All safety precautions must be met, especially in the cell breakdown procedure after the filtration procedure has been completed. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Evenly tighten the set screws with the allen wrench. O-ring Groove Cell Cap (#171-62) Locking Screws (#171-78) OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 9
Back Pressure Receiver (#171-58) T-screw T-screw Cartridge Holder (#143-03) Needle Valve Needle Valve (#170-32) (#170-34) Male Coupling (#171-64) Locking Ring O-ring (#142-58) Back Pressure Receiver OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 10
500 PSI (3,447.5 kPa) above the back pressure. Open the bottom valve to initiate filtration and begin timing. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 11
Measure and report the thickness of the filter cake to the nearest 1/32" (0.8 mm). 16. Clean and dry the apparatus thoroughly after each use. Inspect all o- rings and replace if necessary. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 12
Safety Bleeder Valve (#143-06) (#171-75) Cartridge Barrel (#143-03) Male Needle Valve (#170-32) Graduated Cylinder (#153-16) Filter Paper (#170-19) Stopcock Grease (#153-55) Stand and Base (#171-59) OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...