DIAGnostic:INTerrupt:LINE?

Returned Data
Reset Condition
Related
Commands
Example

DIAGnostic:TEST?

Caution
Chapter 3
DIAGnostic:INTerrupt:LINE? returns the VXIbus interrupt line being used by the
module
Type
Range
int16
0 through 7
*RST does not affect the interrupt line.
DIAGnostic:INTerrupt:LINE
DIAG:INT:LINE?
DIAGnostic:TEST? performs an extensive relay self-test and returns a numerical
and string response indicating the results of the test.
The extended self-test will open and close each relay in the module.
Before performing this test, make sure that external devices will not be
affected by these actions. It is recommended that external devices be
disconnected from the module while executing DIAGnostic:TEST?.
Default
Description
1
VXIbus interrupt line
Query the interrupt level
SCPI Command Reference