Summary of Contents for 3B SCIENTIFIC ATLAS THE ALS SIMULATOR P75

  • Page 1 ATLAS THE ALS SIMULATOR P75 USER MANUAL 1023796, 1024113...
  • Page 2: Table Of Contents

    How to connect Atlas with REALITi360? ..........................................11 How to change the Wi-Fi setting on Atlas? ..........................................15 FAQ ........................................................17 Spare Parts ..................................................... 18 Technical Data ....................................................18 Contact Information ..................................................18 ATLAS THE ALS SIMULATOR P75 | Contents...
  • Page 3: Product Introduction

    ATLAS THE ALS SIMULATOR P75 PRODUCT INTRODUCTION Atlas is the ALS simulator by 3B Scientific used for conducting interdisciplinary and interprofessional emergency training. It offers extensive emergency-specific applications for advanced life support in simulation-based training and further training. DELIVERY CONTENT 1.
  • Page 4: Features

    • REALITi360 extends the therapy options to defibrillation, cardioversion and pacemaker application on the real or generic monitor. The bilateral carotid pulse feature allows to determine the indication and success of applied intervention. ATLAS THE ALS SIMULATOR P75 | Features...
  • Page 5: General Handling

    Battery status indicator on the power button: white light (100%-50%) green light (49%-15%) red light (14%-5%) red blinking (less than 5%) . Atlas is charged via the adapter provided and can be used wirelessly during training sessions. ATLAS THE ALS SIMULATOR P75 | General Handling | Safety Instructions | Installation...
  • Page 6: Airway

    BURP maneuver. To intubate use a classic laryngoscopy, or video laryngoscopy. è è è Recommended tube sizes: • Oraltracheal tube 7.5 • Nasotracheal tube 6.5 – 7 • iGel 4 • Laryngeal mask 4 • Laryngeal tube 4 ATLAS THE ALS SIMULATOR P75 | Airway...
  • Page 7 In order to see the inspiratory tidal volume use the quality CPR Dashboard and generic defibrillator screen from REALITi 360. è è è Lung Specification • Max tidal volume: 1300 mL • May airway pressure: 100 cm H₂ ATLAS THE ALS SIMULATOR P75 | Airway...
  • Page 8: Cardiopulmonary Resuscitation (Cpr)

    ✓ Detection of depth, release, and frequency of compressions ✓ Real-time view of Quality of CPR on the CPR Dashboard ✓ Use of mCPR piston-based devices ✓ Detection of hand position ✓ Compliant with ILCOR resuscitation ATLAS THE ALS SIMULATOR P75 | CPR...
  • Page 9: Circulation

    When performing administration of medication via the IV pad and /or IO access site, use a drain bag which should be connected to the Luer lock adapter. Atlas has one IO drain at the top of the Atlas’ left arm and one at the top of the IV. è Recommended puncture sizes • IV: ≤18G • IO: drilling system ATLAS THE ALS SIMULATOR P75 | Circulation...
  • Page 10 When the day’s sessions are done, flush the IV and IO sites with air to remove any fluid/liquid in the system. Connect a syringe filled with air to the IV/IO catheter and flush the IV pads/IO insert with air until only air exits the overflow tube. Replacement of IV sites Replacement of IO site ATLAS THE ALS SIMULATOR P75 | Circulation...
  • Page 11: Network/Router Settings

    4. On the Monitor iPad Pro, start the REALITi application and router automatically. The power button is located select the “Staging” button: on the right side of the manikin. ATLAS THE ALS SIMULATOR P75 | Network/Router Settings | Connect Atlas with REALITi360...
  • Page 12 Wi-Fi network). 8. After selecting the monitor name on the Control iPad, it will automatically switch to the options of starting with a scenario (red outline) or a monitor (blue outline): ATLAS THE ALS SIMULATOR P75 | Connect Atlas with REALITi360...
  • Page 13 11. Swipe in from the right screen edge of the monitor iPad show the CPR connection dialog on the Defibrillator: 12. Tap “Scan for Manikin”. Select the manikin appearing in the “Available Manikin” table and tap “Connect.” ATLAS THE ALS SIMULATOR P75 | Connect Atlas with REALITi360...
  • Page 14 ATLAS THE ALS SIMULATOR P75 13. The connected manikin will be calibrated automatically: 14. Compress the manikin. You should be able to see the CPR compressions in the ECG waveform and the bar graph. ATLAS THE ALS SIMULATOR P75 | Connect Atlas with REALITi360...
  • Page 15: How To Change The Wi-Fi Setting On Atlas

    4. On the Monitor iPad Pro, start the REALITi application and iSimulate-5.0 automatically. The power button is located select the “Staging” button:: on the right side of the manikin. ATLAS THE ALS SIMULATOR P75 | Network/Router Settings | Wi-Fi Setting...
  • Page 16 NOTE: Atlas connects to 2.4 GHz frequency. We recommend to set up the networks, 2.4GHz & 5.0GHz to the same SSID and password to ensure efficiency in your training. NOTE: Maximum character length of 19 characters for username and password. ATLAS THE ALS SIMULATOR P75 | Wi-Fi Setting...
  • Page 17: Faq

    The following may be the cause: • The blood pressure is set below 50mmHg Systolic. • The connection is lost between the REALITi 360 system and Atlas. • Non-perfusing cardiac rhythm • PEA rhythm ATLAS THE ALS SIMULATOR P75 | FAQ...
  • Page 18: Spare Parts

    15% - 80% CONTACT INFORMATION A worldwide group of companies 3B Scientific GmbH Ludwig-Erhard-Straße 20 • 20459 Hamburg • Germany Phone: +  49   (0)40-73966-0 • Fax: +  49   (0)40-73966-100 3bscientific.com • info@3bscientific.com ATLAS THE ALS SIMULATOR P75 | Spare Parts | Technical Data | Contact Information...