Contents 1 Product introduction 1.1 Product description 1 Product introduction ................1 EDGEAPEX HP is mainly used in Endodontic treatment. It is a cordless endo motor with root canal measurement capability. It can be 2 Installation .................... 5 used as a endo motor for preparation and enlargement of root canals, 3 Function and operation of product ............
Page 3
1.5 Intended Use 1.5.1 The device can be used for preparation and enlargement of root EDGEAPEX HP canals, or device for measuring canal length. 1.5.2 The device must be operated only in hospitals and clinics by 1.5.3 CAUTION: Federal (US) law restricts this device to sale by or on the order of a licensed professional.
1.9.2 Power adapter (Model: ADS-6AM-06N 05050/ 1.7.8 The motor handpiece cannot be autoclaved. Use disinfectant of UE08WCP-050100SPA) neutral pH value or ethyl alcohol to wipe its surface. Input: ~100V-240V 50Hz/60Hz, 0.4A Max 1.7.9 Do not press the push cover of contra angle before the contra Output: 5V angle stops rotating.
Page 5
2.2.1.4 SGP Mode Safety Glide Path Mode F: Forward angle, R: Reverse angle Measuring wire File clip The rotation angle is adjustable, but the forward angle must be equal to the reverse angle. 2.2.1.5 ATR Mode ATR: Adaptive Torque Reverse function. Lip hook Touch probe Disposable insulation sleeves...
Page 6
positioning slot locating pin The meter numbers 1.0, 2.0, 3.0 and digital numbers 00-16 do not represent the actual length from the apical foramen. It simply indicates the file progression towards the apex. The digital numbers -1 and -2 indicate that the file has reached the apex foramen. Subtract 0.5-1mm used to estimate the canal’s working length.
Page 7
2.6 Canal measurement functional connection and slips into place. Release the button to lock the file into the contra This is not required if the canal measurement function will not be angle. used. Connect the measuring wire to the motor handpiece. Line up the measuring wire plug with the notch on the back of the motor and push it all the way in.
2.7 Installation and removal of disposable isolation sleeves e. Preset program selection Long press Setting button “P” to enter preset program during standby 2.7.1 Installation button “P” to entry enter file number selection, press Adjusting button i t a i t a i t a i t a f.
Canal measurement state interface Flash Bar Position a. Canal length indicator bar apical action is triggered. b. Indication number Auto Start Digital numbers 00-16 do not is inserted in the canal. represent the actual length from the Auto Stop apical foramen. It simply indicates is taken out of the canal.
Page 10
c) Press the Main button again, and then the motor handpiece returns The speed setting can be adjusted from 100 rpm to Standby interface. to 2500 rpm. d) Hold down the Setting button “P”, then press Main button to turn Press Adjusting button “+”/“-”...
Page 11
inserted into the canal and the canal length indicator bar lights up more than 2 bars. Forward Angle .In the SGP mode, the Forward P ress Adjusting button “+”/“-” to change. Angles of 20°~400° are available. In the ATR mode, the Forward Angle of into the canal.
Page 12
The parameters of the presets can also be Press Setting button “P” again, the “Auto Standby Scr” can be changed, press Adjusting If you want to change back to default setting, button “+”/“-” to adjust, then press to "Main" long press Setting button “P” to enter preset program during standby state, select a preset This is the automatic return to standby display of motor handpiece when no buttons are...
Page 13
Motor combined with canal preset torque value again. measurement function mode When using motor combined with canal measurement function, the measuring wire must be connected with motor handpiece by USB socket, and white socket connects with patient’s lip by lip hook, keep the black socket idle.
Page 14
4.8 Canal measurement operation Connection testing It is strongly recommended to test the When using in standalone apex locator connection every time before use. Clip mode, we suggest putting the motor the holder onto lip hook and check that handpiece on the charging base to get all the bars on the meter on the screen better visual angle.
Page 15
Broken crown. Extremely dry canal. If the crown is broken and a section If the canal is extremely dry, the meter of the gingival tissue intrudes into the may not move until it is quite close to cavity surrounding the canal opening, the apex.
b) Use tweezers etc. to open the rubber cover and then remove the 5 Troubleshooting screw. Failure Possible cause Solutions c) Remove the battery cover. The motor handpiece Motor is on EAL mode, Change to CW, CCW, d) Remove the old battery and disconnect the connector. does not rotate.
Page 17
Do not use chloride detergent materials. a)Use distilled water or deionized water for cleaning at room 6.2.3 Do not use bleach or chloride disinfectant materials. temperature. 6.2.4 For your own safety, please wear personal protective equipment 6.3.1.2 Manual disinfection steps: (gloves, glasses, mask).
Page 18
6.4 The cleaning, disinfection and sterilization of contra-angle,lip Tools: tray, soft brush, clean and dry soft cloth. Unless otherwise stated, they will be hereinafter referred to as wire from the handpiece in sequence, and then put them into a clean tray; “products”.
Page 19
• The cleaning procedure is suitable for the product, and the irrigating carried out together. c) Cleaning: (c1) The cleaning procedure should be suitable for the It is recommended to use a washer-disinfector in accordance with product to be treated. The flushing period should be sufficient (5-10 minutes).
Page 20
6.4.6 Inspection and maintenance 6.4.6.1 Inspection to EN 285 to comply with EN ISO 17665 or is cleared by FDA; In this chapter, we only check the appearance of the product. The validated sterilization cycle is one (1) fraction cycle, three (3) 1.
Notes: 7.3.3 Keep out of direct sunlight and avoid getting wet in rain and a) The storage environment should be clean and must be disinfected snow during transportation. regularly; 8 Environmental protection b) Product storage must be batched and marked and recorded. Please dispose according to the local laws.
Technical Description Concerning Electromagnetic Emission Appliance compliance Handle with care Table 1: Declaration - electromagnetic emissions WEEE directive Guidance and manufacturer’s declaration - electromagnetic emissions The model is intended for use in the electromagnetic environment specified below. EDGEAPEX HP Humidity limitation Temperature limitation The customer or the user of the model should assure that it is used in such an...
Page 23
Voltage dips, short <5 % UT <5 % UT NOTE I At 80 MHz end 800 MHz. the higher frequency range applies. Mains power quality should be interruptions and (>95% dip in UT.) (>95% dip in UT.) that of a typical commercial or voltage variations on for 0.5 cycle for 0.5 cycle...
Page 24
Endodontic Motor Troubleshooting This guide is suitable for the users who use Endodontic Motor for the first time, and those who get unstable readout because of improper operation. Problems Possible Causes Analysis Solutions No readouts File clip File clip wire broken or poor contact Replace the clip or unstable...
Page 26
EdgeApex HP statement regarding reciprocating function: This EdgeApex HP features a new ATR mode that enables torque-based reciprocation. If the file torque is less than the set trigger torque value, the file will keep rotating in the cutting direction. When the file torque is more than the set trigger torque value, the file will automatically reciprocate in the non-cutting and cutting directions repeatedly at the set angles.
Page 27
Settings for other EdgeEndo systems: You can utilize the open program slots M0-M9 to customize and save as your favorites. The below chart details the recommended settings for customizing program slots M0-M9 for our given file systems. Use the CW mode and adjust to the following ROTARY settings:...