AURA Strap 2 battery contact issue
Using this guide, you agree to the following terms and conditions: Terms of Use*, Terms of Sales*, and Privacy
Policy*. The guide is compiled according to AURA Devices terms and conditions information remains confidential.
Issue: Can't connect AURA Strap 2 or take a measurement.
Solution: Set the right position of the contact on the battery inside the device.
1) Remove the backward cover with the included pick.
2) Remove the battery and check your device battery contact position.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Strap 2 and is the answer not in the manual?

Questions and answers

Summary of Contents for Aura Strap 2

  • Page 1 Using this guide, you agree to the following terms and conditions: Terms of Use*, Terms of Sales*, and Privacy Policy*. The guide is compiled according to AURA Devices terms and conditions information remains confidential. Issue: Can't connect AURA Strap 2 or take a measurement.
  • Page 2 3) Set the contact position as shown in the picture below:...
  • Page 3 4) C lean this contact by an alcohol napkin or a thick cloth. 5) Insert the battery with positive polarity upwards. 5) Place the cover and press until the "click" sound. Now you can take a measurement.