Page 1
Model 136A Tourmaline hydraulic pressure calibration transfer standard, 20k psi, 0.2 Installation and Operating Manual For assistance with the operation of this product, contact PCB Piezotronics, Inc. Toll-free: 800-828-8840 24-hour SensorLine: 716-684-0001 Fax: 716-684-0987 E-mail: info@pcb.com Web: www.pcb.com...
Page 2
Returning Equipment – Following the equipment to PCB Piezotronics for these procedures will insure that your repair. User servicing or repair is not returned materials are handled in the recommended and, if attempted, may most expedient manner.
Page 3
PCB Piezotronics, Inc. Warranty – All equipment and repair 3425 Walden Ave. services provided by PCB Piezotronics, Depew, NY 14043 USA Inc. are covered by a limited warranty Toll-free: (800) 828-8840 against defective...
PCB Model 136A Tourmaline Hydraulic Dynamic Pressure Calibration Transfer Standard NOTE: The oil in the hydraulic system must be DESCRIPTION: non-conductive because the Model 136A has an exposed crystal. Silicon oil is The model 136A Hydraulic Reference Sensor is recommended.
Page 5
Model Number Revision: E CHARGE OUTPUT PRESSURE SENSOR 136A ECN #: 40665 Performance ENGLISH OPTIONAL VERSIONS Sensitivity(± 15 %) 0.2 pC/psi 0.029 pC/kPa Optional versions have identical specifications and accessories as listed for the standard model Measurement Range 15 kpsi 103,400 kPa except where noted below.