Summary of Contents for Diaglobal Duo Photometer DP 200
Page 1
Duo Photometer DP 200 Operating Manual Version 5.4 / 5.7 SI Edition 2021-01 2021-01...
Page 2
Dear customer, We are pleased that you have chosen the Duo Photometer from Diaglobal GmbH and thank you for the confidence you have placed in us. The Duo Photometer belongs to a new generation of small mobile devices developed by Diaglobal GmbH and specially designed for on-site analysis.
Table of contents Page General information regarding the Photometer Installation Description of the device Power supply 3.1.1 Mains power operation 3.1.2 Mains-independent operation Measuring system Service Adjustment and Calibration Maintenance Cleaning Instructions Malfunctions Disposal Required reagents and laboratory accessories Expiration date of consumables Reagents / parameter list Control materials Laboratory aids...
The Duo Photometer fulfils the basic requirements of Appendix I of Directive 98/79/EC regarding in-vitro diagnostics. The conformity of the device with Directive 98/79/EC is confirmed by the use of the CE marking. Manufacturer: Diaglobal GmbH Innovationspark Wuhlheide Köpenicker Str. 325 / Haus 41 12555 Berlin Tel:...
Installation For trouble-free operation of the device, the following environmental conditions must be met: • Ambient temperature: 0 °C ... 40 °C • No direct exposure to sunlight or similar sources of radiant heat • Free from excessive dust • Free from vibrations •...
The Photometer is supplied with a power supply unit for operation on a mains voltage in the range of 100 V ... 240 V AC. The mains plug is marked with a Diaglobal logo (sticker). The connector plug of the power supply unit is connected to the power supply socket on the back of the device.
Non-functional units should be sent to our Berlin address. We will provide a loan device for the duration of the repair. Disposal Diaglobal GmbH will take back and dispose of units that are no longer needed or cannot be repaired, free of charge. 2021-01...
Powder-free nitrile gloves size M LH 035 Safety Lancets Extra, orange 1.8 mm All reagent kits, control materials and other materials are supplied by Diaglobal GmbH and can be stored and transported together with the Duo Photometer in a practical case.
Internal quality assurance is carried out in the form of a weekly accuracy check (calibration) with subsequent documentation of the measured value. The corresponding protocol forms are available from Diaglobal free of charge. We recommend using the blood control HEM QS and ERY QS with target values in the normal concentration area for checking the accuracy of determinations of haemoglobin, haematocrit and erythrocyte counts.
If it becomes obvious during the test that one of the device functions does not correspond to the required settings, <SERVICE> will appear in the display. In this case, switch the device off. Please call Diaglobal GmbH service (Tel. +49 (0) 30 6576 2597) or contact your specialist retailer. Test selection Press the <ON/ENTER>...
Measurement range controls The measurement ranges of all measurement results shown in the display are verified by an integrated measurement range control. If the measurement range is exceeded, an error is displayed. The measurement ranges that are separately defined for each parameter are documented on the respective package inserts as well as in this operating manual, chapter 9, Technical Data.
Page 12
6. Before the capillary is placed in the cuvette, the lower area must be carefully wiped on the outside with a cellulose swab to remove sample particles attached to the capillary. Otherwise, this would lead to incorrect positive measurement results. 7.
Technical data Storage temperature: -20 °C ... 70 °C Operating temperature: 0 °C ... 40 °C Dimensions: 200 x 100 x 50 mm Weight: 450 g Measuring principle: Absorption measurement with single beam photometer (Fig. 2), chopped operation Projector: Spectral apparatus: Interference filter Measuring wavelengths: 546 nm...
General Guidelines and Notes EC Directives 1. Directive 98/79/EC on in-vitro diagnostic devices EN / ISO standards 2. EN ISO 9001:1994, Quality Management Systems, Model for quality assurance in design, development, production, installation and customer service 3. EN ISO 13485, Medical devices, Requirements for regulatory purposes (application of EN ISO 9001) 4.
Page 15
In this case, please contact our customer service at +49 (0) 30 6576 2597. Diaglobal GmbH · Köpenicker Straße 325 · 12555 Berlin · +49 (0)30 6576 2597 · info@diaglobal.de · www.diaglobal.de 2021-01...
Page 16
In regard to series sample (Fig. 3) into photometer measurement: Read measured value After zero point setting any number additional samples can be measured Diaglobal GmbH · Köpenicker Straße 325 · 12555 Berlin · +49 (0)30 6576 2597 · info@diaglobal.de · www.diaglobal.de 2021-01...