Download Print this page

Xetron Elog1 Instruction Manual page 2

Single use temperature data logger

Advertisement

Elog1/1N
Instruction Manual
Single Use Temperature Data Logger
Product instruction diagram
Waterproof
Plastic Bag
USB Port
LED Indicator
Elog1/1N
Button(Start/
Status/Mark/Stop)
Barcode Label
Front
*Please refer to the actual product
Product description
This temperature recorder is mainly used for temperature monitoring and recording of food and medicine storage and transportation
environments, such as refrigerated bags, refrigerated trucks, containers, etc.
This product is connected to a computer via USB , and the recorded data is directly output as a PDF report; the temperature probe is
built-in, powered by a CR2450 wide temperature lithium battery, and the waterproof level reaches IP67 . The product is equipped with a
barcode.
Technical parameters
ll function-related parameters of this recorder are configured by the factory, and users can customize some of these parameters
according to their needs;
Measuring temperature
-30°C~70°C/-22°F~158°F
Storage temperature
-30°C~70°C/-22°F~158°F
±0.5°C/±0.9°F ( -20°C~40°C/-4°F~104°F) ;
Temperature
1°C/1.8°F (other)
measurement accuracy
Sensor
Built-in NTC
Recording
capacity of 16000 groups (maximum)
Recording period
6、 15、 30、 60、 90、 120、 180 days (customizable)
Alarm type
single type ( Sin ), cumulative type ( Cum )
Alarm threshold
Factory default (customizable)
Note: Before the recorder is started, it is recommended to store at room temperature
Start/Stop Info
Mark Info
Status Request
Paste Area
Configuration Info
Back
Data interface
USB2.0/NFC
Report type
PDF format data report
Power supply
Built-in CR2450 wide
temperature lithium battery
Battery life
2 years of storage and use in
normal temperature environment
Protection grade
IP67
Dimensions
93*58*8mm (length * section * height)
Weight
about 20g
Parameter description:
Record capacity:
16000 groups (Max), there is di erence due to the recording cycle.
Delayed startup:
record after 1 or more minutes(custom) later after logger startup.
Alarm threshold:
the upper and lower temp limits of alarm can be set.
H
Over upper temp limit
L
Below lower temp limit
Alarm type:
Single type
Single calculation of continuous over-temperature time
Cumulative type
Cumulative calculation of all continuous over-temperature time
Alarm delay:
Alarm starts after over-temp duration exceeds 1 or more minutes(custom), instead of immediately.
Operating instructions
Long press 5 sec, the green light flashes
Tear o
the barcode if
5 times, logger starts normally.
you need
NFC operation instructions (ONLY for Elog1N)
1. How to install APP?
・Download by scanning the QR code or searching "Xetron iCold" in app store.
・Register a new account then sign in.
2. How to set parameters?
・Connect to the logger:
Click the navigation button
at the upper left corner
Select the NFC temp label (confirm the NFC function of the mobile phone is turned on)
・Set parameters:
Click
at the upper right corner to set parameters, keep close to logger then click "OK" to save settings
(Note: parameters only can be configured before the logger starts)
3. How to view and export data?
・Connect logger and phone app firstly as above mentioned. View data directly once connection is successful.
・Click
to select report format to export.
NFC Temperature Tag
My Device
Place your phone above the Tag
Enter Device Name or Model
APP
NFC
No Device
Long press 5 sec, the red light flashes 5
Insert the logger into the USB
times, logger stops normally.
port of computer to view data
report directly or by software.
NFC Temperature Tag
NFC Temperature Tag
EFM200000000
Wait for be...
EFM200000000
Record
23.2°C
Max-
23.4°C
Max 23.4°C
Configure Time:2021-08-26 14:30:45
Min-
Started:2021-08-26 14:33:20
Min 23.4°C
Running Time:--
Data Points:--
Running Time:00h 00m 00s
Data Points:1
Basic Settings
Basic Settings
EFM200000000
EFM200000000
Serial Number
Serial Number
Temperature Uint
°C
Temperature Uint
°C
UTC+08:00
UTC+08:00
Timezone
Timezone
V1.2
V1.2
Firmware Version
Firmware Version
Logging Settings
Logging Settings
Logging Interval
00h 00m 10s
Logging Interval
00h 00m 10s
Start Delay
00h 00m 00s
Start Delay
00h 00m 00s
Alarm Settings
Alarm Settings
Name
Threshold
Delay
Type
Name
Threshold
Delay
Type
HI
8.0°C
00h 30m 00s
Single
HI
8.0°C
00h 30m 00s
Single
L2
2.0°C
00h 30m 00s
Single
L2
2.0°C
00h 30m 00s
Single