Multi-measuring device for workplace analysis BAPPUevo Inhalt BAPPUevo 1.1. Symbols......................5 1.2. Security alert.....................5 Use personal protective equipment (PPE)..........5 1.3. General information..................6 1.4. Qualification of the operating personnel............6 1.5. Observe accident prevention regulations............6 1.6. Intended use.....................6 1.7. Technical changes....................7 1.8. Initial assembly....................7 1.9.
Page 4
BAPPUevo Multi-measuring device for workplace analysis Function buttons................18 3.2. Keyboard.....................18 3.3. Connections BAPPUevo (from version 4000/11)........18 Connectors VOCOOsx as well as ANEMOMETERevo....19 3.4. Help bar.......................19 3.5. Background Colour..................19 Main menu 4.1. BAPPU-measurement.................20 4.2. Recording....................20 4.3. Settings.......................20 Measurement preparation 5.1. Edit measurement series profile (Profile)............21 5.2.
Page 5
Multi-measuring device for workplace analysis BAPPUevo 8.6. Measurement profile (Profile).................30 BAPPU standard software 9.1. License general terms and conditions............31 9.2. Installation.......................31 9.3. Starting the programme and connecting BAPPU...........32 9.4. Settings......................32 9.5. Online......................32 9.6. Extra........................32 9.7. Defining the type (defining the type of work place).........32 9.8.
Multi-measuring device for workplace analysis BAPPUevo 1.1. Symbols The following symbols are used in this operating manual: This note indicates the risk of injury and / or death as well as health hazard, if certain rules of conduct are disregarded. Important notes and information for installation, operation and Tipp: Tipp:...
BAPPUevo Multi-measuring device for workplace analysis 1.3. General information Risk of burns: The product is subjected to an operating temperature of up to 50 ° C. There is a risk of burns on housing parts. Avoid installation errors: The product may only be installed by qualified personnel of the operator.
Multi-measuring device for workplace analysis BAPPUevo Any use beyond its intended use is deemed to be contrary to the intended purpose. The manufacturer is not liable for resulting damages. Proper use also includes observing and following all instructions in this documentation. 1.7.
Your local government will be able to provide you with the respective collection points and their opening hours. 1.13. Support ELK GmbH | Gladbacher Straße 232 | D-47805 Krefeld Fon: +49 (0) 2151 392829 | Fax: +49 (0) 2151 391846 I info@bappu.de | www.bappu.de 1.14.
Multi-measuring device for workplace analysis BAPPUevo Measuring range and tolerances Measurement Range Tolerance Sensor Resolution Air temperature -20...50 °C +/- 0,5 °C PT 1000 Sensor 0,1 °C 0...70 °C +/- 0,7 °C Integ. temperature Globe semiconductor -20...70 °C +/- 2 °C 0,1 °C temperature* sensor...
BAPPUevo Multi-measuring device for workplace analysis Specifications of the measurement with VOCOOsx Caution: The VOC sensor needs after switching on about 3 minutes, to stabilize. During the heating phase, the “VOC” LED flashes during the heating phase. Available sensors in VOCOOsx When switched on, the LEDs of the VOCOOsx signal the available sensors.
Multi-measuring device for workplace analysis BAPPUevo Frequency weighting and permissible errors Permissible error Nominal frequency in Hz A-evaluation in dB class 2 in dB 31,5 -39,4 +/- 3 1000 +/- 1,5 4000 +/- 3 8000 -1,1 +/- 5 Equivalent continuous sound level In addition to the sound level measurement, the equivalent continuous sound level of the past minute is continuously determined (noise...
BAPPUevo Multi-measuring device for workplace analysis Using the optional sound level calibrator The sound level calibrator can be used to check the "Noise" measurand of the BAPPUevo at a level of 94 dB and a frequency of 1 kHz. To calibrate the "Noise" measurand, proceed as follows: You are inside a measurement of the measurand "Noise".
Multi-measuring device for workplace analysis BAPPUevo Special features for the measurement of illuminance BAPPU-evo is used to measure lighting conditions at a typical work place, including, for example, those specified in DIN 5035-6:2006-11. With other conditions, larger measuring tolerances may occur. Permissible error class C Description BAPPU-evo...
BAPPUevo Multi-measuring device for workplace analysis • Make sure that there are no disturbances such as dust, steam, etc. between the sensor and the measurement object. • The sensor and the measuring surfaces must be free of dirt. • The emissivity must be adjusted to the measurement surface (see table). •...
Multi-measuring device for workplace analysis BAPPUevo Supply voltage The power supply can be provided by the power supply unit, the rechargeable battery or, in exceptional cases, by the battery. If the voltage is too low, the unit switches off. GP ReCyko+ NiMh HR6 210AAHCB type 2050mAh 1.2V batteries are recommended for trouble-free operation.
BAPPUevo Multi-measuring device for workplace analysis 2. Glossary 2.1. BAPPU-measurement In BAPPU-evo, "BAPPU measurement series" refers to the acquisition of instantaneous values of a measured variable and their internal storage under a defined name (workplace). Despite the instantaneous character of some measured quantities, averages over several minutes are available (noise level, air velocity).
Multi-measuring device for workplace analysis BAPPUevo 2.6. Measurement result overview (overview) In the overview, all important measured values and measurement data of the current workplace are displayed. 2.7. Measurement value details The detailed view shows the value stored in the memory for a measured variable and its evaluation.
BAPPUevo Multi-measuring device for workplace analysis Sensitive areas Analogous to the keyboard, the functions of the control pad are also available in many places on the display (see figure Function buttons The function keys F1, F2 and F3 are located at the bottom of the display and trigger different functions according to their labels.
Multi-measuring device for workplace analysis BAPPUevo Connectors VOCOOsx as well as ANEMOMETERevo VOCOOsx ANEMOMETERevo 3.4. Help bar At the top of the screen there is the help line describing the valid keyboard functions. The assignment of the right/left arrow keys, the up/down arrow keys and the function of the OK key are shown. Please note that additional functions may be enabled via the three function keys at the bottom of the screen.
BAPPUevo Multi-measuring device for workplace analysis 4.1. BAPPU-measurement By selecting the menu item "BAPPU measurement series" you will see an overview of the workplace, the workplace list. Edit the workplace (name, workplace type, measurement series profile) and/or start the measurement. Measurement series marked with a "* "...
Multi-measuring device for workplace analysis BAPPUevo 5. Measurement preparation 5.1. Edit measurement series profile (Profile) By means of measurement series profiles you define which measurement types are to be recorded in the measurement series. The "All" (all measured variables) and "Standard" profiles are available. In addition, you can set eight user-defined profiles.
BAPPUevo Multi-measuring device for workplace analysis T i p : T i p : All editing functions can be easily carried out using the software and can be transferred to the device. (please refer to page 35, chapter, 9.8 ”Worksplace names and type definition (data retrieval)”).
Multi-measuring device for workplace analysis BAPPUevo 6. Measurement sequen 6.1. Perform measurement In order to carry out the measurement sequence, select the work place from the work place list. To start measuring, press . According to the profile selected, you will now start your first measurement.
BAPPUevo Multi-measuring device for workplace analysis T i p : T i p : The background colour of the display always shows you the current mode (light background = measuring mode, dark background = save mode). During the illuminance measurement, the self-timer function can be activated by pressing “hold”.
Multi-measuring device for workplace analysis BAPPUevo Using , you can browse the overview, which always comprises two screens. Pressing the keys calls up the first screen of the following work place. Keying in “delete” will delete all of the measured values from the current work place. 6.3.
BAPPUevo Multi-measuring device for workplace analysis On the line below the symbols, you can view the settings for the logging intervals and the memory type. In the middle there is a preview of the recorded data. The measuring range that is outlined with green in the measuring range symbols is always displayed.
Multi-measuring device for workplace analysis BAPPUevo The recording can be stopped by pressing “Stop” (F3) or when the memory capacity is reached. The data can be read using the optional “BAPPU-time” software (please refer to page 39., chapter 10.6, „Operating and reading the internal data counter“). 7.4.
BAPPUevo Multi-measuring device for workplace analysis 8.2. Display/Energy Here you will find the display and energy functions where you can select the option required with . Using the button, you can select the modes that are available 8.3. Measurement settings (measurement) The following sections will explain how the settings to the individual measuring ranges can be made.
Multi-measuring device for workplace analysis BAPPUevo Attention: If the level range is set manually, the occurrence of an overshoot or undershoot of the range is more likely. C-peak measurement (C-peak measurement) By selecting "On", the C-weighted peak sound level (C-peak) of the past minute is measured.
BAPPUevo Multi-measuring device for workplace analysis 8.5. Firmware update We recommend that you update your BAPPU-evo as soon as new firmware is available. The appropriate update data is available from our website. Copy the data onto a USB flash device, and insert this into the USB port of your BAPPU-evo.
All rights, especially the right to reproduce (the exception being for the purpose of installing the program on the user's harddrive) and distribution as well as the processing of the program in any way whatsoever, remain at ELK GmbH. •...
BAPPUevo Multi-measuring device for workplace analysis 9.3. Starting the programme and connecting BAPPU You start the programme by clicking “Start”, “Program”, “BAPPU-time”. Subsequently, the start image will appear followed by the main window (refer to chapter 9.4 - 9.8). When BAPPU-evo is correctly connected to a PC via the USB port, the message “Reading type of work place”...
Page 35
Multi-measuring device for workplace analysis BAPPUevo The set values can be entered in the “Defining the type of work place” dialog field. This window can be opened by clicking the “Type definition” button in the main window. Select the function “Data/new” from the menu item “File”, and enter a set of appropriate initials (a maximum of 4 characters) in the window to identify the type of work place.
Page 36
BAPPUevo Multi-measuring device for workplace analysis The coloured blocks making up the value bar represent the areas where the “Traffic- light” function for conformity control is indicated. You can use the mouse to move the coloured blocks to the values where you would like to set the threshold values for the respective measurement range.
Multi-measuring device for workplace analysis BAPPUevo 9.8. Worksplace names and type definition (data retrieval) Select “Data retrieval” in the main window of the program. The data retrieval window is used to prepare workstation analyses and to display the measurement results. Named identifiers of the storage locations, workstation type definitions and measurement series profile assignments can be sent from this window to your BAPPU or measurement series can be transferred from your BAPPU to the To begin a new project, select “File/New”, or if you want to make any changes to an...
BAPPUevo Multi-measuring device for workplace analysis 9.10. User-defined display In order to prevent the data retrieval window from becoming too big, the system does not, by default, show all of the individual values. Select “Table view/Settings” (user defined), if you would like to set up the window to show additional information.
The license for the Bappu-time software is assigned to a specific multimeter BAPPU in each case. For the operation of an installation with several multimeters BAPPU you need accordingly several licenses. Licenses are available exclusively from ELK GmbH. page 37...
BAPPUevo Multi-measuring device for workplace analysis 10.2. Installation To be able to install BAPPU-time, it is necessary that your computer is equipped with the Windows 10 operating system. The start window of the CD "BAPPU-time" will be opened automatically after inserting the CD into the corresponding drive. If this is not the case, open the start window by double-clicking on "BappuTimeStart.exe"...
Multi-measuring device for workplace analysis BAPPUevo 10.6. Operating and reading the internal data counter Data Logger port To configure or read the BAPPU-evo internal data counter (which will be referred to as “counter”), the correct port has to be selected and the BAPPU-evo should be selected as the device of choice.
BAPPUevo Multi-measuring device for workplace analysis T i p : T i p : Before a new recording, the data in the counter should be deleted in order to be able to get the maximum recording time. You can see if there is any data in the counter by clicking “Counter status”.
Multi-measuring device for workplace analysis BAPPUevo The “Loger status” button is used for calling up details pertaining to the status of the memory content. The “Read counter” function is for transferring data stored on the counter to your PC. A progress bar shows the status of the transfer. When the transfer is complete, the recordings already stored in the memory will be listed in the “Readout window”...
Page 44
BAPPUevo Multi-measuring device for workplace analysis Clicking the respective measurement Air temperature /dew point range symbol will enable you to display the separate recorded areas Rel. humidity consecutively in the window. It is also possible to display several Noise Level measurement ranges simultaneously.
Multi-measuring device for workplace analysis BAPPUevo 10.8. Display options for the graphs The white box at the top right hand side of the window shows the colour legend used for presenting curves. They can be displayed separately by clicking on the abbreviation in the white box or on the header.
BAPPUevo Multi-measuring device for workplace analysis Zoom in/zoom out When the zoom function has been selected, a “+” will appear next to the magnifying glass cursor. You can then click on a point that you want to enlarge or, with the left mouse button pressed down, move the mouse from left to right to draw a frame around the selection you would like to view.
Multi-measuring device for workplace analysis BAPPUevo 10.9. Selecting a recording When part of a recording needs to be edited, you can select this in the “Cut” window. Select “Recording/Cut out recording” in the main menu. Then enter the time frame of the selection you would like to work on in the dialog box.
BAPPUevo Multi-measuring device for workplace analysis When contact between BAPPU-evo and BAPPU-time software has been established, the next step is to select “Recording” and then “Start recording” in order to begin a long term recording. Alternatively, you can press . Following this, the “Recording settings” dialog box will appear.
Multi-measuring device for workplace analysis BAPPUevo Online-Monitor During the recording it is possible to view the measurement trend in the data monitor as well as call up the data which has already been stored in the graphic window. In order to call up the monitor function press in the menu list.
BAPPUevo Multi-measuring device for workplace analysis By clicking the measured values symbol in the top left hand side of the screen, you can switch between the different measurement range views. The end of the recording A recording will end when the specified time frame expires. BAPPU-time confirms this with a pop up window that states: “The recording was terminated automatically”.
Multi-measuring device for workplace analysis BAPPUevo Printing If you would like to print your BAPPU-time graphs, select either “Print file” in the main menu or right click your mouse and select “Print” from the list. Additional options such as print layout and the print quality can be selected in print preview.
BAPPUevo Multi-measuring device for workplace analysis The operator must ensure that all employees who handle the product have read and understood the instructions. In addition, he must instruct the staff at regular intervals and inform them about the dangers. The operator must provide the personnel with the necessary protective equipment and must instruct them to wear the necessary protective equipment.
Page 53
Krefeld / Germany, march 2023 The sole responsibility for issuing this declaration of conformity lies with the manufacturer. ELK GmbH – Ingenieurbüro für Elektronik Gladbacher Str. 232 • D-47805 Krefeld fon. +49 2151-788 86-0 • fax. +49 2151-788 86-02 www.elk.de •...
Page 54
The new standard in instrumentation ELK GmbH Ingenieurbüro für Elektronik Gladbacher Str. 232 • D-47805 Krefeld fon: 02151-788 86-0 • fax: 02151-788 86-02 info@bappu.de • www.bappu.de Brand of ELK GmbH: BAPPUevo...
Need help?
Do you have a question about the BAPPU evo and is the answer not in the manual?
Questions and answers