Voltage dips
and short
interruptions on
AC mains input
lines (IEC 61000-
4-11)
This equipment is intended for use in the electromagnetic environment specified below. The customer or the user
of this equipment should assure that it is used in such an environment.
Immunity Test
Conducted
RF RF coupled
into lines (IEC
61000-4-6)
Radiated RF (IEC
61000-4-3)
EQUIPMENT CLASSIFICATION
• Protection against electric shock: Class II/internally powered equipment
• Degree of protection against electric shock: Type BF applied part
• Ingress protection: IP22
• Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen
or nitrous oxide
• Continuous operation
TROUBLESHOOTING
Problem
The system does
not start
The wrap does not
inflate
The system stopped
pumping
0.5 cycles 1 cycle 25
cycles (50
Hz) 30 cycles
(60 Hz) 250
cycles (50 Hz)
300 cycles
(60 Hz)
Compliance Level
3 Vrms
150 kHz to
80 MHz
6 Vrms in ISM bands
10 V/m
80 MHz to
2.7 GHz
Possible Causes
Power is not turned
on, Battery is not
charged
The session has not
been started
The wrap is punctured or
damaged
The session has been
stopped or ended.
Mains power quality should be that of a typical clinic,
hospital, athletic training, or home environment. If
you require continued operation during power mains
interruptions, ensure that batteries are installed and
charged. Ensure that battery life exceeds longest
anticipated power outages or provide additional
uninterruptible power source.
Electromagnetic Environment – Guidance
This device is suitable for the electromagnetic environment
of typical clinic, hospital, athletic training or home
environments.
Solutions
Press the power button to turn the control unit on.
Check that the battery is charged.
Check that the start / stop LED is ON and the timer is
counting down. If not, press start / stop button to start or
resume session.
Check that there is no air leak error message on the
display screen.
Check that the start / stop LED is ON and the timer is
counting down. If not, press start / stop button to start or
resume session.
Check that there is no air leak error message on the
display screen.
Check that there is no Low Battery message on the
display screen.
Need help?
Do you have a question about the Go and is the answer not in the manual?