TrachFlush Quick Guide
How to start up TrachFlush
Power ON / OFF
2 min alarm silence
Actual cuff pressure
Increase cuff pressure
Decrease cuff pressure
Cuff hold button
Cuff pressure unit of measure
Target cuff pressure
Deflate button
Flush button
LED bar
Power Supply
Airway tube
Cuff pressure tube
TrachFlush Start Up
STEP 1: Connect power supply. Press
and hold the ON/OFF button for 3
seconds to turn on the device.
STEP 3: Connect the Airway Tube
and Cuff Pressure Tube to
TrachFlush by twisting them on
STEP 5: Actual cuff pressure is
measured, and the cuff
controller is OFF. See Page 2 on
how to set up cuff pressure.
Page 1 of 4
STEP 2: Calibrate the device by
pressing the
.
Note: Remove tubes from TrachFlush if
connected
STEP 4: Connect the Airway
Tube to the gas outlet on the
connector or filter, and the
Cuff Pressure Tube to the
pilot balloon
18

Advertisement

loading
Need help?

Need help?

Do you have a question about the TrachFlush and is the answer not in the manual?

Questions and answers

Summary of Contents for AW Technologies TrachFlush

  • Page 1 Power ON / OFF STEP 2: Calibrate the device by 2 min alarm silence pressing the Note: Remove tubes from TrachFlush if Actual cuff pressure connected Increase cuff pressure STEP 3: Connect the Airway Tube...
  • Page 2 TrachFlush Quick Guide How to use the Cuff Controller, Deflate & Cuff Hold Cuff hold Cuff Controller Deflate STEP 1: Press to increase or STEP 1: Press the on main screen. STEP 1: Press the on main screen. decrease the cuff pressure Press to accept Deflate.
  • Page 3 TrachFlush Quick Guide How to use Flush Control and Spont Flush Control Pressure SUPPORT settings Pressure CONTROL settings Troubleshooting To perform a Spont Flush Control, the STEP 1: If ventilator settings are To perform a Flush Control, the following settings are required.
  • Page 4 Cuff pressure < 5 cmH Cuff pressure is above the target pressure, Make sure to return to Pre-Flush or Increase cuff pressure please wait for TrachFlush to reduced Pre-Spont Flush ventilator settings Unstable ventilator signal Cuff regulation error Flush Control in progress...