길 우 트 레 이 딩 02)2636-0009(대) 팩스02)2636-4753 giltron@naver.com www.gilwoo.co.kr 사 용 설 명 서- 영문 MULTI-FUNCTION DATALOGGER WITH PRINTER 9851, Temp./RH logging printer...
INTRODUCTION Thank you for purchasing this multiple function printer! This unit has been developed to meet your max.satisfaction by its user- friendly design. Please review the entire manual for a complete overview of how to operate this meter. MATERIAL SUPPLIED Check for damaged or missing parts in your meter before starting.
COMPARISON TABLE Measured parameters of each meter: PARAMETERS FUNCTION MODEL Temperature, DP Hygro-thermometer 9851 Humidity, Wet Bulb Please refer to below for the difference of each measurement mode. MEMORY POINT MEASUREMENT DESCRIPTION Single Point Single point No Memory Measurement measurement...
ACCESSORY The temperature and humidity sensors are built in the probe. Please keep the small piece of noted enclosed with each probe for calibration purpose. TOP COVER -CONNECTOR The following illustration is the connectors for 9851.
METER KEYPAD-WITH PRINTER Top Cover Based on different functions, the top cover connector and color are different. It is dark green for 9851. Welcome MEAS Corresponding Function keys Keypad SYM. stands for"1,*,.,:,$,+,-,=" stands for"2,a,b,c,A,B,C" PQRS WXYZ stands for"3,d,e,f,D,E,F" stands for"4,g,h,i,G,H,I"...
METER SETTING Before measuring, please check the meter setting first to confirm the basic settings of the meter are what you need. To enter each command, just press the corresponding F1 to F4 key. Welcome MEAS F4 :Meter Setting Pressing key to shift the cursor.
SINGLE MEASUREMENT There are three measurement modes : 1. Single measurement(MEAS) 2. Multiple measurement(MEM) and 3. Auto logging (LOG). Operating the single measurement as a general meter. MEAS F1 :Single Measurement Welcome MEAS Definition Press F1 (MEAS) to proceed single measurement After entering measurement mode , press F4 (PRN) to enter printing mode.
MULTIPLE MEASUREMENT This is to manually record what you measured with real time , and file name (a note or identification for the record ) is edit- able to memorize or recognize . There are total 99 memorize capacities in this measurement. F2 :Multiple measurement Welcome MEAS...
AUTOMATIC LOGGING This is to automatically record what you set up before measurement with real time, first to set records start/stop date and time, sample rate , memory points. The screen content shows Expect memory points and Remain memory points while in setting. F3 :Datalogging Welcome MEAS...
9851 Hygo-Thermometer Setting / Single measurement Welcome MEAS Temp: LCD Cont.(1-5): Prn Cont. (1-9): 60.8% 17.4C Auto Off: 20mins Temp:25.5C Enable 20.0C Cali EXIT EXIT EDIT NEXT Set Clock: RH Cali Mode Select MM-DD-YY HH:MM:SS 06 30 05 18 29...
Page 12
9851 Hygro-Thermometer Calibration Procedure Note: "Input cali.data" method is for RH probe replacement. Welcome "Manual calibration" method is for RH probe replacement & old probe MEAS re-calibration. (Need to plug probe into salt bottles) RH Manual Calibration Temp: 60.8% 17.4C Temp:25.5C...
Page 13
9851 Hygro-Thermometer Multiple measurement Welcome MEAS Temp: Temp: WBT: WBT: EXIT MEAS EDIT NEXT EXIT BACK Note: Temp: After saving #01 memory, press 60.8% 17.4C for next measurement.Then, Temp:25.5C #02 with a cursor will be 20.0C ABORT SAVE highlighted. ROOM1 06-30 23:56:49 Temp: 25.5C...
SOFTWARE The enclosed software is a quick tool for you to download the memorized data to PC for further analysis or upload the pre-edited file names to meters, this may save your setting time. Material & O/S Software CD and RS232 cable with D-sub plug are needed. Operating system need to be WIN98/2000/NT/XP or above.
Print Load MODE There are two data modes. One is Memory, the other one is Logger. To select the mode from main menu or quick selection window. In logger mode, you may select from "Goto" to choose logged data range. There are 100 points in one page. Mode-Memory Mode-Logger...
PORT Select the correct Com. port to ensure the meter is communicated with PC. I n connecting mode, "PC Mode" will appear on the LCD and "COM #" & Edit will show on the left bottom corner of screen. For most PCs, the Com. port number is COM1. There are 8 COM ports for selection in this software.
PRINT Select the data you want to print : There are 3 selection : 1. Print only Logger Data 2. Print only Memory Data 3. Print all ( Both Logger Data and Memory Data in sequel ) Example: (If you choose Print all from 9811 IR meter ) Page1 <<...