Thank you for purchasing a Heidolph Instruments product. This item has been designed, made and inspected in compliance with DIN EN ISO 61010 for long performance and continuous operation. SUMMARY SUMMARY ........................12 STANDARD ITEM PLUS ACCESSORIES................13 GENERAL .........................13 SAFETY INFORMATION ....................13 SET-UP..........................15...
STANDARD ITEM PLUS ACCESSORIES item qty. 230/240V 50/60Hz 115V 50/60Hz heating bath 4000 digit 517-01002-00 517-01002-01 loose parts: power cord 14-007-003-81 14-300-009-29 link wire to heating bath 14-007-045-08 14-300-009-28 Owner's Manual 01-005-004-26 01-005-004-26 GENERAL Unpack your item carefully. Inspect for damage and report such damage or missing parts to your supplier immediately.
Page 4
60°C. Avoid touching hot edge of bowl, evaporator flask and fluid in bowl. Use gloves when replacing evaporator flask. Never run hot bath dry. Repair work is limited to skilled personnel so authorized by Heidolph. Your item needs a solid stand.
SET-UP 1. Item set-up Locate hot bath on a smooth and solid surface, Item is set-up horizontally. Controls facing forward. All of the four feet of the hot bath feature height adjustment, to prevent tipping. Re-adjust feet as required. 2. Electric wiring (fig. 1) Before connecting with your local power supply, double check to be sure: your item has been designed for your local voltage as depicted on the data plate (also refer to "Safety Information"...
Hazard of burns when running hot bath (2) at temperatures of more than 60° C. Beware of metal edge of hot bath, fluid and evaporator flask. fig. 1 4. Adjust hot bath temperature Turn hot bath ON with power switch (2a); button lighted green. Select heating bath temperature with rotary knob (2b).
The item is maintenance-free. Repair work is limited to engineers authorized to do so by HEIDOLPH. So therefore get in touch with your local HEIDOLPH dealer or field respresentative. DISASSEMBLE, FORWARD & STORE Disassemble Turn the item OFF and disconnect mains plug.
Error code Er3 displayed: you are heating your hot bath dry ! Disconnect push button (2a), check whether temperature limiter dropped, reset if required, fill heating bath with fluid. f) Error code Er1 and Er2 displayed: heating bath inspection by a service engineer authorized by Heidolph is mandatory. SPECIFICATIONS heating power...
If any aspect of installation, operation or maintenance remains unanswered in the present Manual, please contact the following address: For repair services please call Heidolph Instruments (phone: +49 - 9122 - 9920-68) or your local, authorized Heidolph Instruments Dealer. Note You will receive approval for sending your defective item to the following address: Heidolph Instruments GmbH &...