. TCT301 Series Termination Box Overview: One triaxial sensor input (three measurement channels), polycarbonate enclosure The TCT301 Termination Boxes are a common cable termination point for bringing compatible cables for triaxial sensors into a termination box for routine data collection with portable data collectors .
Mounting Instructions Molded mounting brackets are included on the enclosure . Wall anchoring screws are not included . Mounting Holes 0 .22 in ./5 .59 mm Figure 3. Cable Termination Box Rear View...
Conduit Entry If you have purchased an enclosure without cable entries provided, you should add your own entry prior to mounting the termination box . CTC does not recommend putting holes in the top of the enclosure as it will cause restricted access to the wire termination connection points and also allow possible moisture ingress .
Grounding Optional grounding is available on the TCT301 Series . Select the “with ground lug” option (TCT301G) when ordering if external grounding is desired . A. Mounting to Earth Ground When mounting TCT301G Series Termination Boxes to earth ground (such as an I-Beam), mount the shield ground wire using a mounting bolt through one of the mounting brackets on the enclosure .
Grounding B. Mounting to Non-Grounded Structure When mounting the TCT301G to a non-grounded structure, ensure the shield ground wire or customer-supplied ground wire is tied to a source of earth ground . Figure 6. Ground Wire Placement Sensor Installation Installation of Sensors/Signal Input Cable Feed blunt-cut end through the cable entry at the bottom of the enclosure .
Page 9
Figure 8. Bottom View with Cord Grips Installed Figure 7. Front View Strip outer jacket of cable back 1¼ in . and remove all of the shielding . Separate the internal wires from the shield . Strip red and black insulation back ¼ in . (-) Common Shield / Drain Wire (+) Channel Y...
Page 10
Sensor Installation SENSOR #1 SENSOR #1 Inputs Sensor Inputs Green = Positive X Axis Red = Positive Y Axis White = Positive Z Axis Black = Common Sensor Inputs Grey = Shield (Ground) Green = Positive X Axis Red = Positive Y Axis White = Positive Z Axis Black = Common Grey = Shield (Ground)
Post Installation Testing The TM1018 Accelerometer Verification Meter can be used to verify cable conductivity, sensor location and proper wiring connections . The Verification Meter will indicate if the sensor, cable and/or junction box is in working condition . It will also confirm bias voltage of the accelerometer, which will inform you of the operation of the internal accelerometer amplifier .
Post Installation Testing The following LED Readout indicates the circuit integrity: Green LED: Normal . Indicates proper connection and an output bias will be given, indicating the health of the sensor (4 – 16 V indicates a healthy accelerometer) . Yellow LED: Open Circuit .
Page 13
Warranty & Refund Warranty All CTC products are backed by our unconditional lifetime warranty . If any CTC product should ever fail, we will repair or replace it at no charge . Refund All stock products can be returned for a 25% restocking fee if returned in new condition within 90 days of shipment .