Operating Manual for Digital Quick-Response Thermometer
GTH 1200 / GTH 1200 A
Specification:
Range:
Resolution:
Accuracy:
Analog output (GTH 1200A): 1mV/°C
Probe connection:
Probe:
Display:
Working temperature:
Nominal temperature:
Power supply:
Battery service life:
Low battery warning:
Dimensions:
Weight:
Other:
GKK 1100:
GKK 1200:
GB 9V:
GNG 10:
Electromagnetic
compatibility:
GREISINGER
D - 93128 Regenstauf, Hans-Sachs-Straße 26
Tel.: 09402 / 8500 od. 8748, Fax: 09402 / 1829
range 1: -65.0 to +199.9°C
range 2: -65 to +1150°C
0.1°C from -65.0 to +199.9°C (switch pos. 1)
1°C from -65 to +1150°C (switch pos. 2)
better than 0.2 % ± 0.5°C respectively ± 1 digit
standard flat mono plug
NiCr-Ni, acc. to ½ DIN 43710 for plug-in operation (not included in scope of supply !)
3 1/2 digit LCD, approx. 13 mm high
0 to 45°C
25°C
9V battery, type IEC 6 F 22 (included)
GTH 1200 A: additional mains power connector for mono plug 2.5 mm diameter
(automatic battery disconnection)
approx. 100 operating hours
"BAT"
approx. 150 x 86 x 30 mm (H x W x D) impact resistant
ABS plastic housing, with integrated pop-up clip for table top or suspended use, lateral
holders for probe handle
approx. 235g (incl. battery)
approx. 20 different standard NiCr-Ni temperature probes available (with double-accuracy
as compared to DIN standards). Please refer to list of probes!
suitable case with foame lining
suitable case with punched lining for this unit, 3 probes with handle, spare battery etc.
spare 9V battery, type IEC 6 F 22
power supply
In accordance with EN50081-1 and EN50082-2 for unrestricted use in housing and
industrial areas. Additional error: <0.5%
H04.0.03-6-e
electronic GmbH
Blatt 1 von 2