temperature Protection control to its maximum setting, if not already set to max. This prevents the heating cutoff system from •...
OPERATION Note: Test the OTP system at least once per year to verify its functionality. Failure to set the OTP voids the manufacturing defect warranty if over temperature damage occurs. (OTP) ET THE EMPERATURE ROTECTION This procedure sets the mechanical heating cutoff to approximately 1˚C above the current bead temperature.
OPERATION RODUCT Load product after the bead bath must run with beads in the tank for at least 8 hours prior to loading product. Briefly stir the beads before loading product. This helps ensure temperature uniformity. Always use a stirring rod to avoid contamination. Avoid using a metal rod to prevent •...
OPERATION ALIDATING PPLICATION ROTOCOLS All water bath application protocols should be validated when switching to bead baths. Applications involving large or frozen product containers may take 2 – 3 times longer to achieve the target setpoint than in a traditional water bath due to the high thermal stability of the beads. Bringing the temperature of frozen product up to 4°C before placing in the bead bath can help optimize heat up times.
Page 34
OPERATION 34 | P a g e cascadesciences.com Tel. 503 847-9047...
USER MAINTENANCE Warning: Disconnect this unit from its power supply prior to performing maintenance or services. Avertissement: Débranchez cet appareil de son alimentation électrique avant d'effectuer la maintenance ou les services. LEANING AND ISINFECTING If a hazardous material or substance has spilled in the tank, immediately initiate your site’s Hazardous Material Spill Containment protocol.
MAINTENANCE Disinfecting the Bath When disinfecting the unit, keep the following in mind: Always turn off and disconnect the unit to safeguard against electrical hazards. • For maximum effectiveness, disinfection procedures are typically performed after cleaning. • Beads must be removed from the tank to successfully disinfect the bath. •...
MAINTENANCE Disinfecting the Beads Remove beads from the tank. 2. Disinfect the beads with a solution of isopropanol (IPA) or 70% ethanol, ensuring that all bead surfaces have been sufficiently coated. 3. The manufacturer recommends allowing the IPA or ethanol solution to air dry. This helps ensure that the beads have had sufficient contact time to be thoroughly disinfected.
The device must be accurate to at least 0.1°C and should be regularly calibrated by a third party. Never use alcohol or mercury-based thermometers. Cascade Sciences bead baths do not normally require calibration. Should your SOP or Quality program require calibrations, follow this guideline.
Page 39
MAINTENANCE 5. The bead temperature must be stable in order to perform an accurate calibration. The bead bath must run loaded with beads for at least 8 hours prior to conducting a • calibration. The temperature is considered stabilized when the filled bead bath has operated with •...
Page 40
MAINTENANCE Suggested Calibration Continued Place the display in temperature calibration mode. a. Press and hold both the Up and Down arrow buttons simultaneously for approximately 5 seconds. b. Release the buttons when the temperature display shows the letters “C O”. The display will begin flashing the current temperature display value.
Page 41
MAINTENANCE Suggested Calibration Continued Compare the reference device reading with the unit temperature Reference Device display again. If the reference device and the unit temperature display • readings are the same, or the difference falls within the range of your protocol, the unit is now calibrated for temperature. -OR- ...
Page 42
MAINTENANCE 42 | P a g e cascadesciences.com Tel. 503 847-9047...
PARTS LIST Description Parts Number Description Parts Number Fuse, 6.3 amp 250V Gable Cover 9751202 3300515 Power Cord, 125 volt, Lab Armor Beads, 15Amp, 9 ft 5 in 5 Liters (2.86m), NEMA 5-15P 1800510 42370-005 43 | P a g e cascadesciences.com Tel.