III. Electrical Specification (23 ° C ± 5 ° C, 5 minutes after power is on)
mA DC Current (1K Ω Max. Load, 24V Loop Supply)
Range
4 - 20mA, 0 - 20mA, 0 -24 mA
Beeper warning when output is open and specified current output > 1mA
mV, V DC Voltage (1mA Supply Current)
Range
0 - 100.00mV
0 - 10.000V
0 - 1.0000V
Beeper warning when output is short and specified voltage output > 10mV
K, J, E, T type Thermocouples( 1 ° C,1 ° F Resolution,1K Ω load Min. )
Range
K: -200 to 0 ° C
K: 0 to 1370 ° C
J: -100 to 0 ° C
J: 0 to 760 ° C
E: -100 to 0 ° C
E: 0 to 700 ° C
T: -200 to 0 ° C
T: 0 to 400 ° C
Resolution
Resolution
10 µ V
1mV
100 µ V
Accuracy
Range
± 1.1 ° C
K: -328 to 32 ° F
± 0.8 ° C
K: 32 to 2400 ° F
± 0.9 ° C
J: -148 to 32 ° F
± 0.7 ° C
J: 32 to 1400 ° F
± 0.9 ° C
E: -148 to 32 ° F
± 0.7 ° C
E: 32 to 1292 ° F
± 1.0 ° C
T: -328 to 32 ° F
± 0.8 ° C
T: 32 to 752 ° F
27
Accuracy
1 µ A
± 0.025% ±3 µ A
Accuracy
± 0.05% ± 30 µ V
± 0.05% ± 3mV
± 0.05% ± 300 µ V
Accuracy
± 2.0 ° F
± 1.5 ° F
± 1.6 ° F
± 1.2 ° F
± 1.6 ° F
± 1.2 ° F
± 1.8 ° F
± 1.5 ° F