Summary of Contents for Precision Acoustics UT Series
Page 1
300 kHz to 60 MHz range. The UT-Series membrane hydrophone is designed to be used in conjunction with a dedicated power supply that powers the submersible preamplifier. The three items are supplied as a complete system by Precision Acoustics Ltd.
Page 2
NPL, London over the frequency range 300 kHz to 60 MHz. Further information on making acoustic measurements with hydrophones can be found on our website (http://www.acoustics.co.uk). In addition, Precision Acoustics Ltd offers training courses covering many aspects of acoustic measurement techniques and practices.
This mount is available from Precision Acoustics Ltd as an optional extra upon request. It is recommended that all hydrophones receive 1h soaking before use. It is recommended that the hydrophone is placed in the water tank at this stage before completing the rest of the setup process.
Page 5
ORIENTATION AND ALIGNMENT All hydrophones exhibit a directional response and the maximum hydrophone signal will be received when the plane of the hydrophone active element is orthogonal to the acoustic axis of the transducer (as shown in the image below). If the membrane hydrophone is rotated about the X or Y axes, there may be a reduction in the amplitude of received signal, even if the active element remains on the acoustic axis of the source.
Page 6
transducer and locating signal there. Once signal is found, move gradually further away from transducer (aligning hydrophone as necessary on the way) until desired field position is achieved. Noise. Where there is poor signal to noise ratio (SNR), it can sometimes be difficult to locate ▪...
Page 7
• Find the pressure field maximum – centred on your current position, move the hydrophone along a line parallel with the X-axis. Move the hydrophone to the location of the field maximum along that line. Repeat this type of line scan alignment in the orthogonal direction of the Y- axis.
Precision Acoustics for inspection. It is unlikely that the membrane can be repaired but a generous discount will be offered against a new membrane.
Page 9
How do I clean the calcium carbonate deposits from the hydrophone? ▪ o Immerse the hydrophone in lemon juice and then rinse in distilled or running water. Never use a cloth/towel/abrasive cleaner on the membrane Do I have to terminate the hydrophone output in 50 Ω? ▪...
All information is based on results gained from experience and tests, and is believed to be accurate but is given without acceptance of liability for loss or damage attributable to reliance thereon as conditions of use lie outside the control of Precision Acoustics Ltd. Warranty Warranty will be for 12 months against defect of hardware component or manufacture only.