Download Print this page

GREISINGER electronic GTMU o. MU Operating Manual

Advertisement

Quick Links

T01.0.0X.6C-01
Specification:
Measuring range:
Sensor element:
Sensor accuracy:
Output signal:
Connection:
Accuracy electronic:
Operating temperature:
Relative humidity:
Storage temperature:
Mounting position:
Housing:
Mounting:
Mounting distance:
Design-types:
Electric connection:
GREISINGER
D - 93128 Regenstauf, Hans-Sachs-Straße 26
Fon: 0049 9402/9383-0, Fax: 0049 9402/9383-33, e-mail: info@greisinger.de
Operating Manual for Temperature Sensor
GTMU o. MU
dependent on the sensor construction (see type plate)
Note: When using with temperatures over 70°C, there must be at least 50mm probe
lenght outside the media.
please refer to type plate
- resistance thermometer Pt100 acc. to DIN IEC 751
- thermocouple NiCr-Ni (typ K), class 1 acc. to DIN IEC 584
please refer to type plate
please refer to type plate
Pt100: 2-, 3- or 4-wire
NiCr-Ni: 2-wire
Pt100: DIN class B (or see type plate)
NiCr-Ni: Class 1
0 to 70°C (housing)
0 to 95 % r.h. (non-condensing)
-45 to 85°C
any position
ABS (IP65), 82 x 80 x 55 mm (without elbow-type plug and pressure connection)
By means of screw thread or mounting holes in housing (accessible after cover has been
removed).
50 x 70mm, max. shaft diameter of mounting screws is 4mm.
design-type 1: with threaded stem for direct screw connection.
design-type 2: with threaded stem at a distance from housing (for higher temperature).
design-type 3: indoor /outdoor probe for direct wall mounting.
design-type 4: duct type probe with centrally mounted sensor tube arrangement with the
shaft of device pointing downwards at a 90° angle
design-type 5: measuring transducer for Pt100 or NiCr-Ni sensors already existing.
Insertion of sensor cable through PG7 screwing.
elbow-type plug conforming to DIN 43650 (IP65),
max. wire cross section: 1.5 mm², wire/cable diameter from 4.5 to 7 mm
electronic GmbH
page 1 v. 2

Advertisement

loading
Need help?

Need help?

Do you have a question about the GTMU o. MU and is the answer not in the manual?

Questions and answers

Summary of Contents for GREISINGER electronic GTMU o. MU

  • Page 1 T01.0.0X.6C-01 page 1 v. 2 Operating Manual for Temperature Sensor GTMU o. MU Specification: Measuring range: dependent on the sensor construction (see type plate) Note: When using with temperatures over 70°C, there must be at least 50mm probe lenght outside the media.
  • Page 2 T01.0.0X.6C-01 page 2 v. 2 General installation instructions of elbow-type plug: To mount the connection cable (2-, 3-, or 4-wire depending on type of device) the angle plug screw has to be loosened and the coupling insert has to be removed by means of a screw driver at the position indicated (arrow). Pull out connection cable through PG glanding and connect to the loose coupling insert as described in the wiring dia- gram.