RKI Instruments 65-2320RK-100 Operator's Manual

Oxygen transmitter

Advertisement

Quick Links

Part Number: 71-0174RK
Revision: A
Released: 7/3/14
1.888.610.7664
65-2320RK-100
Oxygen Transmitter
Operator's Manual
www.calcert.com
15
10
20
5
25
0
30
sales@calcert.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 65-2320RK-100 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for RKI Instruments 65-2320RK-100

  • Page 1 65-2320RK-100 Oxygen Transmitter Operator’s Manual Part Number: 71-0174RK Revision: A Released: 7/3/14 1.888.610.7664 www.calcert.com sales@calcert.com...
  • Page 2 Typical calibration frequencies for most applications are between 3 and 6 months, but can be required more often or less often based on your usage. 65-2320RK-100 Oxygen Transmitter 1.888.610.7664 www.calcert.com sales@calcert.com...
  • Page 3 Product Warranty RKI Instruments, Inc. warrants gas alarm equipment sold by us to be free from defects in materials, workmanship, and performance for a period of one year from date of shipment from RKI Instruments, Inc. Any parts found defective within that period will be repaired or replaced, at our option, free of charge.
  • Page 4: Table Of Contents

    Parts List ..............13 65-2320RK-100 Oxygen Transmitter 1.888.610.7664...
  • Page 5: Overview

    WARNING: When using the 65-2320RK-100, you must follow the instructions and warnings in this manual to assure proper and safe operation of the 65-2320RK-100 and to minimize the risk of personal injury. Be sure to maintain and periodically calibrate the 65-2320RK-100 as described in this manual.
  • Page 6: Description

    NPT mounting threads at the top of the detector allow you to mount the detector to the junction box or a 3/4’’ NPT conduit fitting. Two color-coded leads extend from the top of the detector. The leads allow you to connect the detector to a controller. 2 • 65-2320RK-100 Oxygen Transmitter 1.888.610.7664 www.calcert.com...
  • Page 7: Amplifier

    Use the cover on the front of the junction box to access the interior of the junction box. The detector and amplifier are factory installed in the junction box. Three spacers installed on the back of the junction box control the distance of the junction box 65-2320RK-100 Oxygen Transmitter • 3 1.888.610.7664 www.calcert.com...
  • Page 8: Installation

    1 1/2-20 Thread Oxygen for Calibration Cup Detector Figure 2: Mounting the Oxygen Transmitter At the monitoring site you select, hang or mount the junction box with the detector facing down (see Figure 2). 4 • 65-2320RK-100 Oxygen Transmitter 1.888.610.7664 www.calcert.com sales@calcert.com...
  • Page 9: Wiring The Oxygen Transmitter To A Controller

    The power cable may disrupt the transmission of the transmitter signal to the controller. 11. Connect the wires to the applicable detector/transmitter terminal strip at the controller as shown in Figure 3. 65-2320RK-100 Oxygen Transmitter • 5 1.888.610.7664 www.calcert.com sales@calcert.com...
  • Page 10 12. If shielded cable is used, connect the cable’s drain wire to an available chassis (earth) ground at the controller. RKI controllers typically have a ground stud that can be used to ground the cable’s drain wire. 6 • 65-2320RK-100 Oxygen Transmitter 1.888.610.7664 www.calcert.com...
  • Page 11: Startup

    The procedure below describes applying zero emission air, usually called zero air, using a calibration kit that includes a calibration cup, calibration gas, sample tubing, and a fixed flow regulator with an on/off knob. RKI Instruments, Inc. recommends using a 0.5 LPM (liters per minute) fixed flow regulator.
  • Page 12: Maintenance

    NOTE: If the reading does not decrease, calibrate the detector as described in “Calibration” on page 11. Remove the voltmeter leads from the amplifier test points, then secure the junction box cover to the junction box. 8 • 65-2320RK-100 Oxygen Transmitter 1.888.610.7664 www.calcert.com sales@calcert.com...
  • Page 13: Troubleshooting

    3 continue, replace the detector as months. described later in this section. Some applications 5. If the calibration/response difficulties may require a more continue, contact RKI for further frequent calibration instruction. schedule. 65-2320RK-100 Oxygen Transmitter • 9 1.888.610.7664 www.calcert.com sales@calcert.com...
  • Page 14: Replacing Components Of The Oxygen Transmitter

    Step 5 above. A foam gasket that orients the amplifier and keeps it from rotating is installed on the bottom of the amplifier. Make sure the amplifier is seated flat in the junction box. 10 • 65-2320RK-100 Oxygen Transmitter 1.888.610.7664 www.calcert.com...
  • Page 15: Calibration Frequency

    Set a voltmeter to measure in the millivolt (mV) range. Plug the voltmeter leads into the test points on the amplifier. Plug the positive lead into the red + test point; plug the negative lead into the black - test point. 65-2320RK-100 Oxygen Transmitter • 11 1.888.610.7664 www.calcert.com...
  • Page 16: Setting The Fresh Air Reading

    Turn the regulator knob clockwise to close the regulator. Unscrew the regulator from the calibration cylinder. Unscrew the calibration cup from the detector housing. NOTE: For convenience, leave the regulator and calibration cup connected by the sample tubing. 12 • 65-2320RK-100 Oxygen Transmitter 1.888.610.7664 www.calcert.com sales@calcert.com...
  • Page 17: Returning To Normal Operation

    S2 series oxygen amplifier, w/orienting gasket, 0 - 100% 65-2320RK-100 Oxygen transmitter, non-explosion proof, includes detector 65-2510RK Replacement oxygen detector, non-explosion proof 71-0174RK 65-2320RK-100 Operator’s Manual (this document) 81-0076RK Zero air calibration cylinder, 17 liter steel 81-0076RK-01 Zero air calibration cylinder, 34 liter steel 81-0076RK-03...