Malmet B105BT Operation, Maintenance And Installation Manual

Blanket warming cabinet

Advertisement

Quick Links

Blanket Warming Cabinet
B105BT; B210FS; B210C; B210WB; B210S; B260FS; B260C; B260WB; B260S; B420FS; B420C
BB105FS; FB105FS; BS105FS; BB210FS; FB210FS; BS210FS; BB210C; FB210C; BS210C;
Operation, Maintenance and Installation
design and technical specifications are subject to change without notice.
Serial Number:
Date Installed:
It is important that the name from whom you purchased your device and the name of the installer are recorded above. The installer is responsible for the correct
installation, start up and demonstrating the operation of this device. They are also responsible for issuing relevant certificates of compliance (these may differ from
state to state).
Issue 20
Single Models:
Combination Models:
BB260FS; FB260FS; BS260FS; BB260C; FB260C; BS260C
Manual
Note: Due to Malmet's Policy of continuous product improvement;
Supplied to:
Installed by:
Head Office and Customer Service
ABN 95 001 717 791
9-11 McKay Avenue
PO Box 373
Leeton NSW 2705
Phone: +61 2 6953 7677
Email: info@malmet.com.au
Jun-23

Advertisement

Table of Contents
loading

Summary of Contents for Malmet B105BT

  • Page 1 Phone: +61 2 6953 7677 Email: info@malmet.com.au Blanket Warming Cabinet Single Models: B105BT; B210FS; B210C; B210WB; B210S; B260FS; B260C; B260WB; B260S; B420FS; B420C Combination Models: BB105FS; FB105FS; BS105FS; BB210FS; FB210FS; BS210FS; BB210C; FB210C; BS210C; BB260FS; FB260FS; BS260FS; BB260C; FB260C; BS260C...
  • Page 2: Table Of Contents

    Foreword ..................................2 Certifications and Compliances ............................2 Quality Policy................................. 2 Important Warranty Reminder ............................2 Malmet Head Office and Factory Contact Details ......................2 Safety Instructions - Warnings ............................3 1.0 Design Parameters ..............................4 1.1 Device Operation ............................... 5 1.2 Control Display Features ............................
  • Page 3: Foreword

    Operation, Maintenance and Installation Manual Foreword To obtain maximum life and efficiency from your Malmet Warming Cabinet and to ensure safe operation, please read this manual thoroughly and follow all instructions before operating the device. This manual provides information on the operation of the device. It is recommended that all persons operating the device have access to this manual for training purposes.
  • Page 4: Safety Instructions - Warnings

    Malmet Blanket Warming Cabinet Operation, Maintenance and Installation Manual Safety Instructions - Warnings Please read and understand this manual before using this device, if this device is used in a manner not specified by the manufacturer protection by the device may be impaired.
  • Page 5: Design Parameters

    Operation, Maintenance and Installation Manual 1.0 Design Parameters Single 105 Litre Blanket Warming Cabinet (Model B105BT) is designed to hold 4 blankets maximum; 2 per shelf maximum (depending on type of blanket). Both shelves are adjustable. Single 210 Litre Blanket Warming Cabinets (Models B210FS, B210C, B210WB, B210S) are designed to hold 16 blankets maximum;...
  • Page 6: Device Operation

    Malmet Blanket Warming Cabinet Operation, Maintenance and Installation Manual 1.1 Device Operation NOTE The Blanket Warming Cabinet is factory set to 60°C Before starting the device The device should be run initially on a power supply not protected by an earth leakage circuit breaker for approximately three hours.
  • Page 7: Changing The Set Temperature (Lock Out Feature)

    Malmet Blanket Warming Cabinet Operation, Maintenance and Installation Manual 1.4 Changing the set temperature (lock out feature) AUTHORISED PERSONNEL ONLY Authorised operators can change the temperature by holding the up and down temperature buttons at the same time for a total of 5 seconds. The temperature can then be adjusted. The setting will revert back to tamper proof 5 seconds after the temperature is adjusted or if the up and down temperature buttons are not pressed within 5 seconds.
  • Page 8: Installation

    Malmet Blanket Warming Cabinet Operation, Maintenance and Installation Manual 2.0 Installation Floor Bench Skirt Floor Bench Skirt is factory fitted to the cabinet. To level, remove the four plastic plugs from the inside floor and adjust the feet by turning the slotted screw head. Once the device is level, replace the plugs.
  • Page 9 Malmet Blanket Warming Cabinet Operation, Maintenance and Installation Manual Warming Cabinets with Castors All warming cabinets should be located on a level floor surface and should never be operated on a sloping surface. Check that all swivel wheels are in lock position.
  • Page 10: Service Connections

    Malmet Blanket Warming Cabinet Operation, Maintenance and Installation Manual 2.1 Service Connections Electrical Connection – 240 Volt 10 Amp Plug into 10 Amp G.P.O. 2.2 Device Dimension and Clearances B105BT Blanket Warming Cabinet Bench Top Issue 20 Page 9 Jun-23...
  • Page 11 Malmet Blanket Warming Cabinet Operation, Maintenance and Installation Manual B210FS Blanket Warming Cabinet Freestanding with Floor Skirt Issue 20 Page 10 Jun-23...
  • Page 12 Malmet Blanket Warming Cabinet Operation, Maintenance and Installation Manual B260C Blanket Warming Cabinet Freestanding on Castors Issue 20 Page 11 Jun-23...
  • Page 13 Malmet Blanket Warming Cabinet Operation, Maintenance and Installation Manual B420FS Blanket Warming Cabinet Freestanding with Floor Skirt Note: This Blanket Warming Cabinet is not intended for use by persons (including children) with reduced physical, sensory, or mental capabilities, or lack of experience and knowledge.
  • Page 14: Maintenance

    Devices should have regular Test and Tag procedures undertaken. 3.2 Replacement of Safety Devices The following safety devices must be replaced and tested by a Malmet trained technician, failure to do so may impair the protection by the device. • Over Temperature Cut Out Switches •...
  • Page 15: Trouble Shooting Guide

    Malmet Blanket Warming Cabinet Operation, Maintenance and Installation Manual 3.3 Trouble Shooting Guide Problem Probable Cause Suggested Remedy Display not on or won’t turn on. No Mains power supply. Check power supply and lead plugged in. Power point not turned on.
  • Page 16: Technical Specifications

    Malmet Blanket Warming Cabinet Operation, Maintenance and Installation Manual 4.0 Technical Specifications 4.1 Device Specifications 105 Lt Models: 4 maximum load 210 Lt Models: 16 maximum load Blankets Capacity (Single) 260 Lt Models: 18 maximum load 420 Lt Models: 30 maximum load...
  • Page 17: Wiring Diagram

    Malmet Blanket Warming Cabinet Operation, Maintenance and Installation Manual 4.2 Wiring Diagram Issue 20 Page 16 Jun-23...
  • Page 18: Warranty Statement

    (and paid for) by the owner. • It is the owner's responsibility to provide safe access to the device. Malmet, or any of its authorised service agents, may refuse to perform maintenance or warranty work if access is unsafe, as determined by Malmet or any of its authorised service agents acting reasonably.
  • Page 19 • To the extent permitted by law, no responsibility will be accepted for damage incurred as a result of, or incidental to, electrical surges or brown outs or for any other consequential damages. • If there is no certificate of compliance for plumbing or electrical, Malmet reserves the right to refuse service on non-compliant installations.
  • Page 20 HOW TO MAKE A CLAIM UNDER THIS WARRANTY If you believe there is a defect in a device you have purchased from Malmet, you must notify Malmet in writing of such defect, by sending an email (Notice of Defect) to info@malmet.com.au...
  • Page 21 Malmet Blanket Warming Cabinet Operation, Maintenance and Installation Manual Manufactured by Malmet (Australia) Pty Ltd ABN 95 001 717 791 www.malmet.com.au DCN:2969 P/N 8-M00-83-900...

Table of Contents