Input Filter; Water Ingress Protection (Optional) - SARAD RTM 1688-2 Manual

Radon- and thoron-measurement equipment
Hide thumbs Also See for RTM 1688-2:
Table of Contents

Advertisement

RTM 1688-2

Input Filter

The instrument is equipped with a high efficient multi-stage filter to prevent penetration of
daughter products into the measurement chamber. This filter will be protected against
contamination by an additional syringe filter fitted to the air inlet. Do not use the instrument
without that filter and replace it if required (becomes dark if loaded with dust). Take care for
the correct flow direction. Some filters can block the air flow completely when used in wrong
flow direction.

Water ingress protection (optional)

Please never measure radon in soil gases or in water samples without the stainless-steel
container connected to the water inlet protection! This prevents that water could enter into
the high-voltage chamber of the device. The penetration of water into the device can lead to
its failure and, as a result, to very expensive repairs!
SARAD GmbH offers to its customers a special stainless-steel container with floating contact
to prevent the penetration of water into the device when measuring radon in soil gases and
in water samples. For this purpose, a special stainless-steel container (see photo) is connected
to the air suction line of the radon measuring device between the soil air probe or water
sample and the measuring device and also connected with a plug. If water penetrates the
stainless-steel container, the pump and the air intake are automatically switched off, to
prevent the penetration of water into the high-voltage chamber and which results in a failure
of the device. To continue the measurement, pour out the water that has entered in the
stainless-steel container. The plug should be disconnected from the device and the stainless-
steel container should be removed for emptying.
In order to ensure the correct functioning of the water ingress protection system, the
stainless-steel container must be strictly in vertical position during the measurement!
If the device is used to measure radon in air, the supplied loose plug must be inserted instead
of connecting the special stainless-steel container. Otherwise the built-in pump would no
longer suck air.
fig 2: water ingress protection
 SARAD GmbH 2023
Manual_RTM1688-2_EN_24-02-2023.docx
7