Instruction manual
Pocket – Tester
Max-Planck-Straße 16 | D-31135 Hildesheim | Fon: +49 (0) 5121 7609-0 | Fax: +49 (0) 5121 7609-44 | www.neomeris.de

Advertisement

loading
Need help?

Need help?

Do you have a question about the Basic pH Pocket-Tester and is the answer not in the manual?

Questions and answers

Summary of Contents for Neomeris Basic pH Pocket-Tester

  • Page 1 Instruction manual Pocket – Tester Max-Planck-Straße 16 | D-31135 Hildesheim | Fon: +49 (0) 5121 7609-0 | Fax: +49 (0) 5121 7609-44 | www.neomeris.de...
  • Page 3: Table Of Contents

    Instructions for pH1 / Conductivity 1 Calibration procedure Instructions for pX4 / pH5-Food / ORP5 / Cond5 / PC5 / PC6 Calibration procedure ERRORS DESCRIPTION Max-Planck-Straße 16 I D- 31135 Hildesheim I Fon: +49 (0) 5121 7609-0 I Fax: +49 (0)5121 7609-44 I www.neomeris.de...
  • Page 4: Data Sheet

    Auto Off After 8 minutes Display 3 color LCD backlight display IP-protection IP 67 Power supply 2 x 1,5V AAA batteries Max-Planck-Straße 16 I D- 31135 Hildesheim I Fon: +49 (0) 5121 7609-0 I Fax: +49 (0)5121 7609-44 I www.neomeris.de...
  • Page 5: Product Description

    During Setup: Press to scroll in the menu or increase the value of the selected parameter. During measurement: press to start the calibration of the selected parameter. Press to confirm the calibration and setup value. Max-Planck-Straße 16 I D- 31135 Hildesheim I Fon: +49 (0) 5121 7609-0 I Fax: +49 (0)5121 7609-44 I www.neomeris.de...
  • Page 6 • Properly dispose of waste batteries according to current legislation. • Replace all batteries together with batteries of the same type. Max-Planck-Straße 16 I D- 31135 Hildesheim I Fon: +49 (0) 5121 7609-0 I Fax: +49 (0)5121 7609-44 I www.neomeris.de...
  • Page 7: Instructions For Ph1 / Conductivity 1

    During measurement make sure that pH electrode membrane is free from air bubbles, and that there is not any air bubble around or between conductivity sensor. Max-Planck-Straße 16 I D- 31135 Hildesheim I Fon: +49 (0) 5121 7609-0 I Fax: +49 (0)5121 7609-44 I www.neomeris.de...
  • Page 8: Calibration Procedure

    Note 2: When the first point calibration is confirmed (point 7) if the sensor is not removed from the buffer solution, the instrument may give wrong buffer error. Max-Planck-Straße 16 I D- 31135 Hildesheim I Fon: +49 (0) 5121 7609-0 I Fax: +49 (0)5121 7609-44 I www.neomeris.de...
  • Page 9 During measurement make sure that pH electrode is free from air bubbles, and that there isn’t any air bubble around or between conductivity sensor. Max-Planck-Straße 16 I D- 31135 Hildesheim I Fon: +49 (0) 5121 7609-0 I Fax: +49 (0)5121 7609-44 I www.neomeris.de...
  • Page 10: Calibration Procedure

    12. The calibration process is completed and the meter is ready to use. Note: Anytime press ESC key to abort and exit from calibration procedure. Max-Planck-Straße 16 I D- 31135 Hildesheim I Fon: +49 (0) 5121 7609-0 I Fax: +49 (0)5121 7609-44 I www.neomeris.de...
  • Page 11 Store the sensor into STORAGE Solution (or pH 4) after each use, once rinsed with distilled water. • Never store pH sensor in distilled water!!! Max-Planck-Straße 16 I D- 31135 Hildesheim I Fon: +49 (0) 5121 7609-0 I Fax: +49 (0)5121 7609-44 I www.neomeris.de...
  • Page 12 25 °C 0 - 4 % / °C T coefficient 1,91 % 0 - 4,00 % / °C 1,91 % Reset Max-Planck-Straße 16 I D- 31135 Hildesheim I Fon: +49 (0) 5121 7609-0 I Fax: +49 (0)5121 7609-44 I www.neomeris.de...
  • Page 13: Errors Description

    Verification of the measuring results is the responsibility of the operator and the manufacturer does not respond to any direct or indirect damage occurred while using this instrument. Max-Planck-Straße 16 I D- 31135 Hildesheim I Fon: +49 (0) 5121 7609-0 I Fax: +49 (0)5121 7609-44 I www.neomeris.de...