CEFLA I-MMr Operator's Manual

CEFLA I-MMr Operator's Manual

Micromotor

Advertisement

Quick Links

97050840
Rev. 03
2022-10
EN
.

Advertisement

Table of Contents
loading

Summary of Contents for CEFLA I-MMr

  • Page 1 97050840 Rev. 03 2022-10...
  • Page 2: Table Of Contents

    I-MMr • i-MMr L • i-MMs Contents DESCRIPTION ......................................... 3 1.1. DESCRIPTION OF THE DEVICE ................................3 1.2. INDICATIONS FOR USE ....................................3 1.3. IMPORTANT WARNINGS ...................................3 1.4. SYMBOLS ........................................4 TECHNICAL DATA ......................................5 2.1. MODE OF USE ......................................6 2.2. ENVIRONMENTAL CONDITIONS FOR USE ..............................6 2.3.
  • Page 3: Description

    1.2. INDICATIONS FOR USE The CEFLA Dental Micromotors are brushless electric micromotors controlled by a control unit inside CEFLA Dental Units. They are intended to be connected with an ISO-type handpiece attachment: straight or contra-angle of equal, gear-reducing, or gear increasing speed.
  • Page 4: Symbols

    I-MMr • i-MMr L • i-MMs 1.4. SYMBOLS WARNING! Failure to observe may result in equipment damage or injury to the user and/or patient. Symbol to call the attention to other information found in the equipment User Manual. NOTE: Identifies information that is especially important for the user and/or the assistant.
  • Page 5: Technical Data

    I-MMr • i-MMr L • i-MMs TECHNICAL DATA i-MMr i-MMs i-MMr L Type Brushless Power supply 32Vdc ± 12,5% Max absorption 4,5 A Max Electrical power absorbed 170 W 120 W Max speed 40.000 ± 10% revs/min 40.000 ± 10% revs/min Min speed 100 ±...
  • Page 6: Mode Of Use

    I-MMr • i-MMr L • i-MMs 2.1. MODE OF USE The method of application of the device is intermittent. The operational cycle indicated respects the following times: Type of Instrument Operation (minutes) Pause (minutes) i-MMs i-MMr i-MMr L 2.2. ENVIRONMENTAL CONDITIONS FOR USE •...
  • Page 7: Reprocessing

    I-MMr • i-MMr L • i-MMs REPROCESSING RACCOMANDATION: reprocess shall be performed within 30 minutes of the end of use. The micromotors have been tested for 250 cycles (as required by EN ISO 14457). Before sanitisation operations, the operator must detach the handpiece from the micromotor and the micromotor from the dental units tubing.
  • Page 8: Sterilisation

    4.3. PRODUCTS TO BE USED The Manufacturer recommends to use the following products: CEFLA s.c. does not guarantee proper device operation if products other than those indicated are used. EU market • IC100 (ALPRO MEDICAL GMBH) or other CE marked detergent DETERGENT FLUIDS •...
  • Page 9: Maintenance

    I-MMr • i-MMr L • i-MMs MAINTENANCE At least once a week lubricate the O-Ring seals found in the handpiece coupling with silicone lubricating grease. Wearing disposable gloves, apply a film of grease on the fingertips and lubricate with the fingers.
  • Page 10: Scrapping

    I-MMr • i-MMr L • i-MMs SCRAPPING Scrapping the device must be carried out with respect for the laws in force for electrical and electronic equipment, according to the individual national laws. The materials used for the manufacture are not harmful for human beings or animals that come into contact with or are exposed to them.
  • Page 12 www.cefla.com...