Do you have a question about the ICA-100 and is the answer not in the manual?
Questions and answers
Summary of Contents for Daihan Scientific ICA-100
Page 1
Air-jacketed CO2 Incubator Certifications by - - - ISO 9001 Certified. PL Insurance. User Manual (Version : 1.0.0) written for Air-jacketed CO2 Incubator, ICA-100/-150 models. "ICA-100” Air-jacketed CO2 Incubator, 101 Lit Sales & Basic A/S : DAIHAN Scientific, Air-jacketed CO2 Incubators 1/14...
1.7. Operating ····························································· 11 1.8. Warnings······························································ 12 1.98. Trouble Shooting ·················································· 13 Thank you for purchasing DAIHAN Scientific products. Please 1.2. Getting Started read this manual carefully to get used to the functions of the unit, the precautions to take care of it, and how to operate the unit safely and smoothly.
1.3. Product Specifications DAIHAN Scientific, Air-jacketed CO2 Incubators 3/14...
Page 4
PID-A control method. After reaching the set temperature, the power supplied to the heater is appropriately controlled so that the temperature deviation is minimized, to ensure the stability of temperature. DAIHAN Scientific, Air-jacketed CO2 Incubators 4/14...
◘ CO2 H/L Lamp This is an indicator lamp when CO2 gas inside of the chamber becomes higher (High) or lower (Low) than the set CO2 gas amounts, which lights up with an alarm sound. DAIHAN Scientific, Air-jacketed CO2 Incubators 5/14...
Page 6
This button used when you move to the left column when inputting, Or to turn off the alarm when the alarm sounds. ◘ UV Button This button is used to operate the UV Lamp. It does not work when the unit is running. DAIHAN Scientific, Air-jacketed CO2 Incubators 6/14...
When pressing the START key in the control standby screen, the standby screen changes to the active screen and starts to control of each setting value. At this time, CO2 gas starts control from the moment the temperature reaches 36.0 ℃ for precise control. DAIHAN Scientific, Air-jacketed CO2 Incubators 7/14...
Page 8
Pgm-On Delay : It is displayed only when the control is not in operation (STOP status). It is an item to delay starting of control after set time. DAIHAN Scientific, Air-jacketed CO2 Incubators 8/14...
Page 9
When the sterilization is over after the time entered, the buzzer sounds with the Time-Out message. The UV button returns to the initial screen. DAIHAN Scientific, Air-jacketed CO2 Incubators 9/14...
Page 10
~ 20% higher than the temperature you want to use. This function prevents damage to the sample inside the chamber caused by overheating. Gas Panel ② ③ ① ① Power Plug Socket ② CO2 Gas Inlet Port ③ CO2 Gas Calibration Outlet Port DAIHAN Scientific, Air-jacketed CO2 Incubators 10/14...
Page 11
1.7. How to replace HEPA Filter(Option) DAIHAN Scientific, Air-jacketed CO2 Incubators 11/14...
8. Before use, please clean up the inner chamber. - Please use soaked gauze with isopropyl alcohol(70%) - Please disassemble the shelves and clean up bottom and Upper side. - Please dry around the sensor and fan before use. DAIHAN Scientific, Air-jacketed CO2 Incubators 12/14...
Over 60℃ heat Triac Failure Check Triac Replace Triac uuuu Short or part failure Check wiring Replace parts If any troubles persists after the trouble shootings, please contact the place of purchase or customer service center. DAIHAN Scientific, Air-jacketed CO2 Incubators 13/14...
Page 14
Air-jacketed CO2 Incubator Model No. ICA-100 / ICA-150 Serial No. Date Customer DAIHAN Scientific Co., Ltd. 326, Sinpyeongseokhwa-ro, jijeong-myeon, Wonj-si, Gangwon-do, Korea TEL: +82-2-967-5235 FAX: +82-2-963 5231 Email: info@daihansci.co.kr -Sci.com www.DAIHAN www.ALL for LAB.com DAIHAN Scientific, Air-jacketed CO2 Incubators 14/14...
Page 15
)인증! PL Insurance. (제조물 배상책임보험) 벤 처기업 사용자 설명서 (Version : 1.0.0) written for Air-jacketed CO2 Incubator, ICA-100/-150 models. "ICA-100” Air-jacketed CO2 Incubator, 101 Lit Sales & Basic A/S : 판매 및 기본 A/S : DAIHAN Scientific, Air-jacketed CO2 Incubators...
Page 16
이 표시는 고온의 히터를 사용하므로 화상의 위험이 있음을 알립니다. 폭 발 위 험 이 표시는 휘발성, 폭발성 화학물질을 사용할 경우 폭발의 위 험이 있음을 알립니다. 접 근 금 지 기기의 사용 중에 사용자의 접근을 제한함을 알립니다. DAIHAN Scientific, Air-jacketed CO2 Incubators 2/14...
도달한 후에는 히터에 공급되는 전력이 적절하게 조절되어 온도편차를 극소화 함과 동시에 우수한 온도의 안전성이 유 지되도록 하였습니다. 또한 챔버의 미관과 고도의 밀폐성을 유지하기 위하여 스테 인레스 재질을 사용하였으며 고급단열재를 사용하여 기기 외 부의 과열을 최대한 방지하였습니다. DAIHAN Scientific, Air-jacketed CO2 Incubators 4/14...
챔버 내부의 온도가 설정한 온도보다 높아졌거나(High) 낮아 졌을 때(Low) Alarm 과 함께 점등되는 램프입니다.. ◘ CO2 H/L Lamp 챔버 내부의 CO2 가스량이 설정한 CO2 값보다 높아졌거나 낮 아졌을 때 Allarm 과 함께 점등되는 램프입니다. . DAIHAN Scientific, Air-jacketed CO2 Incubators 5/14...
Page 20
◘ ← Button(Mute Button) 입력 시 왼쪽 칸으로 이동하는 버튼 입니다. 또는 알람이 울릴 때 알람 소리를 끄는 버튼 입니다. ◘ UV Button UV Lamp 를 동작시킬 때 사용하는 버튼입니다. 기기가 동 작 중일 때는 작동하지 않습니다. DAIHAN Scientific, Air-jacketed CO2 Incubators 6/14...
이 화면에서 1 분 이상 입력이 없을 시, 자동적으로 현재 값 표시 화면으로 돌아갑니다. 제어 대기 화면 상태에서 START 키를 누르면 대기 화면은 실행 화면으로 바뀌며 각 설정 값의 제어를 시작합니다. 이 때, CO2 가스는 정밀한 제어를 위해 온도가 36.0℃가 된 순간부터 제어를 시작합니다. DAIHAN Scientific, Air-jacketed CO2 Incubators 7/14...
Page 22
UV-LAMP Time : 온도제어 시작 전 UV LAMP 를 켜는 시간을 설정하는 항목입니다. 입력한 시간만큼 UV LAMP 가 켜진 뒤 온도 제어를 시작합니다. Pgm-On Delay : 제어를 하지 않는 상태(STOP 상태)에서만 표시되며 입력한 시간 뒤에 제어가 시작되도록 지연을 주는 항목입니다. DAIHAN Scientific, Air-jacketed CO2 Incubators 8/14...
Page 23
Lamp 가 점등되며 입력한 시간만큼 살균이 진행 됩니다. 왼쪽 화면과 같이 살균 중 이라는 문구가 표시 되며 남은 살 균 시간을 시간:분 단위로 표시 합니다. 입력한 시간이 지나 살균이 종료되면 Time-Out 문구와 함께 부저가 울립니다. UV 버튼으로 초기화면으로 돌아갑니다. DAIHAN Scientific, Air-jacketed CO2 Incubators 9/14...
Page 24
정도 높은 온도에 맞추어 두시면 됩니다. 이 기능은 과열되어 발생하는 챔버 내부의 시료의 훼손을 방 지하여 줍니다. ※ Gas Panel ② ③ ① ① Power Plug Socket ② CO2 Gas Inlet Port ③ CO2 Gas Calibration Outlet Port DAIHAN Scientific, Air-jacketed CO2 Incubators 10/14...
컨트롤러 확인 컨트롤러 교체 uuuu 60℃ 이상 과열 트라이악 불량 트라이악 확인 트라이악 교체 uuuu 합선 및 부품 불량 배선 확인 부품 교체 점검 후 이상이 지속될 경우 구입처나 본사 고객상담실로 서비스를 의뢰하여 주십시오. DAIHAN Scientific, Air-jacketed CO2 Incubators 12/14...
Page 27
4. 천재지변(화재, 염해, 가스해, 수해, 기타)에 의한 고장인 경우. (주) 대 한 과 학 강원도 원주시 지정면 월송리 신평석화로 326 Call Center: 080-008-3000 A/S 만족 서울 사업소 : 02)967-5235 FAX: 033)737-7575 본사, 02)967-5258~9 서울사업소. Email: info@daihansci.co.kr www.DAIHAN-Sci.com (www.ALL for LAB.com) DAIHAN Scientific, Air-jacketed CO2 Incubators 13/14...
Page 28
경상북도 포항시 남구 효자로 118 번길 46-2 054-275-1018 054-274-1014 하남 (주)에스엔씨 (대표 성진재) 경기도 하남시 조정대로 150 하남아이테코 441 02-1644-5778 031-790-1790 (주) 대 한 과 학 ㈜대한과학 Call Center 080-008-3000(본사) 080-008-9009(대전) Call Center 080-008-3000(본사) 080-008-9009(대전) 서울사업소 : 02-967-5235 E-mail: info@daihansci.co.kr DAIHAN Scientific, Air-jacketed CO2 Incubators 14/14...
Need help?
Do you have a question about the ICA-100 and is the answer not in the manual?