OCULUS Binoptometer 4P Instruction Manual page 12

Vision test device
Table of Contents

Advertisement

4 Safety Instructions
Instructions on Electrical Safety
8 / 55
Caution
Risk of personal injury or damage to property due to an incorrect level of
safety
Connecting the Binoptometer® 4P with its non-medical electrical
equipment (e.g. data processing equipment) to a medical electrical
system must not result in a patient safety level below that prescribed by
DIN EN ISO 60601-1. If making this connection leads to the leakage
current threshold being exceeded, protective measures including a circuit
breaker must be in place.
Ensure that connections with non-medical devices are made correctly.
Use only a computer that meets the specifications given in this
instruction manual,
sec. 17, page
Use of a multiple socket extension cord
Risk of personal injury or material damage caused by unsafe multiple
socket extension cord
If you use a multiple socket extension cord to connect the Binoptometer® 4P
to the power supply, you must heed the following information:
Use an extension cord that complies with the requirements of DIN
EN ISO 60601-1: 20005, section 16.
Do not place the multiple socket extension cord on the floor.
Do not use more than one multiple socket extension cord.
Plug only the Binoptometer® 4P and the computer that is being used
with the unit (if applicable) into the multiple socket extension cord.
If you are using a multiple socket extension chord it has to be supplied
with a isolation transformer.
If you are using a new computer for the Binoptometer® 4P, you must have
the electrical safety checked. Call OCULUS Service for this purpose.
Electromagnetic Compatibility (EMC) / Cables
Risk of personal injury or damage to property due to electromagnetic
interference
Portable and mobile RF communications equipment can affect medical
electrical equipment
sec. 18, page
Make sure that portable and mobile RF communications equipment
do not cause interference.
Recommendation: Maintain a minimum distance of 4 m. If the
distance is shorter, you must ensure that the Binoptometer® 4P
functions correctly.
Instruction Manual Binoptometer® 4P (G/59860/1016/en)