Table of contents LB 8110 Shield Table of Contents About this operating manual ................5 Some prior remarks ...................... 5 Storage .......................... 5 Target Group ........................ 5 Validity of the Manual ....................6 Structure of the Manual....................6 Copyrights ........................6 Representation ......................
1 About this operating manual About this operating manual Some prior remarks The product is handed over to you by the manufacturer BERTHOLD TECHNOLOGIES GmbH & Co. KG in a complete and functionally reliable condition. This operating manual illustrates how to: •...
Version and release date of this operating manual can be found in the bottom of each page. An alteration service is not provided by the manufacturer BERTHOLD TECHNOLOGIES GmbH & Co. KG. The manufacturer reserves the right to make changes to this operating manual at any time without stating reasons.
LB 8110 Shield 1 About this operating manual Symbols Used 1.8.1 Structure of Warnings j Signal word Source and consequence Explanation, if required Measure In case of emergency • Warning symbols: (warning triangle) draws attention to the hazard • Signal word: states the severity of the hazard •...
1 About this operating manual LB 8110 Shield IMPORTANT Sections marked with this symbol point out important information on the product or on handling the product. Provides tips on application and other useful information. 1.8.2 Symbols used on the device Ionising radiation The shield is equipped with a radioactive source.
• Restructuring or changing the system components. • Manipulation or avoidance of existing safety equipment. BERTHOLD TECHNOLOGIES GmbH & Co. KG shall only accept liability for / guarantee the correspondence of the product to its publicised specifications. If the product is used in a way which is not described in this manual, the product's protection is compromised and the warranty becomes void.
2 Safety LB 8110 Shield Ambient conditions during operation and storage The shield was specifically designed for use in rough ambient conditions. The compliance with the operating conditions specified below contributes to guaranteeing the permanent functionality of the shield and the prevention of damage.
Page 11
BERTHOLD TECHNOLOGIES for particular tasks. When dealing with radioactive materials, a radiation safety officer must also be consulted. Radiation Safety Officer In order to ensure proper handling and compliance with the statutory requirements, the company has to appoint a radiation safety officer in accordance with the applicable national law (in Germany: Strahlenschutzverordnung [German radiation protection regulation]).
2 Safety LB 8110 Shield Radiation Protection 2.4.1 Basic principles and regulations The amount of radiation absorbed by the body (exposure to radiation) is determined by three parameters from which the basic radiation protection regulations can be derived: Distance If work close to equipment containing radioactive substances is required, the largest distance possible is to be kept.
In the case of fire, these measures limit the access to this area. Avoid the incorporation by keeping sufficient distance. Notify BERTHOLD TECHNOLOGIES of the situation; you will promptly receive information on immediate measures. 36682BA2 Rev. 00, 07/2018...
2 Safety LB 8110 Shield Operator's obligations The operator of the product must regularly train his personnel in the following topics: • Observation and use of the operating manual and the legal provisions. • Proper use of the product. • Observation of the plant security instructions and the operating instructions of...
The LB 8110 shield is intended as shield/protective container for radioactive rod sources and may only be used for this purpose. The radioactive substance is contained in a leak-proof welded source capsule and installed in the LB 8110 shield. Apart from source and shield, additional system components such as detectors and evaluation units are required for a complete measuring system.
4 Installation LB 8110 Shield Installation Safety Instructions j WARNING Danger of injury by falling loads Never stand underneath a lifted or suspended load, keep at a safe distance. Only use tested sling gear components appropriate for the transport ...
LB 8110 Shield 4 Installation Packaging The shield with radioactive source fulfils the regulations for the transport of radioactive substances UN 2910 – excepted package – with a maximum activity of < 10 GBq Am-241 acc. IAEA SSR-6. Intermediate storage of the source...
4 Installation LB 8110 Shield Installation of the shield To avoid unnecessary exposure to radiation, install the shield as final system component. The shield is placed on a mounting base at the measuring point and fixed. The shield on the container is secured against tilting by means of a tipping protection.
LB 8110 Shield 4 Installation Marking The warning sign indicates the beginning of the control area as long as the control area is outside the shield. If the control area is within the shield, then the radiation warning sign already applied to the shield is sufficient.
5 Usage LB 8110 Shield Usage Safety instructions j CAUTION Danger caused by nuclear radiation Shields usually contain radioactive sources. An increased exposure to radiation may lead to damage to health. Consult the radiation safety officer responsible for your company.
Page 23
LB 8110 Shield 5 Usage Padlock locking cylinder Bolt Lever rod Fig. 4 Opening the locking mechanism Opening the locking mechanism: If applicable, remove the padlock (Fig. 4, item 1). Remove the bolt (Fig. 4, item 2). Turn the locking cylinder to the OPEN position using the lever rod (Fig. 4, item 3).
BERTHOLD TECHNOLOGIES GmbH & Co KG before the system is used again. Only spare parts specified by BERTHOLD TECHNOLOGIES GmbH & Co. KG may be used and required repair work may be carried out exclusively by authorized persons.
LB 8110 Shield 6 Maintenance and Repair Visual inspection The visual inspection must be carried out by persons who are at least employees with general knowledge at the following intervals: • before initial commissioning • with every repair that may be required •...
Page 26
6 Maintenance and Repair LB 8110 Shield • If the source is installed in a device, a drawing is to be provided. In the drawing, the position of the source and of all parts intended to protect the source against external influences must be clearly indicated. Recommendations regarding the most appropriate test method should be given, e.g.
LB 8110 Shield 6 Maintenance and Repair Checking the locking mechanism The functional test of the locking mechanism must be carried out by a person who is at least an employee with general knowledge. We recommend conducting the test of the locking mechanism at the same intervals as the visual inspection if the operating condition of the system allows.
6 Maintenance and Repair LB 8110 Shield Source replacement When to replace the source Generally, the radioactive source used enables a service life of 15 years. A replacement of the source is only required if the statistical fluctuations of the output signal are inadmissibly large and compensation by increasing the time constant is no longer possible, e.g.
LB 8110 Shield 7 Decommissioning Decommissioning Safety instructions j WARNING Danger of injury by falling loads Never stand underneath a lifted or suspended load, keep at a safe distance. Only use tested sling gear components appropriate for the transport ...
7 Decommissioning LB 8110 Shield Disassemble the shield j WARNING Danger of injury by falling loads Falling loads can endanger people. Never stand underneath a lifted or suspended load, keep at a safe distance. Cordon off the danger zone with barrier tape.
LB 8110 Shield 7 Decommissioning Shipment of radioactive substances If you wish to dispatch radioactive material, you must observe the international regulations for the transport of dangerous goods and for radiation protection as well as any applicable national regulations. It is the full responsibility of the sender to comply with these regulations.
Page 32
BERTHOLD TECHNOLOGIES GmbH & Co. KG will not assume any costs for customs clearance or transportation. • BERTHOLD TECHNOLOGIES GmbH & Co. KG is to be informed in advance about any return transport. BERTHOLD TECHNOLOGIES GmbH & Co. KG reserves the right to reject any radioactive substances sent to Berthold without prior notice.
LB 8110 Shield 8 Appendices Appendices Check lists for tests Visual inspection Requirement not met The shield must be free from damage. The type plates must be legible. If there is corrosion, it must not affect the function and stability of the shield.
Page 34
8 Appendices LB 8110 Shield Checking the functions of the locking mechanism Requirement not met It must be possible to move the locking mechanism completely back and forth between the two positions (OPEN / CLOSED). The measuring signal of the...