SPECIFICATION Electrics Power supply 230 V / 50 (opt. 60) Hz., 16 A fuse Power consumption Approx. 350 VA with cooling Environment Medium temperature 0 to +40 °C Ambient temperature –20 to +50 °C Suction height Max. 8 m (at 1013h Pa) volume accuracy +- 5 % at 250 ml average in a set of 10 samples General specifications...
Safety information Please read the entire manual before the equipment is unpacked, set up or operated. Pay attention to all danger and caution statements. Personal injury or damage to the equipment could occur if they are not observed. To ensure that the protection provided by this equipment is not impaired, do not use or install this equipment in any manner other than that specified in this manual.
Warning labels Read all labels and notices attached to the equipment. Personal injury or damage to the equipment could occur if they are not observed. Any symbol on the equipment will appear along with a caution statement in the manual. This symbol, if noted on the instrument, references the user manual for operation and/or safety information This symbol, when noted on a product enclosure or barrier, indicates that a risk of...
Depending on analysis requirements, we can also offer you alternative materials, e.g. different types of silicone or Teflon. Important Note: The sampler is based on the model SP5 B For the part of the basic programming, we refer to the standard manual “Programming instructions for SP-TP-P6“...
DEVICE FEATURES 1. Hood 2. Control with display and keypad 3. Connection for suction hose 4. Distributor arm 5. 24 segment Bottles 1 L 6. Bottle tray 7. Notch for position of bottle tray 8. Peristaltic Pump 9. Capacitive Sensor 2 10.
Details Direct distributor 24 x 1 L It´s a direct distributor, with 24 bottles Bottles from top with filling direction page 11 of 24...
Page 12
The bottles are in a bottle tray, which can be moved out At the bottom of the tray is a bolt for positioning. If you move the tray back, take care to position the bolt in the notch. Page 12 of 24...
ELECTRICAL CONNECTIONS Electrical installation Prepare the electrical installation Loosen the screws and remove the cover (SB5 B) Feed cable through (SP5 B) page 13 of 24...
Page 14
Wiring diagram (SP5 B) Complete the electrical installation (SP5 B) Attach cover Page 14 of 24...
PERISTALTIC PUMP - Important Note: Before you put the sampler into service, the system has to be calibrated! How to do the calibration is described in the ANNEX PUMP TUBE REPLACEMENT Important Note: Use of tubing other than that supplied by the manufacturer may cause excessive wear on mechanical parts and/or poor pump performance! Inspect and clean the pump tubing and rollers on a regular basis.
Page 18
Important Note: The sampler measures the sampling volume with 2 capacitive Sensors. Depending to the sampling point, after some time can be dirt in the silicon tube. If you get error messages (error water sensor), you have to clean the tube! capacitive sensor 2 capacitive Sensor 1 TIP: A very simple way to clean the tube inside is the use of a pig (sponge).
PROGRAMMING MAIN MENU STRUCTURE STATUS/STOP START PROGRAMS CHANGE IN PRESENT BOTTLE MANUAL SAMPLING IN BOTTLE X SAMPLING DATA TEMPERATURE DATA DATA MEMORY BOTTLE PROTOCOL BOTTLE ARCHIV COMPONENT TEST TEMPERATURE BOARD DIGITAL INPUTS DIAGNOSTICS/ TEST ANALOG INPUTS VERSION INFO DATE/TIME SYSTEM SETTINGS SETUP SLEEP MODE...
ANNEX – Calibration Volume of PERISTALTIC PUMP - The single sample volume is defined by 2 capacitive sensors and a measuring tube and is highly accurate. We recommend to calibrate the system every time a new sampling point and from time to time, depending on the customer-specific...
Page 21
Please collect the samples of all 3 cycles in a measuring glass! (Capacity >=2 liters) 3. ACTUAL VALUE upper value Please enter the total volume of the 3 cycles. The upper value is calibrated 4. Start LOWER value There are 3 consecutive calibration cycles which are started. Please collect the samples of all 3 cycles in a measuring glass! (Capacity >=200 ml)