Lactat Photometer plus
DP 110
Operating Manual
Version 5.9 / 5.9 SI
Edition 2021-01
2021-01

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Lactat Photometer plus DP 110 and is the answer not in the manual?

Questions and answers

Summary of Contents for Diaglobal Lactat Photometer plus DP 110

  • Page 1 Lactat Photometer plus DP 110 Operating Manual Version 5.9 / 5.9 SI Edition 2021-01 2021-01...
  • Page 2 Dear customer, We are pleased that you have chosen the Lactat Photometer plus from Diaglobal GmbH and thank you for the confidence you have placed in us. The Lactat Photometer plus belongs to a new generation of small mobile devices developed by Diaglobal GmbH and specially designed for on-site analysis.
  • Page 3: Table Of Contents

    Table of contents Page General information regarding the Photometer Installation Description of the device Power supply 3.1.1 Mains power operation 3.1.2 Mains-independent operation Measuring system Service Adjustment and Calibration Maintenance Cleaning Instructions Malfunctions Disposal Required reagents and laboratory accessories Expiration date of consumables Reagents / parameter list Control materials Laboratory aids...
  • Page 4: General Information Regarding The Photometer

    The Lactat Photometer plus fulfils the basic requirements of Appendix I of Directive 98/79/EC regarding in-vitro diagnostics. The conformity of the device with Directive 98/79/EC is confirmed by the use of the CE marking. Manufacturer: Diaglobal GmbH Innovationspark Wuhlheide Köpenicker Str. 325 / Haus 41 12555 Berlin Tel:...
  • Page 5: Installation

    Installation For trouble-free operation of the device, the following environmental conditions must be met: • Ambient temperature: 0 °C ... 40 °C • No direct exposure to sunlight or similar sources of radiant heat • Free from excessive dust • Free from vibrations •...
  • Page 6: Power Supply

    The Photometer is supplied with a power supply unit for operation on a mains voltage in the range of 100 V ... 240 V AC. The mains plug is marked with a Diaglobal logo (sticker). The connector plug of the power supply unit is connected to the power supply socket on the back of the device.
  • Page 7: Service

    Non-functional units should be sent to our Berlin address. We will provide a loan device for the duration of the repair. Disposal Diaglobal GmbH will take back and dispose of units that are no longer needed or cannot be repaired, free of charge. 2021-01...
  • Page 8: Required Reagents And Laboratory Accessories

    Powder-free nitrile gloves size M LH 035 Safety Lancets Extra, orange 1.8 mm All reagent kits, control materials and other materials are supplied by Diaglobal GmbH and can be stored and transported together with the Lactat Photometer plus in a practical case.
  • Page 9: Quality Control According To The Guideline Of The German Medical Association

    (calibration) with subsequent documentation of the measured value. The corresponding protocol forms are available from Diaglobal free of charge. We recommend using the Diaglobal control solutions LAC QS and GLU QS to check the accuracy of lactate and glucose determinations In agreement with the requirements of the German Medical Association, a test of the device function (see operating instructions, chapter 8.5) is integrated in the...
  • Page 10: Measuring Procedure

    Measuring process Multi-point measurement with consideration of the sample blank value and recognition of the endpoint After measuring the sample blank value (=measurement 1) the colour reaction in the cuvette is started. The reaction process is monitored by the device (=measurement 2).
  • Page 11: Measurement

    If it becomes obvious during the test that one of the device functions does not correspond to the required settings, <SERVICE> will appear in the display. In this case, switch the device off. Please call Diaglobal GmbH service (Tel. +49 (0) 30 6576 2597) or contact your specialist retailer. Test selection Press the <ON/ENTER>...
  • Page 12: Notes On Taking Samples And Carrying Out Measurements

    Measurement range controls The measurement ranges of all measurement results shown in the display are verified by an integrated measurement range control. If the measurement range is exceeded, an error is displayed. The measurement ranges that are separately defined for each parameter are documented on the respective package inserts as well as in this operating manual, chapter 9, Technical Data.
  • Page 13 If the sample is below the black ring mark, this will lead to incorrect negative measurement results. In this case, correction is hardly possible due to the air bubble that will form when trying to collect more blood. 7. Before the capillary is placed in the cuvette, the lower area must be carefully wiped on the outside with a cellulose swab to remove sample particles attached to the capillary.
  • Page 14: Technical Data

    Technical data Storage temperature: -20 °C ... 70 °C Operating temperature: 0 °C ... 40 °C Dimensions: 200 x 100 x 50 mm Weight: 450 g Measuring principle: Absorption measurement with single beam photometer (Fig. 2), chopped operation Projector: Spectral apparatus: Interference filter Measuring wavelength: 520 nm...
  • Page 15: General Guidelines And Notes

    General Guidelines and Notes EC Directives 1. Directive 98/79/EC on in-vitro diagnostic devices EN / ISO standards 2. EN ISO 9001:1994, Quality Management Systems, Model for quality assurance in design, development, production, installation and customer service 3. EN ISO 13485, Medical devices, Requirements for regulatory purposes (application of EN ISO 9001) 4.
  • Page 16 In this case, please contact our customer service at +49 (0) 30 6576 2597. Diaglobal GmbH · Köpenicker Straße 325 · 12555 Berlin · +49 (0)30 6576 2597 · info@diaglobal.de · www.diaglobal.de 2021-01...
  • Page 17 9. Time is displayed, wait for 7. Turn cuvette upside down 8. First press ON/ENTER several times measured value Then insert cuvette into photometer Diaglobal GmbH · Köpenicker Straße 325 · 12555 Berlin · +49 (0)30 6576 2597 · info@diaglobal.de · www.diaglobal.de 2021-01...
  • Page 18 Note: Ensure the correct or- Time is displayed, wait for measured value, remove der of the samples! measured value cuvette, and so on Diaglobal GmbH · Köpenicker Straße 325 · 12555 Berlin · +49 (0)30 6576 2597 · info@diaglobal.de · www.diaglobal.de 2021-01...